FDA Enforcement
Class II
Ongoing
Global Total LP single step medium, 60mL, bicarbonate-buffered media for D1-5 embryo culture and transfer, Reference Number H5GT-060
Recall: Z-1348-2023
·
Reported April 12, 2023
Enforcement
- Recall Number
- Z-1348-2023
- Event ID
- 91794
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- CooperSurgical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 12, 2023
- Initiation Date
- February 13, 2023
- Classification Date
- April 6, 2023
- Address
- 95 Corporate Dr, Trumbull, CT, 06611-1350, United States
Description
Global Total LP single step medium, 60mL, bicarbonate-buffered media for D1-5 embryo culture and transfer, Reference Number H5GT-060
Reason
It has come to CooperSurgical's attention that the affected Product may contain a medium other than the Global Total LP single step medium which does not contain the protein source required for embryo culture and development.
Code Info
UDI-DI: 00815965020433; LOT 221118-010694
Distribution
Worldwide - US Nationwide distribution in the states of NY, UT and the countries of Argentina, Belgium, Bulgaria, Ecuador, France, Germany, Hong Kong, Italy, Malaysia, Mexico, Peru, Singapore, Spain, Switzerland, Thailand, Turkey, the United Kingdom.
Quantity
190 units