FDA Enforcement Class II Terminated

Medtronic Dual Chamber Temporary External Pacemaker, Model 5392. Product Usage: The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic, or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient. Specific indications for temporary cardiac pacing include, but are not limited to, the following: - Complete heart block - Sinus bradycardia - Sick sinus syndrome - Bradycardia with congestive heart failure - Atrial and/or ventricular arrhythmias - Cardiac arrest - Support, management, and evaluation of a patient before permanent pacemaker implantation - Support during permanent pacemaker replacement - Cardiac complications during invasive or surgical procedures - Support following cardiac surgery - Acute myocardial infarction complicated by heart block - Atrial tachyarrhythmias that require high-rate burst pacing for treatment

Recall: Z-1356-2015 · Reported April 8, 2015

Enforcement

Recall Number
Z-1356-2015
Event ID
70763
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Medtronic Inc. Cardiac Rhythm Disease Management
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 8, 2015
Initiation Date
March 17, 2015
Classification Date
March 31, 2015
Termination Date
October 8, 2015
Address
8200 Coral Sea St NE, N/A, Saint Paul, MN, 55112-4391, United States

Description

Medtronic Dual Chamber Temporary External Pacemaker, Model 5392. Product Usage: The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual chamber pacing in a clinical environment by trained personnel. The temporary pacemaker can be used where short-term demand (synchronous) or asynchronous pacing is indicated for therapeutic, prophylactic, or diagnostic purposes. The temporary pacemaker must be used in an environment where the patient is monitored continuously to ensure that it is operating properly and delivering appropriate therapy to the patient. Specific indications for temporary cardiac pacing include, but are not limited to, the following: - Complete heart block - Sinus bradycardia - Sick sinus syndrome - Bradycardia with congestive heart failure - Atrial and/or ventricular arrhythmias - Cardiac arrest - Support, management, and evaluation of a patient before permanent pacemaker implantation - Support during permanent pacemaker replacement - Cardiac complications during invasive or surgical procedures - Support following cardiac surgery - Acute myocardial infarction complicated by heart block - Atrial tachyarrhythmias that require high-rate burst pacing for treatment

Reason

Possible performance issue when used with specific AA-sized (LR6) batteries. The negative terminal of a commercially available AA (LR6) battery did not maintain a sufficient connection with the battery drawer electrical contact. This issue could prevent the EPG from powering on or cause the EPG to abruptly lose primary battery power, potentially stopping delivery of pacing therapy.

Code Info

serial numbers equal or lower than DJH009999P

Distribution

Worldwide Distribution - US Nationwide in the states of: AL, AK, AZ, AR, CA, CO, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MA, MI, MM, MS, MO, MT, NE NV,N, NM, NY, NC, ND, OH,OK, PA, RI, SC, TN, TX, UT, VT, VA, WA, WV, WI, WY, PR, VI and the countries of:Australia, Austria, Bahamas, Belgium, Bolivia ,Botswana, Brazil ,Brunei, Darussalam, Bulgaria, Burkina Faso, Canada, Colombia, Croatia, Cyprus, Czech Republic, Denmark,, Dominican Republic, Egypt, France, Georgia, Germany, Greece, Guyana, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Korea, Kuwait, Lebanon, Luxembourg, Macedonia, Malaysia, Mauritius, Morocco, Mozambique, Namibia, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Papua, New Guinea, Peru, Philippines, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, St. Pierre and Miquelon,Sweden, Switzerland, Tanzania, Thailand, Trinidad and Tobago, Tunisia, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, and Yemen. .

Quantity

6040 devices