FDA Enforcement Class II Terminated

Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.

Recall: Z-1326-2016 · Reported April 13, 2016

Enforcement

Recall Number
Z-1326-2016
Event ID
73561
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Synthes (USA) Products LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 13, 2016
Initiation Date
February 23, 2016
Classification Date
April 5, 2016
Termination Date
September 8, 2016
Address
1301 Goshen Pkwy, N/A, West Chester, PA, 19380-5986, United States

Description

Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.

Reason

Out of specification at the thread. The thread height of these screws is under-sized. The cross section of the affected area is less than that of conforming screws such that the thread height of the screws is undersized.

Code Info

Titanium Low Profile Neuro Screw, Self-Drilling, 3mm, part number # 400.833, Lot numbers: 9960359, 9962300, 9967875

Distribution

Worldwide Distribution - US Nationwide in the states of MA, VA, TN, GA, WA, IA, NJ, TN, PA, NY, CO, CT, VT, and the countries of Canada and Switzerland.

Quantity

500 units distributed in USA, and 35 units distributed OUS