FDA Enforcement
Class II
Terminated
Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical endoscopic camera system during surgical procedures.
Recall: Z-1340-2013
·
Reported May 29, 2013
Enforcement
- Recall Number
- Z-1340-2013
- Event ID
- 64647
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Linvatec Corp. dba ConMed Linvatec
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- May 29, 2013
- Initiation Date
- March 18, 2013
- Classification Date
- May 20, 2013
- Termination Date
- November 6, 2013
- Address
- 11311 Concept Blvd, N/A, Largo, FL, 33773-4908, United States
Description
Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical endoscopic camera system during surgical procedures.
Reason
The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key strokes (Caps lock and Num Lock) which may lock up and requires the user to reboot to continue.
Code Info
VP1600
Distribution
Worldwide Distribution-USA (nationwide) including the Virgin Islands and the states of CA, CT, FL, GA, IN, KY, KS, LA, MD, MI, MN, MO, MS, MT, NE, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, and WI, and the countries of Austria, Belgium, Canada, Denmark, Hong Kong and Zambia.
Quantity
267 units