FDA Enforcement Class II Ongoing

NUVASIVE PRECICE Intramedullary Limb Lengthening (IMLL) and (Short) System End Cap for PRECICE Intramedullary Limb Lengthening System Implant; Packaged End Cap 8.5/10.7mm,¿ +0mm; Packaged End Cap 8.5/10.7mm,¿ +5mm; Packaged End Cap 8.5/10.7mm,¿ +10mm; Packaged End Cap 8.5/10.7mm,¿ +15mm; Packaged End Cap 8.5/10.7mm,¿ +20mm; Packaged End Cap 12.5mm,¿ +0mm; Packaged End Cap 12.5mm,¿ +5mm; Packaged End Cap 12.5mm,¿ +10mm; Packaged End Cap 12.5mm,¿ +15mm; Packaged End Cap 12.5mm,¿ +20mm; 3.5 mm Locking Screw, 20 mm Length; 3.5 mm Locking Screw 25 mm Length; 3.5 mm Locking Screw 30 mm Length; 3.5 mm Locking Screw 35 mm Length; 3.5 mm Locking Screw 40 mm Length; 3.5 mm Locking Screw 45 mm Length; 3.5 mm Locking Screw 50 mm Length; 3.5 mm Locking Screw 55 mm Length; 3.5 mm Locking Screw 60 mm Length; Locking Screw, 3.5x65mm; Locking Screw, 3.5x70mm; Locking Screw, 3.5x75mm; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 4.0 mm X 65 mm Locking Screw, Peg; 4.0 mm X 70 mm Locking Screw, Peg; 4.0 mm X 75 mm Locking Screw, Peg; 4.0 mm X 80 mm Locking Screw, Peg; 4.0 mm X 85 mm Locking Screw, Peg; 4.0 mm X 90 mm Locking Screw, Peg; 4.0 mm X 95 mm Locking Screw, Peg; 4.0 mm X 100 mm Locking Screw, Peg; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 5.0 mm X 80 mm Locking Screw, Peg; 5.0 mm X 85 mm Locking Screw, Peg; 5.0 mm X 90 mm Locking Screw, Peg; 5.0 mm X 95 mm Locking Screw, Peg; 5.0 mm X 100 mm Locking Screw, Peg; Orthopedic intramedullary limb lengthening rod, retrograde femur 10 degree, 10.7 mm diameter, 170 mm length; Orthopedic intramedullary limb lengthening rod, antegrade tibia, 10.7 mm diameter, 160 mm length; Orthopedic intramedullary limb lengthening rod, antegrade femur trochanteric, 10.7 mm diameter, 170 mm length; Orthopedic intramedullary limb lengthening rod, universal femur, 10.7 mm diameter, 170 mm length; Orthopedic intramedullary limb lengthening rod, retrograde femur, 10 degree, 10.7 mm diameter, 215 mm length; Orthopedic intramedullary limb lengthening rod, retrograde femur, 10 degree, 10.7 mm diameter, 230 mm length; Orthopedic intramedullary limb lengthening rod, antegrade femur (piriformis), 10.7 mm diameter, 215 mm length; Orthopedic intramedullary limb lengthening rod, antegrade femur (piriformis), 10.7 mm diameter, 230 mm length; Orthopedic intramedullary limb lengthening rod, antegrade tibia, 10.7 mm diameter, 195 mm length; Orthopedic intramedullary limb lengthening rod, antegrade tibia, 10.7 mm diameter, 215 mm length; Orthopedic intramedullary limb lengthening

Recall: Z-1349-2021 · Reported April 14, 2021

Enforcement

Recall Number
Z-1349-2021
Event ID
87353
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Nuvasive Specialized Orthopedics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 14, 2021
Initiation Date
February 12, 2021
Classification Date
April 5, 2021
Address
101 Enterprise Ste 100, N/A, Aliso Viejo, CA, 92656-2604, United States

Description

NUVASIVE PRECICE Intramedullary Limb Lengthening (IMLL) and (Short) System End Cap for PRECICE Intramedullary Limb Lengthening System Implant; Packaged End Cap 8.5/10.7mm,¿ +0mm; Packaged End Cap 8.5/10.7mm,¿ +5mm; Packaged End Cap 8.5/10.7mm,¿ +10mm; Packaged End Cap 8.5/10.7mm,¿ +15mm; Packaged End Cap 8.5/10.7mm,¿ +20mm; Packaged End Cap 12.5mm,¿ +0mm; Packaged End Cap 12.5mm,¿ +5mm; Packaged End Cap 12.5mm,¿ +10mm; Packaged End Cap 12.5mm,¿ +15mm; Packaged End Cap 12.5mm,¿ +20mm; 3.5 mm Locking Screw, 20 mm Length; 3.5 mm Locking Screw 25 mm Length; 3.5 mm Locking Screw 30 mm Length; 3.5 mm Locking Screw 35 mm Length; 3.5 mm Locking Screw 40 mm Length; 3.5 mm Locking Screw 45 mm Length; 3.5 mm Locking Screw 50 mm Length; 3.5 mm Locking Screw 55 mm Length; 3.5 mm Locking Screw 60 mm Length; Locking Screw, 3.5x65mm; Locking Screw, 3.5x70mm; Locking Screw, 3.5x75mm; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 4.0 mm Locking Screw 20 mm Length; 4.0 mm Locking Screw 25 mm Length; 4.0 mm Locking Screw 30 mm Length; 4.0 mm Locking Screw 35 mm Length; 4.0 mm Locking Screw 40 mm Length; 4.0 mm Locking Screw 45 mm Length; 4.0 mm Locking Screw 50 mm Length; 4.0 mm Locking Screw 55 mm Length; 4.0 mm Locking Screw 60 mm Length; 4.0 mm X 65 mm Locking Screw, Peg; 4.0 mm X 70 mm Locking Screw, Peg; 4.0 mm X 75 mm Locking Screw, Peg; 4.0 mm X 80 mm Locking Screw, Peg; 4.0 mm X 85 mm Locking Screw, Peg; 4.0 mm X 90 mm Locking Screw, Peg; 4.0 mm X 95 mm Locking Screw, Peg; 4.0 mm X 100 mm Locking Screw, Peg; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 5.0 mm X 80 mm Locking Screw, Peg; 5.0 mm X 85 mm Locking Screw, Peg; 5.0 mm X 90 mm Locking Screw, Peg; 5.0 mm X 95 mm Locking Screw, Peg; 5.0 mm X 100 mm Locking Screw, Peg; Orthopedic intramedullary limb lengthening rod, retrograde femur 10 degree, 10.7 mm diameter, 170 mm length; Orthopedic intramedullary limb lengthening rod, antegrade tibia, 10.7 mm diameter, 160 mm length; Orthopedic intramedullary limb lengthening rod, antegrade femur trochanteric, 10.7 mm diameter, 170 mm length; Orthopedic intramedullary limb lengthening rod, universal femur, 10.7 mm diameter, 170 mm length; Orthopedic intramedullary limb lengthening rod, retrograde femur, 10 degree, 10.7 mm diameter, 215 mm length; Orthopedic intramedullary limb lengthening rod, retrograde femur, 10 degree, 10.7 mm diameter, 230 mm length; Orthopedic intramedullary limb lengthening rod, antegrade femur (piriformis), 10.7 mm diameter, 215 mm length; Orthopedic intramedullary limb lengthening rod, antegrade femur (piriformis), 10.7 mm diameter, 230 mm length; Orthopedic intramedullary limb lengthening rod, antegrade tibia, 10.7 mm diameter, 195 mm length; Orthopedic intramedullary limb lengthening rod, antegrade tibia, 10.7 mm diameter, 215 mm length; Orthopedic intramedullary limb lengthening

Reason

Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.

Code Info

All Lots manufactured between April 2015 to February 2021. PRECICE Intramedullary Limb Lengthening (IMLL) Product Codes: End caps: CPA1-000; CPA2-000; CPA2-005; CPA2-010; CPA2-015; CPA2-020; CPA3-000; CPA3-005; CPA3-010; CPA3-015; CPA3-020; Locking screws: LSB3-020; LSB3-025; LSB3-030; LSB3-035; LSB3-040; LSB3-045; LSB3-050; LSB3-055; LSB3-060; LSB3-065; LSB3-070; LSB3-075; LSB4-020; LSB4-025; LSB4-030; LSB4-035; LSB4-040; LSB4-045; LSB4-050; LSB4-055; LSB4-060; LSB5-020; LSB5-025; LSB5-030; LSB5-035; LSB5-040; LSB5-045; LSB5-050; LSB5-055; LSB5-060; LSB5-065; LSB5-070; LSB5-075; LSC4-020; LSC4-025; LSC4-030; LSC4-035; LSC4-040; LSC4-045; LSC4-050; LSC4-055; LSC4-060; LSC4-065; LSC4-070; LSC4-075; LSC4-080; LSC4-085; LSC4-090; LSC4-095; LSC4-100; LSC5-020; LSC5-025; LSC5-030; LSC5-035; LSC5-040; LSC5-045; LSC5-050; LSC5-055; LSC5-060; LSC5-065; LSC5-070; LSC5-075; LSC5-080; LSC5-085; LSC5-090; LSC5-095; LSC5-100; Limb lengthening intramedullary rods 10.7 mm diameter: P10.7-30H170; P10.7-30J160; P10.7-30K170; P10.7-30U170; P10.7-50A215; P10.7-50A230; P10.7-50B215; P10.7-50B230; P10.7-50C195; P10.7-50C215; P10.7-50C230; P10.7-50D215; P10.7-50D230; P10.7-50E215; P10.7-50E230; P10.7-50H190; P10.7-50J180; P10.7-50K190; P10.7-50M175; P10.7-50N160; P10.7-50P175; P10.7-50Q155; P10.7-50U190; P10.7-50V215; P10.7-50V230; P10.7-50X215; P10.7-50X230; P10.7-80A245; P10.7-80A275; P10.7-80A305; P10.7-80A335; P10.7-80A365; P10.7-80B245; P10.7-80B275; P10.7-80B305; P10.7-80B335; P10.7-80B365; P10.7-80C245; P10.7-80C275; P10.7-80C305; P10.7-80C335; P10.7-80C365; P10.7-80D245; P10.7-80D275; P10.7-80D305; P10.7-80D335; P10.7-80D365; P10.7-80E245; P10.7-80E275; P10.7-80E305; P10.7-80E335; P10.7-80E365; P10.7-80V245; P10.7-80V275; P10.7-80V305; P10.7-80V335; P10.7-80V365; P10.7-80X245; P10.7-80X275; P10.7-80X305; P10.7-80X335; P10.7-80X365; Limb lengthening intramedullary rods 12.5 mm diameter: P12.5-30H170; P12.5-30J160; P12.5-30K170; P12.5-30U170; P12.5-50A215; P12.5-50A230; P12.5-50B215; P12.5-50B230; P12.5-50C195; P12.5-50C215; P12.5-50C230; P12.5-50D215; P12.5-50D230; P12.5-50E215; P12.5-50E230; P12.5-50H190; P12.5-50J180; P12.5-50K190; P12.5-50U190; P12.5-50V215; P12.5-50V230; P12.5-50X215; P12.5-50X230; P12.5-80A245; P12.5-80A275; P12.5-80A305; P12.5-80A335; P12.5-80A365; P12.5-80B245; P12.5-80B275; P12.5-80B305; P12.5-80B335; P12.5-80B365; P12.5-80C245; P12.5-80C275; P12.5-80C305; P12.5-80C335; P12.5-80C365; P12.5-80D245; P12.5-80D275; P12.5-80D305; P12.5-80D335; P12.5-80D365; P12.5-80E245; P12.5-80E275; P12.5-80E305; P12.5-80E335; P12.5-80E365; P12.5-80V245; P12.5-80V275; P12.5-80V305; P12.5-80V335; P12.5-80V365; P12.5-80X245; P12.5-80X275; P12.5-80X305; P12.5-80X335; P12.5-80X365; Limb lengthening intramedullary rods 8.5 mm diameter: P8.5-30H170; P8.5-30J160; P8.5-30K170; P8.5-30U170; P8.5-50A215; P8.5-50A230; P8.5-50B215; P8.5-50B230; P8.5-50C195; P8.5-50C215; P8.5-50C230; P8.5-50D215; P8.5-50D230; P8.5-50E215; P8.5-50E230; P8.5-50H190; P8.5-50J180; P8.5-50K190; P8.5-50M165; P8.5-50N150 P8.5-50P165; P8.5-50Q155; P8.5-50U190; P8.5-50V215; P8.5-50V230; P8.5-50X215; P8.5-50X230; P8.5-80A245; P8.5-80A275; P8.5-80A305; P8.5-80A335; P8.5-80A365; P8.5-80B245; P8.5-80B275; P8.5-80B305; P8.5-80B335; P8.5-80B365; P8.5-80C245; P8.5-80C275; P8.5-80C305; P8.5-80C335; P8.5-80C365; P8.5-80D245; P8.5-80D275; P8.5-80D305; P8.5-80D335; P8.5-80D365; P8.5-80E245; P8.5-80E275; P8.5-80E305; P8.5-80E335; P8.5-80E365; P8.5-80V245; P8.5-80V275; P8.5-80V305; P8.5-80V335; P8.5-80V365; P8.5-80X245; P8.5-80X275; P8.5-80X305; P8.5-80X335; P8.5-80X365; Limb lengthening intramedullary rods 10.5 mm diameter: PR10.5-50A215; PR10.5-50A230; PR10.5-50B215; PR10.5-50B230; PR10.5-50C195; PR10.5-50C215; PR10.5-50C230; PR10.5-50D215; PR10.5-50D230; PR10.5-50E215; PR10.5-50E230; PR10.5-80A245; PR10.5-80A275; PR10.5-80A305; PR10.5-80A335; PR10.5-80A365; PR10.5-80B245; PR10.5-80B275; PR10.5-80B305; PR10.5-80B335; PR10.5-80B365; PR10.5-80C245; PR10.5-80C275; PR10.5-80C305; PR10.5-80C335; PR10.5-80C365; PR10.5-80D245; PR10.5-80D275; PR10.5-80D305; PR10.5-80D335; PR10.5-80D365; PR10.5-80E245; PR10.5-80E275; PR10.5-80E305; PR10.5-80E335; PR10.5-80E365; Limb lengthening intramedullary rods 11.5 mm diameter: PR11.5-50A215; PR11.5-50A230; PR11.5-50B215; PR11.5-50B230; PR11.5-50C195; PR11.5-50C215; PR11.5-50C230; PR11.5-50D215; PR11.5-50D230; PR11.5-50E215; PR11.5-50E230; PR11.5-80A245; PR11.5-80A275; PR11.5-80A305; PR11.5-80A335; PR11.5-80A365; PR11.5-80B245; PR11.5-80B275; PR11.5-80B305; PR11.5-80B335; PR11.5-80B365; PR11.5-80C245; PR11.5-80C275; PR11.5-80C305; PR11.5-80C335; PR11.5-80C365; PR11.5-80D245; PR11.5-80D275; PR11.5-80D305; PR11.5-80D335; PR11.5-80D365; PR11.5-80E245; PR11.5-80E275; PR11.5-80E305; PR11.5-80E335; PR11.5-80E365; Limb lengthening intramedullary rods 12.5 mm diameter: PR12.5-50A215; PR12.5-50A230; PR12.5-50B215; PR12.5-50B230; PR12.5-50C195; PR12.5-50C215; PR12.5-50C230; PR12.5-50D215; PR12.5-50D230; PR12.5-50E215; PR12.5-50E230; PR12.5-80A245; PR12.5-80A275; PR12.5-80A305; PR12.5-80A335; PR12.5-80A365; PR12.5-80B245; PR12.5-80B275; PR12.5-80B305; PR12.5-80B335; PR12.5-80B365; PR12.5-80C245; PR12.5-80C275; PR12.5-80C305; PR12.5-80C335; PR12.5-80C365; PR12.5-80D245; PR12.5-80D275; PR12.5-80D305; PR12.5-80D335; PR12.5-80D365; PR12.5-80E245; PR12.5-80E275; PR12.5-80E305; PR12.5-80E335; PR12.5-80E365; Limb lengthening intramedullary rods 9.0 mm diameter: PR9.0-50A215; PR9.0-50A230; PR9.0-50B215; PR9.0-50B230; PR9.0-50C195; PR9.0-50C215; PR9.0-50C230; PR9.0-50D215; PR9.0-50D230; PR9.0-50E215; PR9.0-50E230; PR9.0-80A245; PR9.0-80A275; PR9.0-80A305; PR9.0-80A335; PR9.0-80A365; PR9.0-80B245; PR9.0-80B275; PR9.0-80B305; PR9.0-80B335; PR9.0-80B365; PR9.0-80C245; PR9.0-80C275; PR9.0-80C305; PR9.0-80C335; PR9.0-80C365; PR9.0-80D245; PR9.0-80D275; PR9.0-80D305; PR9.0-80D335; PR9.0-80D365; PR9.0-80E245; PR9.0-80E275; PR9.0-80E305; PR9.0-80E335; PR9.0-80E365; Limb lengthening intramedullary rods 9.5 mm diameter: PR9.5-50A215; PR9.5-50A230; PR9.5-50B215; PR9.5-50B230; PR9.5-50C195; PR9.5-50C215; PR9.5-50C230; PR9.5-50D215; PR9.5-50D230; PR9.5-50E215; PR9.5-50E230; PR9.5-80A245; PR9.5-80A275; PR9.5-80A305; PR9.5-80A335; PR9.5-80A365; PR9.5-80B245; PR9.5-80B275; PR9.5-80B305; PR9.5-80B335; PR9.5-80B365; PR9.5-80C245; PR9.5-80C275; PR9.5-80C305; PR9.5-80C335; PR9.5-80C365; PR9.5-80D245; PR9.5-80D275; PR9.5-80D305; PR9.5-80D335; PR9.5-80D365; PR9.5-80E245; PR9.5-80E275; PR9.5-80E305; PR9.5-80E335; PR9.5-80E365; Trauma intramedullary rods 10.7 mm diameter: T10.7-50A205; T10.7-50A225; T10.7-50A240; T10.7-50B205; T10.7-50B225; T10.7-50B240; T10.7-50C205; T10.7-50C225; T10.7-50C240; T10.7-50D205; T10.7-50D225; T10.7-50D240; T10.7-50E205; T10.7-50E225; T10.7-50E240; T10.7-80A255; T10.7-80A285; T10.7-80A315; T10.7-80A345; T10.7-80A375; T10.7-80B255; T10.7-80B285; T10.7-80B315; T10.7-80B345; T10.7-80B375; T10.7-80C255; T10.7-80C285; T10.7-80C315; T10.7-80C345; T10.7-80C375; T10.7-80D255; T10.7-80D285; T10.7-80D315; T10.7-80D345; T10.7-80D375; T10.7-80E255; T10.7-80E285; T10.7-80E315; T10.7-80E345; T10.7-80E375; Trauma intramedullary rods 12.5 mm diameter: T12.5-50A205; T12.5-50A225; T12.5-50A240; T12.5-50B205; T12.5-50B225; T12.5-50B240; T12.5-50C205; T12.5-50C225; T12.5-50C240; T12.5-50D205; T12.5-50D225; T12.5-50D240; T12.5-50E205; T12.5-50E225; T12.5-50E240; T12.5-80A255; T12.5-80A285; T12.5-80A315; T12.5-80A345; T12.5-80A375; T12.5-80B255; T12.5-80B285; T12.5-80B315; T12.5-80B345; T12.5-80B375; T12.5-80C255; T12.5-80C285; T12.5-80C315; T12.5-80C345; T12.5-80C375; T12.5-80D255; T12.5-80D285; T12.5-80D315; T12.5-80D345; T12.5-80D375; T12.5-80E255; T12.5-80E285; T12.5-80E315; T12.5-80E345; T12.5-80E375; Trauma intramedullary rods 8.5 mm diameter: T8.5-50A205; T8.5-50A225; T8.5-50A240; T8.5-50B205; T8.5-50B225; T8.5-50B240; T8.5-50C205; T8.5-50C225; T8.5-50C240; T8.5-50D205; T8.5-50D225; T8.5-50D240; T8.5-50E205; T8.5-50E225; T8.5-50E240; T8.5-80A255; T8.5-80A285; T8.5-80A315; T8.5-80A345; T8.5-80A375; T8.5-80B255; T8.5-80B285; T8.5-80B315; T8.5-80B345; T8.5-80B375; T8.5-80C255; T8.5-80C285; T8.5-80C315; T8.5-80C345; T8.5-80C375; T8.5-80D255; T8.5-80D285; T8.5-80D315; T8.5-80D345; T8.5-80D375; T8.5-80E255; T8.5-80E285; T8.5-80E315; T8.5-80E345; T8.5-80E375; Locking screws: TSA4.5-020; TSA4.5-025; TSA4.5-030; TSA4.5-035; TSA4.5-040; TSA4.5-045; TSA4.5-050; TSA4.5-055; TSA4.5-060; TSA4.5-065; TSA4.5-070; TSA4.5-075; TSA4-020; TSA4-022.5; TSA4-025; TSA4-027.5; TSA4-030; TSA4-032.5; TSA4-035; TSA4-037.5; TSA4-040; TSA4-042.5; TSA4-045; TSA4-047.5; TSA4-050; TSA4-055; TSA4-060; TSA4-065; TSA4-070; TSA4-075; TSA4-080; TSA4-085; TSA4-090; TSA4-095; TSA4-100; TSA4-100; TSA5-020; TSA5-020; TSA5-022.5; TSA5-022.5; TSA5-025; TSA5-025; TSA5-027.5; TSA5-027.5; TSA5-030; TSA5-030; TSA5-032.5; TSA5-032.5; TSA5-035; TSA5-035; TSA5-037.5; TSA5-037.5; TSA5-040; TSA5-040; TSA5-042.5; TSA5-042.5; TSA5-045; TSA5-045; TSA5-047.5; TSA5-047.5; TSA5-050; TSA5-050; TSA5-052.5; TSA5-055; TSA5-055; TSA5-057.5; TSA5-060; TSA5-060; TSA5-062.5; TSA5-065; TSA5-065; TSA5-067.5; TSA5-070; TSA5-070; TSA5-072.5; TSA5-075; TSA5-075; TSA5-080; TSA5-080; TSA5-085; TSA5-085; TSA5-090; TSA5-090; TSA5-095; TSA5-095; TSA5-100; TSA5-100; TSA5-105; TSA5-105; TSA5-110; TSA5-110; GTIN Numbers: 00812258020002; 00887517001139; 00887517001146; 00887517001153; 00887517001160; 00887517001177; 00887517001184; 00887517001191; 00887517001207; 00887517001214; 00887517001221; 00812258021337; 00812258021344; 00812258021351; 00812258021368; 00812258021375; 00812258021382; 00812258021399; 00812258021405; 00812258021412; 00887517936233; 00887517936240; 00887517936257; 00856719002763; 00856719002770; 00856719002787; 00856719002794; 00856719002800; 00856719002817; 00856719002824; 00856719002831; 00856719002848; 00856719002886; 00856719002893; 00856719002909; 00856719002916; 00856719002923; 00856719002930; 00856719002947; 00856719002954; 00856719002961; 00856719002978; 00856719002985; 00856719002992; 00812258021429; 00812258021436; 00812258021443; 00812258021450; 00812258021467; 00812258021474; 00812258021481; 00812258021498; 00812258021504; 00887517000958; 00887517000965; 00887517000972; 00887517000989; 00887517000996; 00887517001009; 00887517001016; 00887517001023; 00812258021511; 00812258021528; 00812258021535; 00812258021542; 00812258021559; 00812258021566; 00812258021573; 00812258021580; 00812258021597; 00812258021603; 00812258021610; 00812258021627; 00887517001030; 00887517001047; 00887517001054; 00887517001061; 00887517001078; 00812258026073; 00812258026134; 00812258026196; 00812258026257; 00812258020057; 00812258020064; 00812258020231; 00812258020248; 00812258020422; 00812258020439; 00812258020446; 00812258020620; 00812258020637; 00812258020804; 00812258020811; 00812258026080; 00812258026141; 00812258026202; 00812258028244; 00812258028268; 00812258028282; 00812258028220; 00812258026264; 00887517934819; 00887517934826; 00887517934833; 00887517934840; 00812258020071; 00812258020088; 00812258020095; 00812258020101; 00812258020118; 00812258020255; 00812258020262; 00812258020279; 00812258020286; 00812258020293; 00812258020453; 00812258020460; 00812258020477; 00812258020484; 00812258020491; 00812258020644; 00812258020651; 00812258020668; 00812258020675; 00812258020682; 00812258020828; 00812258020835; 00812258020842; 00812258020859; 00812258020866; 00887517934857; 00887517934864; 00887517934871; 00887517934888; 00887517934895; 00887517935502; 00887517935519; 00887517935526; 00887517935533; 00887517935540; 00812258026097; 00812258026158; 00812258026219; 00812258026271; 00812258020125; 00812258020132; 00812258020309; 00812258020316; 00812258020507; 00812258020514; 00812258020521; 00812258020699; 00812258020705; 00812258020873; 00812258020880; 00812258026103; 00812258026165; 00812258026226; 00812258026288; 00887517935557; 00887517935564; 00887517935571; 00887517935588; 00812258020149; 00812258020156; 00812258020163; 00812258020170; 00812258020187; 00812258020323; 00812258020330; 00812258020347; 00812258020354; 00812258020361; 00812258020538; 00812258020545; 00812258020552; 00812258020569; 00812258020576; 00812258020712; 00812258020729; 00812258020736; 00812258020743; 00812258020750; 00812258020897; 00812258020903; 00812258020910; 00812258020927; 00812258020934; 00887517935595; 00887517935601; 00887517935618; 00887517935625; 00887517935632; 00887517935649; 00887517935656; 00887517935663; 00887517935670; 00887517935687; 00812258026059; 00812258026110; 00812258026172; 00812258026233; 00812258020019; 00812258020026; 00812258020194; 00812258020200; 00812258020378; 00812258020385; 00812258020392; 00812258020583; 00812258020590; 00812258020767; 00812258020774; 00812258026066; 00812258026127; 00812258026189; 00812258028237; 00812258028251; 00812258028275; 00812258028213; 00812258026240; 00887517935694; 00887517935700; 00887517935717; 00887517935724; 00812258020033; 00812258020040; 00812258022358; 00812258022365; 00812258022372; 00812258020217; 00812258020224; 00812258022389; 00812258022396; 00812258022402; 00812258020408; 00812258020415; 00812258022419; 00812258022426; 00812258022433; 00812258020606; 00812258020613; 00812258022440; 00812258022457; 00812258022464; 00812258020781; 00812258020798; 00812258022471; 00812258022488; 00812258022495; 00887517935731; 00887517935748; 00887517935755; 00887517935762; 00887517935779; 00887517935786; 00887517935793; 00887517935809; 00887517935816; 00887517935823; 00812258025359; 00812258025366; 00812258024253; 00812258024260; 00812258024956; 00812258024963; 00812258024970; 00812258024604; 00812258024611; 00812258025700; 00812258025717; 00812258025373; 00812258025380; 00812258025397; 00812258025403; 00812258025410; 00812258024277; 00812258024284; 00812258024291; 00812258024307; 00812258024314; 00812258024987; 00812258024994; 00812258025007; 00812258025014; 00812258025021; 00812258024628; 00812258024635; 00812258024642; 00812258024659; 00812258024666; 00812258025724; 00812258025731; 00812258025748; 00812258025755; 00812258025762; 00812258025427; 00812258025434; 00812258024321; 00812258024338; 00812258025038; 00812258025045; 00812258025052; 00812258024673; 00812258024680; 00812258025779; 00812258025786; 00812258025441; 00812258025458; 00812258025465; 00812258025472; 00812258025489; 00812258024345; 00812258024352; 00812258024369; 00812258024376; 00812258024383; 00812258025069; 00812258025076; 00812258025083; 00812258025090; 00812258025106; 00812258024697; 00812258024703; 00812258024710; 00812258024727; 00812258024734; 00812258025793; 00812258025809; 00812258025816; 00812258025823; 00812258025830; 00812258025632; 00812258025649; 00812258024390; 00812258024406; 00812258025113; 00812258025120; 00812258025137; 00812258024741; 00812258024758; 00812258025984; 00812258025991; 00812258025656; 00812258025663; 00812258025670; 00812258025687; 00812258025694; 00812258024413; 00812258024420; 00812258024437; 00812258024444; 00812258024451; 00812258025144; 00812258025151; 00812258025168; 00812258025175; 00812258025182; 00812258024765; 00812258024772; 00812258024789; 00812258024796; 00812258024802; 00812258026004; 00812258026011; 00812258026028; 00812258026035; 00812258026042; 00812258025496; 00812258025502; 00812258024468; 00812258024475; 00812258025199; 00812258025205; 00812258025212; 00812258024819; 00812258024826; 00812258025847; 00812258025854; 00812258025519; 00812258025526; 00812258025533; 00812258025540; 00812258025557; 00812258024482; 00812258024499; 00812258024505; 00812258024512; 00812258024529; 00812258025229; 00812258025236; 00812258025243; 00812258025250; 00812258025267; 00812258024833; 00812258024840; 00812258024857; 00812258024864; 00812258024871; 00812258025861; 00812258025878; 00812258025885; 00812258025892; 00812258025908; 00812258025564; 00812258025571; 00812258024536; 00812258024543; 00812258025274; 00812258025281; 00812258025298; 00812258024888; 00812258024895; 00812258025915; 00812258025922; 00812258025588; 00812258025595; 00812258025601; 00812258025618; 00812258025625; 00812258024550; 00812258024567; 00812258024574; 00812258024581; 00812258024598; 00812258025304; 00812258025311; 00812258025328; 00812258025335; 00812258025342; 00812258024901; 00812258024918; 00812258024925; 00812258024932; 00812258024949; 00812258025939; 00812258025946; 00812258025953; 00812258025960; 00812258025977; 00812258023836; 00812258023843; 00812258023850; 00812258023119; 00812258023126; 00812258023133; 00812258023591; 00812258023607; 00812258023614; 00812258023355; 00812258023362; 00812258023379; 00812258024079; 00812258024086; 00812258024093; 00812258023867; 00812258023874; 00812258023881; 00812258023898; 00812258023904; 00812258023140; 00812258023157; 00812258023164; 00812258023171; 00812258023188; 00812258023621; 00812258023638; 00812258023645; 00812258023652; 00812258023669; 00812258023386; 00812258023393; 00812258023409; 00812258023416; 00812258023423; 00812258024109; 00812258024116; 00812258024123; 00812258024130; 00812258024147; 00812258023911; 00812258023928; 00812258023935; 00812258023195; 00812258023201; 00812258023218; 00812258023676; 00812258023683; 00812258023690; 00812258023430; 00812258023447; 00812258023454; 00812258024154; 00812258024161; 00812258024178; 00812258023942; 00812258023959; 00812258023966; 00812258023973; 00812258023980; 00812258023225; 00812258023232; 00812258023249; 00812258023256; 00812258023263; 00812258023706; 00812258023713; 00812258023720; 00812258023737; 00812258023744; 00812258023461; 00812258023478; 00812258023485; 00812258023492; 00812258023508; 00812258024185; 00812258024192; 00812258024208; 00812258024215; 00812258024222; 00812258023751; 00812258023768; 00812258023775; 00812258023034; 00812258023041; 00812258023058; 00812258023515; 00812258023522; 00812258023539; 00812258023270; 00812258023287; 00812258023294; 00812258023997; 00812258024000; 00812258024017; 00812258023782; 00812258023799; 00812258023805; 00812258023812; 00812258023829; 00812258023065; 00812258023072; 00812258023089; 00812258023096; 00812258023102; 00812258023546; 00812258023553; 00812258023560; 00812258023577; 00812258023584; 00812258023300; 00812258023317; 00812258023324; 00812258023331; 00812258023348; 00812258024024; 00812258024031; 00812258024048; 00812258024055; 00812258024062; 00887517936011; 00887517936028; 00887517936035; 00887517936042; 00887517936059; 00887517936066; 00887517936073; 00887517936080; 00887517936097; 00887517936202; 00887517936219; 00887517936226; 00887517000033; 00887517000040; 00887517000057; 00887517000064; 00887517000071; 00887517000088; 00887517000095; 00887517000101; 00887517000118; 00887517000125; 00887517000132; 00887517000149; 00887517000156; 00887517000163; 00887517000170; 00887517000187; 00887517000194; 00887517000200; 00887517000217; 00887517000224; 00887517000231; 00887517000248; 00887517000255; 00887517000255; 00812258027728; 00887517000262; 00812258027735; 00887517000279; 00812258027742; 00887517000286; 00812258027759; 00887517000293; 00812258027766; 00887517000309; 00812258027773; 00887517000316; 00812258027780; 00887517000323; 00812258027797; 00887517000330; 00812258027803; 00887517000347; 00812258027810; 00887517000354; 00812258027827; 00887517000361; 00812258027834; 00887517000378; 00812258027841; 00887517000385; 00812258027858; 00812258027865; 00887517000392; 00812258027872; 00812258027889; 00887517000408; 00812258027896; 00812258027902; 00887517000415; 00812258027919; 00812258027926; 00887517000422; 00812258027933; 00812258027940; 00887517000439; 00812258027957; 00887517000446; 00812258027964; 00887517000453; 00812258027971; 00887517000460; 00812258027988; 00887517000477; 00812258027995; 00887517000484; 00812258028008; 00812258028008; 00812258028015; 00812258028015;

Distribution

Worldwide distribution, including U.S. Nationwide.

Quantity

10,657 units