FDA Enforcement Class III Terminated

TLC-2000 Therapeutic Medical Laser System Power Pack, Model Number TLC-2002

Recall: Z-1348-2020 · Reported March 4, 2020

Enforcement

Recall Number
Z-1348-2020
Event ID
84842
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Theralase Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
March 4, 2020
Initiation Date
September 26, 2019
Classification Date
February 24, 2020
Termination Date
December 3, 2020
Address
41 Hollinger Rd, N/A, East York, N/A, N/A, Canada

Description

TLC-2000 Therapeutic Medical Laser System Power Pack, Model Number TLC-2002

Reason

Power Pack (component) was shipped with an internal battery pack tested to UN-38.3, UL-2054 and IEC-62133 (2nd Edition) standards; however, not tested to IEC-60601-1-2 (3rd edition) standards

Code Info

Serial Numbers: 3104-3106 (inclusive); 3109 to 3141 (inclusive), 3143, 3146, 3147, 3155, 3158, 3159, 3163-3165 (inclusive), 3168-3173 (inclusive), 3180-3184 (inclusive), 3188-3332 (inclusive), 3334

Distribution

U.S.A.; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania

Quantity

202 units