FDA Enforcement
Class II
Ongoing
MINI STICK MAX 5F X 10 CM STIFF .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-758
Recall: Z-1340-2024
·
Reported March 27, 2024
Enforcement
- Recall Number
- Z-1340-2024
- Event ID
- 94035
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Angiodynamics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 27, 2024
- Initiation Date
- January 30, 2024
- Classification Date
- March 21, 2024
- Address
- 603 Queensbury Ave, Queensbury, NY, 12804-7619, United States
Description
MINI STICK MAX 5F X 10 CM STIFF .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-758
Reason
Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
Code Info
UDI: 15051684023016 UPN: H965457581 Lot Number: 5805060
Distribution
Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.
Quantity
15 units