FDA Enforcement Class II Terminated

BECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370

Recall: Z-1362-2018 · Reported April 18, 2018

Enforcement

Recall Number
Z-1362-2018
Event ID
79566
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
April 18, 2018
Initiation Date
March 2, 2018
Classification Date
April 11, 2018
Termination Date
July 6, 2023
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232, United States

Description

BECKMAN COULTER iQ200 Series Urine Microscopy Analyzer, Catalog Numbers: 700-3345, C10684, 700-3375, C10683, 700-3325, 700-3347, 700-3320, 700-3322, 700-3370

Reason

Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.

Code Info

All

Distribution

Worldwide

Quantity

5247 units