FDA Enforcement Class II Terminated

Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw

Recall: Z-1350-2020 · Reported March 4, 2020

Enforcement

Recall Number
Z-1350-2020
Event ID
84638
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ulrich medical USA Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 4, 2020
Initiation Date
December 31, 2019
Classification Date
February 24, 2020
Termination Date
November 17, 2025
Address
18221 Edison Ave, N/A, Chesterfield, MO, 63005-3703, United States

Description

Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw

Reason

Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.

Code Info

Lot Numbers: U010577, U016620, U019084

Distribution

AZ, TX, GA, FL, AR, CA

Quantity

179 units