FDA Enforcement Class II Terminated

cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.

Recall: Z-1346-2013 · Reported May 29, 2013

Enforcement

Recall Number
Z-1346-2013
Event ID
64930
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Roche Molecular Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 29, 2013
Initiation Date
April 1, 2013
Classification Date
May 21, 2013
Termination Date
August 7, 2014
Address
1080 Us Highway 202 S, N/A, Branchburg, NJ, 08876-3733, United States

Description

cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.

Reason

In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported

Code Info

Material Number 05200881001 - all lot/serial numbers.

Distribution

Nationwide Distribution.

Quantity

425 instruments