FDA Enforcement
Class II
Terminated
Zimmer Sterile Disposable Tourniquet Cuff with Protective Sleeve and PLC, Dual Port, Dual Bladder, Item No. 60708015200, UDI (01)00889024377547 (17)230812(10)27968701 The product is a tourniquet - Product Usage: Tourniquets are intended to be used by qualified medical professionals to exert enough pressure on the arterial blood flow in a limb to produce a bloodless operating field.
Recall: Z-1356-2020
·
Reported March 4, 2020
Enforcement
- Recall Number
- Z-1356-2020
- Event ID
- 84752
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer Surgical Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 4, 2020
- Initiation Date
- January 3, 2020
- Classification Date
- February 25, 2020
- Termination Date
- October 30, 2020
- Address
- 200 W Ohio Ave, N/A, Dover, OH, 44622-9642, United States
Description
Zimmer Sterile Disposable Tourniquet Cuff with Protective Sleeve and PLC, Dual Port, Dual Bladder, Item No. 60708015200, UDI (01)00889024377547 (17)230812(10)27968701 The product is a tourniquet - Product Usage: Tourniquets are intended to be used by qualified medical professionals to exert enough pressure on the arterial blood flow in a limb to produce a bloodless operating field.
Reason
When inflating/deflating a single bladder, both bladders could potentially inflate/deflate due to a leak between the bladders.
Code Info
Lot 27968701
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, FL, IA, ID, IL, LA, MA, MO, NC, NY, OH, OR, PA, SC, TN, TX, WA, WI, and WV and country of Netherlands.
Quantity
58