FDA Enforcement
Class II
Ongoing
Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad)
Recall: Z-1350-2023
·
Reported April 19, 2023
Enforcement
- Recall Number
- Z-1350-2023
- Event ID
- 91758
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Church & Dwight Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- April 19, 2023
- Initiation Date
- January 25, 2023
- Classification Date
- April 7, 2023
- Address
- 469 N Harrison St, N/A, Princeton, NJ, 08540-3510, United States
Description
Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad)
Reason
Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user.
Code Info
All lots with the Easy Read App
Distribution
US Nationwide distribution.
Quantity
780000 downloads