FDA Enforcement
Class II
Terminated
Amicus Exchange Kit Therapeutics - Product Usage: intended for use in the collection of blood components and mononuclear cells.
Recall: Z-1345-2021
·
Reported April 14, 2021
Enforcement
- Recall Number
- Z-1345-2021
- Event ID
- 87407
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Fenwal Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 14, 2021
- Initiation Date
- February 25, 2021
- Classification Date
- April 2, 2021
- Termination Date
- April 18, 2023
- Address
- 3 Corporate Dr Ste 300, N/A, Lake Zurich, IL, 60047-8930, United States
Description
Amicus Exchange Kit Therapeutics - Product Usage: intended for use in the collection of blood components and mononuclear cells.
Reason
Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using therapeutic protocols only. The defect is a blood leak at the boot, elbow, or in the channel on the separation chamber of the centrifuge pack.
Code Info
Product Code X6R2349; UDI: 04086000101950; Batch Numbers FA20E25150 (Exp. 05/31/2022) and FA20I22142 (Exp. 09/30/2022)
Distribution
US Nationwide distribution in the states of GA, TX, TN, MD, CA, KS, MN, MA, NM, NC, LA, CO, KY, NV, PA, UT, MI, NJ, IL, MD, OH, NY, FL.
Quantity
24 units