FDA Enforcement
Class II
Terminated
Spirit MB Bracket, UR Central +14T +5A.018, Part Number 494-0110, Lot Number 081351013. Product Usage: An orthodontic bracket. Spirit MB Brackets are a fixed attachment on a tooth which holds an archwire during orthodontic treatment.
Recall: Z-1330-2014
·
Reported April 9, 2014
Enforcement
- Recall Number
- Z-1330-2014
- Event ID
- 67849
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ormco/Sybronendo
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 9, 2014
- Initiation Date
- March 25, 2014
- Classification Date
- April 3, 2014
- Termination Date
- April 18, 2014
- Address
- 1332 S Lone Hill Ave, N/A, Glendora, CA, 91740-5339, United States
Description
Spirit MB Bracket, UR Central +14T +5A.018, Part Number 494-0110, Lot Number 081351013. Product Usage: An orthodontic bracket. Spirit MB Brackets are a fixed attachment on a tooth which holds an archwire during orthodontic treatment.
Reason
Ormco Corporation initiated this recall of Spirit MB Brackets, Part Number 494-0110, Lot Number 081351013, because the ID dot, which designates the bracket as distal/gingival, is misplaced. The misidentification may lead an orthodontist to place the bracket on the wrong tooth, leading to movement 10 degrees in the wrong direction.
Code Info
Part Number 494-0110, Lot Number 081351013.
Distribution
Internationally Distributed in the country of Japan.
Quantity
150 units