FDA Enforcement Class II Ongoing

CELLSEARCH Circulating Tumor Cell Kit (Epithelial)-IVD intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood Part Number: 7900001

Recall: Z-1345-2022 · Reported July 13, 2022

Enforcement

Recall Number
Z-1345-2022
Event ID
90417
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Menarini Silicon Biosystems
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
July 13, 2022
Initiation Date
May 20, 2022
Classification Date
July 7, 2022
Address
3401 Masons Mill Rd Ste 100, N/A, Huntingdon Valley, PA, 19006-3574, United States

Description

CELLSEARCH Circulating Tumor Cell Kit (Epithelial)-IVD intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood Part Number: 7900001

Reason

High number of total images/unassigned events including (dual positives) and the potential for false positive results being placed into the image gallery in some patient samples

Code Info

UDI-DI: (01)10888385000016(17)221004(10)S108 Lot Number: S108 Exp. Date: 04Oct2022

Distribution

CA FL IL IN MA MI NC NM NY OH ON PA TX Foreign: Italy, Shanghai, Singapore,

Quantity

484 units