FDA Enforcement Class II Ongoing

MINI STICK MAX 5F X 10 CM STD .018 NI/TU ECHO 2.75" PG Catalog Number: 45-756

Recall: Z-1339-2024 · Reported March 27, 2024

Enforcement

Recall Number
Z-1339-2024
Event ID
94035
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Angiodynamics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 27, 2024
Initiation Date
January 30, 2024
Classification Date
March 21, 2024
Address
603 Queensbury Ave, Queensbury, NY, 12804-7619, United States

Description

MINI STICK MAX 5F X 10 CM STD .018 NI/TU ECHO 2.75" PG Catalog Number: 45-756

Reason

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Code Info

UDI: 15051684022996 UPN: H965457561 Lot Number: 5796084 5796085 5796086 5796087 5796088 5796089 5796090 5796659 5796660 5796661 5796662 5796663 5796666 5797456 5797465 5798836 5798841 5798842 5799620 5800404 5803526 5804396

Distribution

Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.

Quantity

5176 units