FDA Enforcement
Class II
Terminated
Arrowg+ard Blue(R) PSI Kit, (1) For use with 7 - 7.5 Fr. Catheters Part Number (PN): ASK-29804-UHN2; (2) For use with 7.5 - 8 Fr. Catheters PN: ASK-29903-UCD
Recall: Z-1334-2018
·
Reported April 18, 2018
Enforcement
- Recall Number
- Z-1334-2018
- Event ID
- 79512
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Arrow International Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 18, 2018
- Initiation Date
- December 22, 2017
- Classification Date
- April 10, 2018
- Termination Date
- May 27, 2020
- Address
- 2400 Bernville Rd, Reading, PA, 19605-9607, United States
Description
Arrowg+ard Blue(R) PSI Kit, (1) For use with 7 - 7.5 Fr. Catheters Part Number (PN): ASK-29804-UHN2; (2) For use with 7.5 - 8 Fr. Catheters PN: ASK-29903-UCD
Reason
Certain lots of cardiovascular catheters may not be completely sealed and so sterility of the product cannot be guaranteed.
Code Info
Lot Numbers: (1) 13F17D0188, 13F17E0570; (2) 13F17E0660
Distribution
Distributed in all 50 U.S. states and the District of Columbia. Distributed in Bahamas, Canada, Mexico, Panama, and Singapore.