FDA Enforcement Class II Ongoing

thermage THERMAGE CPT SYSTEM REF TG-2B *** Solta Medical, Inc. thermage RF GENERATOR REF TG-2B *** Solta Medical, Inc. Dermatologic and general surgical procedures for electrocoagulation and hemostasis; " Non-invasive treatment of periorbital wrinkles and rhytids including upper and lower eyelids; " Non-invasive treatment of wrinkles and rhytids. The Thermage CPT System ( System ) consists of the RF Generator/ Cooling System, 16.00 cm2 Handpiece ( Body Handpiece ), Standard Handpiece with Vibration ( Face Handpiece ), optional electric Footswitch, accessory cables, Return Pad, Return Pad Cable, and Treatment Tips.

Recall: Z-1354-2023 · Reported April 19, 2023

Enforcement

Recall Number
Z-1354-2023
Event ID
91847
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Solta Medical Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 19, 2023
Initiation Date
February 16, 2023
Classification Date
April 7, 2023
Address
11720 N Creek Pkwy N Ste 100, N/A, Bothell, WA, 98011-8244, United States

Description

thermage THERMAGE CPT SYSTEM REF TG-2B *** Solta Medical, Inc. thermage RF GENERATOR REF TG-2B *** Solta Medical, Inc. Dermatologic and general surgical procedures for electrocoagulation and hemostasis; " Non-invasive treatment of periorbital wrinkles and rhytids including upper and lower eyelids; " Non-invasive treatment of wrinkles and rhytids. The Thermage CPT System ( System ) consists of the RF Generator/ Cooling System, 16.00 cm2 Handpiece ( Body Handpiece ), Standard Handpiece with Vibration ( Face Handpiece ), optional electric Footswitch, accessory cables, Return Pad, Return Pad Cable, and Treatment Tips.

Reason

A service unit was potentially improperly tested and calibrated during manufacturing on a failed test station.

Code Info

Model Number: TG-2B UDI-DI Code: 00850608002124 Serial Number: 001933

Distribution

U.S. Distribution to state of: MD

Quantity

1 unit