FDA Enforcement Class I Ongoing

Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF, Product Description): 10CN10H, 10.0MM SHILEY CUFFED TRACH CAN; 10CN10R, 10.0MM ADT FLEX TRACH W TG CUFF; 10UN10A, 10.0MM UNCUFF TRACH TUBE X1; 10UN10H, 10.0MM SHILEY CUFFLESS TRACH C; 10UN10R, 10.0MM ADT FLEX TRACH CUFFLESS; 4CN65A, 6.5MM TRACH TUBE W TG CUFF X1; 4CN65H, 6.5MM SHILEY CUFFED TRACH CANN; 4CN65R, 6.5MM ADT FLEX TRACH W TG CUFF X1; 4UN65A, 6.5MM UNCUFF TRACH TUBE X1; 4UN65H, 6.5MM SHILEY CUFFLESS TRACH CA; 4UN65R, 6.5MM ADT FLEX TRACH CUFFLESSX1; 5CN70A, 7.0MM TRACH TUBE W TG CUFF X1; 5CN70H, 7.0MM SHILEY CUFFED TRACH CANN; 5CN70R, 7.0MM ADT FLEX TRACH W TG CUFF X1; 5UN70A, 7.0MM UNCUFF TRACH TUBE X1; 5UN70H, 7.0MM SHILEY CUFFLESS TRACH CA; 5UN70R, 7.0MM ADT FLEX TRACH CUFFLESSX1; 6CN75A, 7.5MM TRACH TUBE W TG CUFF X1; 6CN75H, 7.5MM SHILEY CUFFED TRACH CANN; 6CN75R, 7.5MM ADT FLEX TRACH W TG CUFF X1; 6UN75H, 7.5MM SHILEY CUFFLESS TRACH CA; 6UN75R, 7.5MM ADT FLEX TRACH CUFFLESSX1; 7CN80A, 8.0MM TRACH TUBE W TG CUFF X1; 7CN80H, 8.0MM SHILEY CUFFED TRACH CANN; 7CN80R, 8.0MM ADT FLEX TRACH W TG CUFF X1; 7UN80A, 8.0MM UNCUFF TRACH TUBE X1; 7UN80H, 8.0MM SHILEY CUFFLESS TRACH CA; 7UN80R, 8.0MM ADT FLEX TRACH CUFFLESSX1; 8CN85A, 8.5MM TRACH TUBE W TG CUFF X1; 8CN85H, 8.5MM SHILEY CUFFED TRACH CANN; 8CN85R, 8.5MM ADT FLEX TRACH W TG CUFF X1; 8UN85A, 8.5MM UNCUFF TRACH TUBE X1; 8UN85H, 8.5MM SHILEY CUFFLESS TRACH CA; 8UN85R, 8.5MM ADT FLEX TRACH CUFFLESSX1; 9CN90H, 9.0MM SHILEY CUFFED TRACH CANN; 9CN90R, 9.0MM ADT FLEX TRACH W TG CUFF X1; 9UN90H, 9.0MM SHILEY CUFFLESS TRACH CA; 9UN90R, 9.0MM ADT FLEX TRACH CUFFLESSX1

Recall: Z-1357-2023 · Reported April 26, 2023

Enforcement

Recall Number
Z-1357-2023
Event ID
91943
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Covidien
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 26, 2023
Initiation Date
March 24, 2023
Classification Date
April 19, 2023
Address
6135 Gunbarrel Ave, N/A, Boulder, CO, 80301-3214, United States

Description

Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF, Product Description): 10CN10H, 10.0MM SHILEY CUFFED TRACH CAN; 10CN10R, 10.0MM ADT FLEX TRACH W TG CUFF; 10UN10A, 10.0MM UNCUFF TRACH TUBE X1; 10UN10H, 10.0MM SHILEY CUFFLESS TRACH C; 10UN10R, 10.0MM ADT FLEX TRACH CUFFLESS; 4CN65A, 6.5MM TRACH TUBE W TG CUFF X1; 4CN65H, 6.5MM SHILEY CUFFED TRACH CANN; 4CN65R, 6.5MM ADT FLEX TRACH W TG CUFF X1; 4UN65A, 6.5MM UNCUFF TRACH TUBE X1; 4UN65H, 6.5MM SHILEY CUFFLESS TRACH CA; 4UN65R, 6.5MM ADT FLEX TRACH CUFFLESSX1; 5CN70A, 7.0MM TRACH TUBE W TG CUFF X1; 5CN70H, 7.0MM SHILEY CUFFED TRACH CANN; 5CN70R, 7.0MM ADT FLEX TRACH W TG CUFF X1; 5UN70A, 7.0MM UNCUFF TRACH TUBE X1; 5UN70H, 7.0MM SHILEY CUFFLESS TRACH CA; 5UN70R, 7.0MM ADT FLEX TRACH CUFFLESSX1; 6CN75A, 7.5MM TRACH TUBE W TG CUFF X1; 6CN75H, 7.5MM SHILEY CUFFED TRACH CANN; 6CN75R, 7.5MM ADT FLEX TRACH W TG CUFF X1; 6UN75H, 7.5MM SHILEY CUFFLESS TRACH CA; 6UN75R, 7.5MM ADT FLEX TRACH CUFFLESSX1; 7CN80A, 8.0MM TRACH TUBE W TG CUFF X1; 7CN80H, 8.0MM SHILEY CUFFED TRACH CANN; 7CN80R, 8.0MM ADT FLEX TRACH W TG CUFF X1; 7UN80A, 8.0MM UNCUFF TRACH TUBE X1; 7UN80H, 8.0MM SHILEY CUFFLESS TRACH CA; 7UN80R, 8.0MM ADT FLEX TRACH CUFFLESSX1; 8CN85A, 8.5MM TRACH TUBE W TG CUFF X1; 8CN85H, 8.5MM SHILEY CUFFED TRACH CANN; 8CN85R, 8.5MM ADT FLEX TRACH W TG CUFF X1; 8UN85A, 8.5MM UNCUFF TRACH TUBE X1; 8UN85H, 8.5MM SHILEY CUFFLESS TRACH CA; 8UN85R, 8.5MM ADT FLEX TRACH CUFFLESSX1; 9CN90H, 9.0MM SHILEY CUFFED TRACH CANN; 9CN90R, 9.0MM ADT FLEX TRACH W TG CUFF X1; 9UN90H, 9.0MM SHILEY CUFFLESS TRACH CA; 9UN90R, 9.0MM ADT FLEX TRACH CUFFLESSX1

Reason

A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.

Code Info

REF/UDI-DI/Lot: 10CN10H/10884521172371/21C0402JZX, 21D0461JZX, 21G0457JZX, 21I0293JZX, 21I0294JZX, 21I0552JZX, 21J1435JZX, 21J1436JZX, 21K0275JZX, 21K0276JZX, 21K0277JZX, 21K0278JZX; 10CN10R/10884521205475/202107218X, 202107296X, 202108128X, 202108285X, 21H0792JZX, 21H0793JZX, 21H0794JZX, 22I0596JZX; 10UN10A/20884521175690/21D0463JZX; 10UN10H/20884521176291/21E0661JZX, 21G0473JZX; 10UN10R/10884521205543/202105307X, 202105308X, 202106148X, 202209335X; 4CN65A/10884521172456/21C0432JZX, 21I0369JZX; 4CN65H/20884521176307/21C0358JZX, 21C0359JZX, 21C0360JZX, 21C0361JZX, 21D0440JZX, 21D0441JZX, 21E0512JZX, 21E0513JZX, 21E0514JZX, 21E0515JZX, 21E0516JZX, 21E0517JZX, 21E0518JZX, 21E0519JZX, 21E0520JZX, 21E0521JZX, 21E0522JZX, 21E0523JZX, 21E0524JZX, 21E0525JZX, 21E0526JZX, 21E0527JZX, 21E1773JZX, 21E1774JZX, 21E1775JZX, 21E1776JZX, 21E1777JZX, 21E1778JZX, 21E1779JZX, 21E1780JZX, 21E1781JZX, 21E1782JZX, 21E1783JZX, 21H1324JZX, 21J0181JZX, 21J1094JZX, 21K0237JZX, 21K0239JZX, 21K0639JZX, 21K0643JZX, 21K0645JZX, 22C1465JZX, 22C1466JZX, 22C1467JZX, 22D0798JZX, 22D0799JZX, 22D0800JZX, 22D0801JZX, 22D0891JZX, 22D0892JZX, 22D0893JZX, 22D0897JZX, 22D0898JZX, 22D0899JZX, 22D0900JZX, 22D0901JZX, 22D0909JZX, 22F0918JZX, 22F0919JZX, 22F0920JZX, 22F0921JZX, 22F0922JZX, 22F0923JZX, 22F0924JZX, 22F0925JZX, 22F0926JZX, 22F0927JZX, 22F0928JZX, 22F0929JZX, 22J0073JZX, 22J0074JZX, 22J0075JZX, 22J0076JZX, 22J0077JZX, 22J0080JZX, 22J0081JZX, 22J0083JZX; 4CN65R/10884521205024/202105082X, 202105310X, 202106267X, 202106268X, 202107294X, 202107295X, 202111088X, 202111270X, 202111399X, 202112135X, 202201303X, 202202186X; 4UN65A/20884521175706/21D0466JZX, 21E0659JZX, 21E0660JZX; 4UN65H/20884521176239/21C0413JZX, 21C0414JZX, 21C0416JZX, 21C0417JZX, 21D0898JZX, 21D0899JZX, 21D0900JZX, 21D0901JZX, 21E1784JZX, 21E1785JZX, 22D0930JZX, 22D0931JZX, 22D0932JZX, 22D0933JZX, 22E0679JZX, 22E0680JZX; 4UN65R/10884521205482/202107297X, 202107298X, 202107299X, 202107300X, 202108122X, 202108123X, 202108124X, 202108125X, 202108139X, 202108140X, 202108228X, 202108229X, 21E1139JZX; 5CN70A/10884521172463/22D0927JZX, 22D0928JZX, 22D0929JZX, 22H0031JZX, 22H0032JZX, 22H0033JZX, 22H0860JZX, 22H0861JZX, 22H0863JZX, 22H0865JZX, 22J0071JZX, 22J0072JZX; 5CN70H/20884521176314/21C0404JZX, 21C0405JZX, 21C0406JZX, 21D0462JZX, 21G1130JZX, 21I0283JZX, 21I0284JZX, 21I0285JZX, 21I0286JZX, 21I0519JZX, 21I0520JZX, 21I0521JZX, 21I0522JZX, 21I0544JZX, 21I0545JZX, 21I0546JZX, 21I0630JZX, 21I0631JZX, 22D0924JZX, 22D0925JZX, 22D0926JZX, 22H0152JZX; 5CN70R/10884521205420/202111135X, 202208294X, 202208295X, 202210208X; 5UN70A/20884521175713/21B0062JZX, 21E0645JZX, 21E0646JZX, 21E0647JZX; 5UN70H/20884521176246/21E0655JZX, 21E0656JZX, 21G1008JZX, 21G1009JZX, 21H0735JZX; 5UN70R/10884521205499/202208296X; 6CN75A/10884521172470/22A0137JZX, 22H0868JZX; 6CN75H/20884521176321/21D0420JZX, 21D0421JZX, 21D0422JZX, 21D0423JZX, 21D0424JZX, 21D0425JZX, 21D0426JZX, 21D0427JZX, 21D0428JZX, 21D0429JZX, 21D0430JZX, 21D0431JZX, 21D0432JZX, 21D0433JZX, 21D0434JZX, 21D0435JZX, 21D0436JZX, 21D0437JZX, 21D0438JZX, 21D0439JZX, 21D0741JZX, 21D0742JZX, 21D0743JZX, 21D0744JZX, 21D0745JZX, 21D0746JZX, 21D0747JZX, 21D0748JZX, 21D0749JZX, 21E0632JZX, 21E0633JZX, 21E0634JZX, 21E0635JZX, 21E0636JZX, 21E0637JZX, 21E0638JZX, 21E0639JZX, 21E0640JZX, 21E0641JZX, 21E0642JZX, 21E0643JZX, 21E0644JZX, 21E1100JZX, 21E1101JZX, 21E1673JZX, 21E1674JZX, 21E1675JZX, 21E1676JZX, 21E1677JZX, 21E1678JZX, 21E1679JZX, 21E1680JZX, 21E1681JZX, 21E1682JZX, 21E1683JZX, 21E1684JZX, 21E1685JZX, 21E1686JZX, 21E1687JZX, 21E1688JZX, 21E1689JZX, 21E1690JZX, 21E1691JZX, 21E1694JZX, 21E1695JZX, 21E1696JZX, 21E1697JZX, 21E1769JZX, 21F0692JZX, 21F0693JZX, 21F0694JZX, 21G0296JZX, 21G0297JZX, 21G0298JZX, 21G0299JZX, 21G0426JZX, 21G0427JZX, 21G0428JZX, 21G0429JZX, 21G0430JZX, 21H0624JZX, 21I0260JZX, 21I0261JZX, 21I0262JZX, 21I0263JZX, 21I0264JZX, 21I0265JZX, 21I0266JZX, 21I0267JZX, 21I0268JZX, 21I0269JZX, 21I0270JZX, 21I0271JZX, 21I0491JZX, 21I0492JZX, 21I0680JZX, 21I0681JZX, 21I0682JZX, 21I0683JZX, 21J0090JZX, 21J0168JZX, 21J0169JZX, 21J0170JZX, 21J0171JZX, 21J0172JZX, 21J0176JZX, 21J0177JZX, 21J1088JZX, 21J1410JZX, 21J1411JZX, 21J1417JZX, 21J1418JZX, 21L0549JZX, 21L0550JZX, 21L0551JZX, 21L0552JZX, 21L0553JZX, 21L0756JZX, 21L0777JZX, 21L0778JZX, 21L0779JZX, 21L0796JZX, 21L0797JZX, 21L0801JZX, 21L0802JZX, 21L0803JZX, 21L0804JZX, 21L0805JZX, 21L0806JZX, 21L0807JZX, 21L0808JZX, 21L0809JZX, 21L0810JZX, 21L0811JZX, 21L0812JZX, 21L0813JZX, 21L0814JZX, 21L0815JZX, 21L0816JZX, 21L0817JZX, 21L0818JZX, 21L0819JZX 22A0665JZX, 22A0666JZX, 22A0670JZX, 22A0671JZX, 22A0672JZX, 22A0673JZX, 22A0674JZX, 22A0771JZX, 22A0772JZX, 22A0900JZX, 22A0901JZX, 22A0902JZX, 22A0903JZX, 22A0904JZX, 22A0905JZX, 22A0906JZX, 22A0907JZX, 22A0908JZX, 22A0909JZX, 22A0910JZX, 22A0911JZX, 22A0912JZX, 22A0913JZX, 22A0914JZX, 22A0915JZX, 22A0916JZX, 22B0277JZX, 22B0280JZX, 22B0288JZX, 22B0289JZX, 22B0290JZX, 22B0291JZX, 22B0334JZX, 22B0335JZX, 22B0336JZX, 22B0337JZX, 22C0758JZX, 22C0759JZX, 22D0936JZX, 22D0937JZX, 22D0938JZX, 22D0939JZX, 22D0940JZX, 22D0941JZX, 22D0942JZX, 22D0943JZX, 22D0944JZX, 22D0945JZX, 22D0946JZX, 22D0947JZX, 22D0948JZX, 22D0951JZX, 22D0952JZX, 22D0953JZX, 22D0954JZX, 22D0955JZX, 22D0956JZX, 22D0957JZX, 22D0958JZX, 22D0960JZX, 22D0961JZX, 22D0962JZX, 22D0963JZX, 22D0964JZX, 22D0965JZX, 22D0966JZX, 22D0967JZX, 22D0968JZX, 22D0969JZX, 22D0970JZX, 22D0971JZX, 22D0972JZX, 22G0403JZX, 22G0404JZX, 22G0405JZX, 22G0406JZX, 22G0407JZX, 22G0408JZX, 22G0414JZX, 22G0420JZX, 22H0872JZX, 22H0873JZX, 22H0874JZX, 22H0875JZX, 22H0877JZX, 22H0881JZX, 22H0886JZX, 22H0887JZX, 22J0084JZX, 22J0085JZX, 22J0086JZX, 22J0093JZX, 22J0094JZX, 22J0095JZX, 22J0096JZX, 22J0097JZX, 22J0098JZX, 22J0099JZX, 22J0100JZX, 22J0101JZX, 22J0102JZX, 22J0103JZX, 22J0104JZX, 22J0187JZX, 22J0189JZX, 22J0190JZX, 22J0191JZX, 22J0192JZX, 22J0193JZX, 22J0194JZX, 22J0196JZX, 22J0197JZX, 22J0198JZX, 22J0206JZX, 22K0244JZX, 22K1921JZX, 22K1922JZX, 22K1983JZX; 6CN75R/10884521205437/202110171X, 202110231X, 202110310X, 202110387X, 202111268X, 202203160X, 202203304X, 202203305X, 202203306X, 202203406X, 202204167X, 202206076X, 202206077X, 202206078X, 202207117X, 202207266X, 202207267X, 202207268X, 202208201X, 202208202X, 202209220X, 202210429X, 21G0479JZX, 21G0480JZX, 21G0481JZX, 21G0482JZX, 21L0522JZX, 21L0523JZX, 21L0524JZX, 21L0526JZX; 6UN75H/20884521176253/21C0420JZX, 21C0421JZX, 21C0422JZX, 21C0423JZX, 21C0424JZX, 21C0425JZX, 21C0426JZX, 21C0427JZX, 21C0428JZX, 21C0429JZX, 21C0430JZX, 21E1758JZX, 21E1759JZX, 21E1760JZX, 21E1761JZX, 21E1762JZX, 21E1763JZX, 21E1764JZX, 21E1765JZX, 21E1766JZX, 21E1767JZX, 21F0225JZX, 21G1158JZX, 21G1160JZX, 21G1161JZX, 21G1162JZX, 21G1163JZX, 21H0749JZX, 21H1009JZX, 21I0022JZX, 22B1035JZX, 22B1036JZX, 22B1039JZX, 22B1040JZX, 22B1041JZX, 22B1042JZX, 22B1043JZX, 22B1044JZX, 22B1045JZX, 22B1046JZX, 22B1047JZX, 22B1048JZX, 22B1049JZX, 22B1052JZX, 22B1053JZX, 22B1054JZX, 22C1511JZX, 22C1512JZX, 22C1513JZX, 22C1514JZX, 22C1515JZX, 22C1516JZX, 22C1518JZX, 22C1519JZX, 22D0792JZX, 22D0793JZX, 22D0794JZX, 22D0795JZX, 22D0796JZX, 22D0902JZX, 22D0903JZX, 22D0904JZX, 22D0905JZX, 22D0906JZX, 22D0907JZX, 22D0908JZX, 22D0913JZX, 22D0914JZX, 22D0915JZX, 22D0916JZX, 22D0919JZX, 22D0920JZX, 22D0921JZX, 22D0922JZX, 22E1002JZX, 22E1003JZX, 22E1004JZX, 22E1005JZX, 22E1008JZX, 22E1009JZX, 22E1010JZX, 22E1012JZX, 22E1013JZX, 22E1014JZX, 22E1015JZX, 22E1016JZX, 22E1017JZX, 22K0254JZX, 22K0255JZX, 22K0256JZX, 22K0257JZX, 22K0258JZX, 22K0259JZX, 22K0260JZX, 22K0261JZX, 22K0262JZX, 22K0263JZX, 22K0264JZX, 22K0265JZX, 22K0267JZX, 22K0268JZX, 22K0269JZX, 22K0270JZX, 22K0271JZX, 22K0272JZX, 22K0273JZX, 22K0274JZX, 22K0275JZX; 6UN75R/10884521205505/202107305X, 202107306X, 202108126X, 202108178X, 202108179X, 202109214X, 202109215X, 202109340X, 202111221X, 202111222X, 202111266X, 202201019X, 21G0487JZX, 21H0958JZX, 22B0728JZX, 22C1156JZX, 22C1157JZX, 22C1158JZX, 22C1159JZX, 22C1160JZX, 22E1001JZX, 22H0064JZX, 22H0065JZX, 22H0067JZX, 22H0068JZX, 22H0069JZX, 22H0072JZX, 22H0073JZX, 22H0075JZX, 22H0076JZX; 7CN80A/10884521172487/21C0355JZX, 22D0975JZX, 22D0976JZX, 22E0964JZX; 7CN80H/20884521176338/21C0381JZX, 21C0382JZX, 21C0383JZX, 21C0384JZX, 21D0442JZX, 21D0443JZX, 21E1698JZX, 21G0454JZX, 21I0402JZX, 21K0651JZX, 21K0652JZX, 21K0653JZX, 21K0654JZX, 21K0655JZX; 7CN80R/10884521205444/202104382X, 202104383X, 202105228X, 202105284X, 202106225X, 202203162X, 202203309X, 202204065X, 202206207X, 202206208X, 202206209X, 202206297X, 21G0791JZX; 7UN80A/20884521175737/20J0050JZX, 21C0356JZX, 21E1744JZX, 22C0313JZX, 22C0314JZX; 7UN80H/20884521176260/21C0431JZX, 21E1768JZX, 22C1527JZX, 22C1528JZX; 7UN80R/10884521205512/202203307X, 21G0774JZX; 8CN85A/10884521172494/21C0434JZX, 21C0435JZX, 21C0436JZX, 21I0287JZX, 21I0366JZX, 22A0136JZX, 22E1151JZX, 22E1152JZX, 22E1156JZX, 22E1157JZX; 8CN85H/20884521176345/21C0386JZX, 21C0387JZX, 21C0388JZX, 21C0389JZX, 21C0390JZX, 21C0391JZX, 21C0392JZX, 21C0393JZX, 21C0394JZX, 21C0395JZX, 21C0396JZX, 21C0397JZX, 21C0398JZX, 21C0399JZX, 21C0400JZX, 21C0401JZX, 21D0444JZX, 21D0445JZX, 21D0446JZX, 21D0447JZX, 21D0448JZX, 21D0449JZX, 21D0450JZX, 21D0451JZX, 21D0452JZX, 21D0453JZX, 21D0454JZX, 21D0455JZX, 21D0456JZX, 21D0457JZX, 21D0460JZX, 21E0623JZX, 21E0624JZX, 21E0625JZX, 21E0626JZX, 21E0627JZX, 21E0628JZX, 21E0629JZX, 21E0630JZX, 21E0631JZX, 21E1094JZX, 21E1095JZX, 21E1142JZX, 21E1143JZX, 21E1146JZX, 21E1147JZX, 21E1550JZX, 21E1551JZX, 21E1552JZX, 21E1711JZX, 21E1714JZX, 21E1715JZX, 21E1716JZX, 21E1717JZX, 21E1718JZX, 21E1719JZX, 21E1720JZX, 21E1721JZX, 21E1722JZX, 21E1723JZX, 21E1724JZX, 21E1725JZX, 21E1726JZX, 21E1727JZX, 21E1728JZX, 21E1729JZX, 21E1730JZX, 21E1731JZX, 21E1732JZX, 21E1733JZX, 21E1734JZX, 21E1735JZX, 21E1736JZX, 21E1737JZX, 21E1738JZX, 21F0524JZX, 21G0287JZX, 21G0288JZX, 21G0289JZX, 21G0290JZX, 21G0291JZX, 21G1149JZX, 21G1150JZX, 21H0796JZX, 21H0797JZX, 21I0330JZX, 21I0331JZX, 21I0332JZX, 21I0333JZX, 21I0334JZX, 21I0367JZX, 21I0368JZX, 21J0186JZX, 21J0187JZX, 21J0188JZX, 21J0189JZX, 21J0190JZX, 21J0191JZX, 21J0193JZX, 21L0333JZX, 22A0694JZX, 22A0695JZX, 22A0779JZX, 22A0780JZX, 22A0781JZX, 22A0782JZX, 22A0783JZX, 22A0784JZX, 22A0785JZX, 22A0786JZX, 22A0787JZX, 22A0788JZX, 22A0820JZX, 22A0821JZX, 22A0822JZX, 22A0823JZX, 22A0824JZX, 22A0825JZX, 22A0826JZX, 22A0827JZX, 22A0828JZX, 22A0829JZX, 22A0830JZX, 22A0831JZX, 22A0832JZX, 22A0833JZX, 22A0835JZX, 22A0836JZX, 22C0795JZX, 22C0797JZX, 22C0798JZX, 22C0799JZX ,22C0800JZX ,22C0802JZX 22C0813JZX, 22C0814JZX, 22C0815JZX, 22C0819JZX, 22C0820JZX, 22E1231JZX, 22E1232JZX, 22E1233JZX, 22E1234JZX, 22E1235JZX, 22E1236JZX, 22E1237JZX, 22E1238JZX, 22E1239JZX, 22E1240JZX, 22E1243JZX, 22E1244JZX, 22E1245JZX, 22E1246JZX, 22E1247JZX, 22E1248JZX, 22E1249JZX, 22E1260JZX, 22E1261JZX, 22E1262JZX, 22E1263JZX, 22E1264JZX, 22E1265JZX, 22E1266JZX, 22E1267JZX, 22F0476JZX, 22F0477JZX, 22F0478JZX, 22F0479JZX, 22F0480JZX, 22F0481JZX, 22F0482JZX, 22F0483JZX, 22F0484JZX, 22F0485JZX, 22F0486JZX, 22F0487JZX, 22F0488JZX, 22F0489JZX, 22F0490JZX, 22F0491JZX, 22F0492JZX, 22F0493JZX, 22F0494JZX, 22F0495JZX, 22F0496JZX, 22F0497JZX, 22F0498JZX, 22F0499JZX, 22F0500JZX, 22F0501JZX, 22F0502JZX, 22F0503JZX, 22F0504JZX, 22F0505JZX, 22F0506JZX, 22F0862JZX, 22F0863JZX, 22F0864JZX, 22F0897JZX, 22F0898JZX, 22F0899JZX, 22F0900JZX, 22I0530JZX, 22I0531JZX, 22I0532JZX, 22I0533JZX, 22I0534JZX, 22J0120JZX, 22J0121JZX, 22J0122JZX, 22J0123JZX, 22J0124JZX, 22J0125JZX, 22J0126JZX, 22J0127JZX, 22J0128JZX, 22J0129JZX, 22J0130JZX, 22J0131JZX, 22J0132JZX, 22J1084JZX, 22J1085JZX, 22J1096JZX, 22J1097JZX, 22J1098JZX, 22J1099JZX, 22J1100JZX, 22J1101JZX, 22J1102JZX, 22J1103JZX, 22J1104JZX, 22J1105JZX, 22J1107JZX, 22J1108JZX, 22J1109JZX, 22J1115JZX; 8CN85R/10884521205451/202106382X, 202107194X, 202107195X, 202107196X, 202109037X, 202109124X, 202109168X, 202110189X, 202111087X, 202111106X, 202111115X, 202201302X, 202202117X, 202202118X, 202202120X, 202202323X, 202202438X, 202204062X, 21H0762JZX, 21H0763JZX, 21H0764JZX, 21H0795JZX, 21H0951JZX, 21H0952JZX, 21J0382JZX, 21J0383JZX, 22A0732JZX, 22F0889JZX, 22H0040JZX; 8UN85A/20884521175744/22C1488JZX; 8UN85H/20884521176277/21C0408JZX, 21C0409JZX, 21C0410JZX, 21E0653JZX, 21E0654JZX, 21E1745JZX, 21E1746JZX, 21E1747JZX, 21E1748JZX, 21E1749JZX, 21E1750JZX, 21E1751JZX, 21E1752JZX, 21E1753JZX, 21F0523JZX, 21H0479JZX, 21L0296JZX, 21L0297JZX, 21L0298JZX, 21L0299JZX, 21L0300JZX, 21L0301JZX, 21L0302JZX, 21L0303JZX, 21L0304JZX, 21L0305JZX, 21L0380JZX, 22B1059JZX, 22B1060JZX, 22C1484JZX, 22C1485JZX; 8UN85R/10884521205529/202105199X, 202105200X, 202105201X, 202105383X, 202105384X, 202105385X, 202106264X, 202106265X, 202106313X, 202106314X, 202106315X, 202106316X, 202107063X, 202107064X, 202107065X, 202107113X, 202107114X, 202107115X, 202107116X, 202107202X, 202108121X, 202108187X, 202108188X, 202108230X, 202202326X, 202203066X, 202203409X, 202208021X, 202209336X, 21G0922JZX, 22C1430JZX, 22C1432JZX; 9CN90H/20884521176352/21C0411JZX, 21C0412JZX, 21D0464JZX, 21D0465JZX, 21D0601JZX, 21I0555JZX, 21I0556JZX, 21K0262JZX, 21K0263JZX; 9CN90R/10884521205468/202105281X, 202110114X, 202202327X, 202204166X, 202205514X; 9UN90H/20884521176284/21E0657JZX, 21E0658JZX; 9UN90R/10884521205536/202203308X, 21F0328JZX

Distribution

Worldwide - US Nationwide distribution including in the states of PR, CA, FL, MN, PA, AZ, IL, TN, OH, MI, GA, KS, HI, MD, AK, NY, NE, IA, OK, VA, ND, MO, NC, SC, KY, OR, IN, WA, TX, WI, CT, MA, WV, CO, VT, NJ, SD, AR, AL, MS, LA, DE, MT, WY, DC, UT, NV, NH, RI, ID, NM, ME and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bolivia, Plurinational State Of Bosnia And Herzegovina, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Jordan, Kosovo, Kuwait, Lebanon, Lithuania, Luxembourg, Macao, Malaysia, Malta, Martinique, Mexico, Namibia, Netherlands, New Caledonia, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Republic Of Tanzania, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, Viet Nam.

Quantity

599,187