FDA Enforcement Class II Ongoing

Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 34X4,1BLA,1PRT QUICK Catalog number: 5921-034-135NS Indicated for use in surgical procedures that require the temporary occlusion of blood flow in a patient s extremities

Recall: Z-1341-2023 · Reported April 12, 2023

Enforcement

Recall Number
Z-1341-2023
Event ID
91849
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 12, 2023
Initiation Date
March 1, 2023
Classification Date
April 4, 2023
Address
1941 Stryker Way, Portage, MI, 49002-9711, United States

Description

Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 34X4,1BLA,1PRT QUICK Catalog number: 5921-034-135NS Indicated for use in surgical procedures that require the temporary occlusion of blood flow in a patient s extremities

Reason

Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications

Code Info

GTIN: 07613327606171 Lot Numbers: 2022111805 2022112801 2022112803 2022112804 2022112805 2022112806 2022112807 2022112808 2022120506 2022120507 2022120508

Distribution

US Nationwide distribution.

Quantity

1979 units (packs of 10)