Abbott ARCHITECT Estradiol Reagent Kit - Product Usage: The ARCHITECT Estradiol assay is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of Estradiol in human serum and plasma.
Enforcement
- Recall Number
- Z-1353-2019
- Event ID
- 82129
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Abbott Ireland Diagnostics Division
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 22, 2019
- Initiation Date
- February 5, 2019
- Classification Date
- May 16, 2019
- Termination Date
- August 21, 2020
- Address
- Lisnamuck, Co. Longford, N/A, Ireland
Description
Abbott ARCHITECT Estradiol Reagent Kit - Product Usage: The ARCHITECT Estradiol assay is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of Estradiol in human serum and plasma.
Patient results may be falsely elevated. This patient impact only applies to patients currently being treated with or recently treated with the drug Mifepristone.
List Number 7K72-20 - ALL LOTS, List Number 7K72-25 - ALL LOTS, and List number 7K72-35 - ALL LOTS.
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR. AZ, CA, CT, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, KS, MT, NC, ND, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, WI, AND PUERTO RICO, and countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Belarus, Belgium, Bosnia and Herzegovi, Brazil, Brunei, Botswana, Canada, Chile, China, Columbia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Gaza and Jericho, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland,Israel, Italy, Jamaica, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Mauritius, Mexico, Moldova, Montenegro, Morocco, Mozambique, Namibia, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland,Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turks, UAE, Uganda, United Kingdom, and Vietnam.
2,724