FDA Enforcement
Class II
Terminated
BECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370
Recall: Z-1362-2018
·
Reported April 18, 2018
Enforcement
- Recall Number
- Z-1362-2018
- Event ID
- 79566
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- April 18, 2018
- Initiation Date
- March 2, 2018
- Classification Date
- April 11, 2018
- Termination Date
- July 6, 2023
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232, United States
Description
BECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370
Reason
Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.
Code Info
All
Distribution
Worldwide
Quantity
5247 units