Class II
Ongoing
Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 30X4,1BLA,2PRT QUICK Catalog number: 5921-030-235NS Indicated for use in surgical procedures that require the temporary occlusion of blood flow in a patient s extremities
Recall: Z-1340-2023
·
Reported April 12, 2023
Enforcement
- Recall Number
- Z-1340-2023
- Event ID
- 91849
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Stryker Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 12, 2023
- Initiation Date
- March 1, 2023
- Classification Date
- April 4, 2023
- Address
- 1941 Stryker Way, Portage, MI, 49002-9711, United States
Description
Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 30X4,1BLA,2PRT QUICK Catalog number: 5921-030-235NS Indicated for use in surgical procedures that require the temporary occlusion of blood flow in a patient s extremities
Reason
Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications
Code Info
GTIN: 07613327606195 Lot Numbers: 2022111804 2022112902 2022121909 2022121910 2022122603 2022122604
Distribution
US Nationwide distribution.
Quantity
1206 units (packs of 10)