Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
Enforcement
- Recall Number
- Z-1327-2016
- Event ID
- 73561
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Synthes (USA) Products LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 13, 2016
- Initiation Date
- February 23, 2016
- Classification Date
- April 5, 2016
- Termination Date
- September 8, 2016
- Address
- 1301 Goshen Pkwy, N/A, West Chester, PA, 19380-5986, United States
Description
Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
Out of specification at the thread. The thread height of these screws is under-sized. The cross section of the affected area is less than that of conforming screws such that the thread height of the screws is undersized.
Titanium Low Profile Neuro Screw, Self-Drilling, 3mm, part number # 400.833E, Lot numbers: 9958565, 9951621, 9952976, 9955377, 9962314, 9968868, 9969645
Worldwide Distribution - US Nationwide in the states of MA, VA, TN, GA, WA, IA, NJ, TN, PA, NY, CO, CT, VT, and the countries of Canada and Switzerland.
150 units distributed in US, 1439 units distributed OUS (foreign)