FDA Enforcement
Class II
Terminated
cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.
Recall: Z-1346-2013
·
Reported May 29, 2013
Enforcement
- Recall Number
- Z-1346-2013
- Event ID
- 64930
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Molecular Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 29, 2013
- Initiation Date
- April 1, 2013
- Classification Date
- May 21, 2013
- Termination Date
- August 7, 2014
- Address
- 1080 Us Highway 202 S, N/A, Branchburg, NJ, 08876-3733, United States
Description
cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.
Reason
In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported
Code Info
Material Number 05200881001 - all lot/serial numbers.
Distribution
Nationwide Distribution.
Quantity
425 instruments