FDA Enforcement
Class I
Ongoing
Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US 2. Medtronic HVAD Controller DC Adapter, Product Number: 1440
Recall: Z-1337-2021
·
Reported April 14, 2021
Enforcement
- Recall Number
- Z-1337-2021
- Event ID
- 87446
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Heartware, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- N/A
- Report Date
- April 14, 2021
- Initiation Date
- February 26, 2021
- Classification Date
- April 6, 2021
- Address
- 14400 Nw 60th Ave, N/A, Miami Lakes, FL, 33014-2807, United States
Description
Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US 2. Medtronic HVAD Controller DC Adapter, Product Number: 1440
Reason
There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.
Code Info
All lot codes
Distribution
Worldwide
Quantity
17992 units (8054 US)