FDA Enforcement Class II Ongoing

Bio-Rad, REF: 12005660, QXDx BCR-ABL %IS Kit

Recall: Z-1358-2024 · Reported April 3, 2024

Enforcement

Recall Number
Z-1358-2024
Event ID
94020
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Bio-Rad Laboratories
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 3, 2024
Initiation Date
February 15, 2024
Classification Date
March 22, 2024
Address
5731 W Las Positas Blvd, N/A, Pleasanton, CA, 94588-4084, United States

Description

Bio-Rad, REF: 12005660, QXDx BCR-ABL %IS Kit

Reason

Due to incorrect calibrations and controls, their is a potential of incorrect reporting of lower %IS and molecular response values.

Code Info

Lot # 64562656 & 64571846/ UDI: 03610521162834

Distribution

US Nationwide distribution in the states of PA, SD, TX, FL, MN.

Quantity

33 units