FDA Enforcement Class II Ongoing

PRECICE STRYDE - Product Usage: intended for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

Recall: Z-1323-2021 · Reported April 7, 2021

Enforcement

Recall Number
Z-1323-2021
Event ID
87352
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Nuvasive Specialized Orthopedics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 7, 2021
Initiation Date
February 12, 2021
Classification Date
March 30, 2021
Address
101 Enterprise Ste 100, Aliso Viejo, CA, 92656-2604, United States

Description

PRECICE STRYDE - Product Usage: intended for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

Reason

Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

Code Info

All Lots affected. Below are: Primary DI Number(GTIN)/Product Code (REF)/Product Description: -887517981370 BT10-100C360-5 Antegrade Tibia 10¿ Bend, 10.0mm x 360mm, 5/10cm ST; 887517981554 BT10-10A360-5 Retrograde Femur 10¿ Bend, 10.0mm x 360mm, 5/10cm ST; 887517981547 BT10-10A360-7 Retrograde Femur 10¿ Bend, 10.0mm x 360mm, 7/10cm ST; 887517981530 BT10-10A380-5 Retrograde Femur 10¿ Bend, 10.0mm x 380mm, 5/10cm ST; 887517981523 BT10-10A380-7 Retrograde Femur 10¿ Bend, 10.0mm x 380mm, 7/10cm ST; 887517981516 BT10-10A400-5 Retrograde Femur 10¿ Bend, 10.0mm x 400mm, 5/10cm ST; 887517981509 BT10-10A400-7 Retrograde Femur 10¿ Bend, 10.0mm x 400mm, 7/10cm ST; 887517981493 BT10-10B360-5 Antegrade Femur Piriformis Straight, 10.0mm x 360mm, 5/10cm ST; 887517981486 BT10-10B360-7 Antegrade Femur Piriformis Straight, 10.0mm x 360mm, 7/10cm ST; 887517981479 BT10-10B380-5 Antegrade Femur Piriformis Straight, 10.0mm x 380mm, 5/10cm ST; 887517981462 BT10-10B380-7 Antegrade Femur Piriformis Straight, 10.0mm x 380mm, 7/10cm ST; 887517981455 BT10-10B400-5 Antegrade Femur Piriformis Straight, 10.0mm x 400mm, 5/10cm ST; 887517981448 BT10-10B400-7 Antegrade Femur Piriformis Straight, 10.0mm x 400mm, 7/10cm ST; 887517981431 BT10-10BT360-5 Antegrade Femur Piriformis Straight, 10.0mm x 360mm, 5/10cm ST; 887517981424 BT10-10BT360-7 Antegrade Femur Piriformis Straight, 10.0mm x 360mm, 7/10cm ST; 887517981417 BT10-10BT380-5 Antegrade Femur Piriformis Straight, 10.0mm x 380mm, 5/10cm ST; 887517981400 BT10-10BT380-7 Antegrade Femur Piriformis Straight, 10.0mm x 380mm, 7/10cm ST; 887517981394 BT10-10BT400-5 Antegrade Femur Piriformis Straight, 10.0mm x 400mm, 5/10cm ST; 887517981387 BT10-10BT400-7 Antegrade Femur Piriformis Straight, 10.0mm x 400mm, 7/10cm ST; 887517981363 BT10-10C360-7 Antegrade Tibia 10¿ Bend, 10.0mm x 360mm, 7/10cm ST; 887517981356 BT10-10C380-5 Antegrade Tibia 10¿ Bend, 10.0mm x 380mm, 5/10cm ST; 887517981349 BT10-10C380-7 Antegrade Tibia 10¿ Bend, 10.0mm x 380mm, 7/10cm ST; 887517981332 BT10-10C400-5 Antegrade Tibia 10¿ Bend, 10.0mm x 400mm, 5/10cm ST; 887517981325 BT10-10C400-7 Antegrade Tibia 10¿ Bend, 10.0mm x 400mm, 7/10cm ST; 887517981318 BT10-10D360-5 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 360mm, 5/10cm ST; 887517981301 BT10-10D360-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 360mm, 7/10cm ST; 887517981295 BT10-10D380-5 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 380mm, 5/10cm ST; 887517981288 BT10-10D380-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 380mm, 7/10cm ST; 887517981271 BT10-10D400-5 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 400mm, 5/10cm ST; 887517981264 BT10-10D400-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 400mm, 7/10cm ST; 887517981257 BT10-10DT360-5 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 360mm, 5/10cm ST; 887517981240 BT10-10DT360-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 360mm, 7/10cm ST; 887517981233 BT10-10DT380-5 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 380mm, 5/10cm ST; 887517981226 BT10-10DT380-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 380mm, 7/10cm ST; 887517981219 BT10-10DT400-5 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 400mm, 5/10cm ST; 887517981202 BT10-10DT400-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 400mm, 7/10cm ST; 887517981196 BT10-10E360-5 Retrograde Femur Straight, 10.0mm x 360mm, 5/10cm ST; 887517981189 BT10-10E360-7 Retrograde Femur Straight, 10.0mm x 360mm, 7/10cm ST; 887517981172 BT10-10E380-5 Retrograde Femur Straight, 10.0mm x 380mm, 5/10cm ST; 887517981165 BT10-10E380-7 Retrograde Femur Straight, 10.0mm x 380mm, 7/10cm ST; 887517981158 BT10-10E400-5 Retrograde Femur Straight, 10.0mm x 400mm, 5/10cm ST; 887517981141 BT10-10E400-7 Retrograde Femur Straight, 10.0mm x 400mm, 7/10cm ST; 887517981134 BT10-10SA360-5 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 360mm, 5/10cm ST; 887517981127 BT10-10SA360-7 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 360mm, 7/10cm ST; 887517981110 BT10-10SA380-5 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 380mm, 5/10cm ST; 887517981103 BT10-10SA380-7 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 380mm, 7/10cm ST; 887517981097 BT10-10SA400-5 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 400mm, 5/10cm ST; 887517981080 BT10-10SA400-7 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 400mm, 7/10cm ST; 887517981073 BT10-10SB360-5 Recon Antegrade Femur Piriformis Straight, 10.0mm x 360mm, 5/10cm ST; 887517981066 BT10-10SB360-7 Recon Antegrade Femur Piriformis Straight, 10.0mm x 360mm, 7/10cm ST; 887517981059 BT10-10SB380-5 Recon Antegrade Femur Piriformis Straight, 10.0mm x 380mm, 5/10cm ST; 887517981042 BT10-10SB380-7 Recon Antegrade Femur Piriformis Straight, 10.0mm x 380mm, 7/10cm ST; 887517981035 BT10-10SB400-5 Recon Antegrade Femur Piriformis Straight, 10.0mm x 400mm, 5/10cm ST; 887517981028 BT10-10SB400-7 Recon Antegrade Femur Piriformis Straight, 10.0mm x 400mm, 7/10cm ST; 887517981011 BT10-10SD360-5 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 360mm, 5/10cm ST; 887517981004 BT10-10SD360-7 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 360mm, 7/10cm ST; 887517980991 BT10-10SD380-5 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 380mm, 5/10cm ST; 887517980984 BT10-10SD380-7 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 380mm, 7/10cm ST; 887517980977 BT10-10SD400-5 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 400mm, 5/10cm ST; 887517980960 BT10-10SD400-7 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 400mm, 7/10cm ST; 887517980953 BT10-10SE360-5 Retrograde Femur Straight  Three Holes, 10.0mm x 360mm, 5/10cm ST; 887517980946 BT10-10SE360-7 Retrograde Femur Straight  Three Holes, 10.0mm x 360mm, 7/10cm ST; 887517980939 BT10-10SE380-5 Retrograde Femur Straight  Three Holes, 10.0mm x 380mm, 5/10cm ST; 887517980922 BT10-10SE380-7 Retrograde Femur Straight  Three Holes, 10.0mm x 380mm, 7/10cm ST; 887517980915 BT10-10SE400-5 Retrograde Femur Straight  Three Holes, 10.0mm x 400mm, 5/10cm ST; 887517980908 BT10-10SE400-7 Retrograde Femur Straight  Three Holes, 10.0mm x 400mm, 7/10cm ST; 887517980892 BT10-10SJ360-5 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 360mm, 5/10cm ST; 887517980885 BT10-10SJ360-7 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 360mm, 7/10cm ST; 887517980878 BT10-10SJ380-5 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 380mm, 5/10cm ST; 887517980861 BT10-10SJ380-7 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 380mm, 7/10cm ST; 887517980854 BT10-10SJ400-5 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 400mm, 5/10cm ST; 887517980847 BT10-10SJ400-7 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 400mm, 7/10cm ST; 887517980830 BT10-10V360-5 Retrograde Femur 10¿ Bend, 10.0mm x 360mm, 5/10cm ST; 887517980823 BT10-10V360-7 Retrograde Femur 10¿ Bend, 10.0mm x 360mm, 7/10cm ST; 887517980816 BT10-10V380-5 Retrograde Femur 10¿ Bend, 10.0mm x 380mm, 5/10cm ST; 887517980809 BT10-10V380-7 Retrograde Femur 10¿ Bend, 10.0mm x 380mm, 7/10cm ST; 887517980793 BT10-10V400-5 Retrograde Femur 10¿ Bend, 10.0mm x 400mm, 5/10cm ST; 887517980786 BT10-10V400-7 Retrograde Femur 10¿ Bend, 10.0mm x 400mm, 7/10cm ST; 887517980779 BT10-10X360-5 Retrograde Femur Straight, 10.0mm x 360mm, 5/10cm ST; 887517980762 BT10-10X360-7 Retrograde Femur Straight, 10.0mm x 360mm, 7/10cm ST; 887517980755 BT10-10X380-5 Retrograde Femur Straight, 10.0mm x 380mm, 5/10cm ST; 887517980748 BT10-10X380-7 Retrograde Femur Straight, 10.0mm x 380mm, 7/10cm ST; 887517980731 BT10-10X400-5 Retrograde Femur Straight, 10.0mm x 400mm, 5/10cm ST; 887517980724 BT10-10X400-7 Retrograde Femur Straight, 10.0mm x 400mm, 7/10cm ST; 887517980717 BT10-60A280-6 Retrograde Femur 10¿ Bend, 10.0mm x 280mm, 6/6cm ST; 887517980700 BT10-60B280-6 Antegrade Femur Piriformis Straight, 10.0mm x 280mm, 6/6cm ST; 887517980694 BT10-60BT280-6 Antegrade Femur Piriformis Straight, 10.0mm x 280mm, 6/6cm ST; 887517980687 BT10-60C280-6 Antegrade Tibia 10¿ Bend, 10.0mm x 280mm, 6/6cm ST; 887517980670 BT10-60D280-6 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 280mm, 6/6cm ST; 887517980663 BT10-60DT280-6 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 280mm, 6/6cm ST; 887517980656 BT10-60E280-6 Retrograde Femur Straight, 10.0mm x 280mm, 6/6cm ST; 887517980649 BT10-60SA280-6 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 280mm, 6/6cm ST; 887517980632 BT10-60SB280-6 Recon Antegrade Femur Piriformis Straight, 10.0mm x 280mm, 6/6cm ST; 887517980625 BT10-60SD280-6 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 280mm, 6/6cm ST; 887517980618 BT10-60SE280-6 Retrograde Femur Straight  Three Holes, 10.0mm x 280mm, 6/6cm ST; 887517980601 BT10-60SJ280-6 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 280mm, 6/6cm ST; 887517980595 BT10-60V280-6 Retrograde Femur 10¿ Bend, 10.0mm x 280mm, 6/6cm ST; 887517980588 BT10-60X280-6 Retrograde Femur Straight, 10.0mm x 280mm, 6/6cm ST; 887517980571 BT10-70A300-5 Retrograde Femur 10¿ Bend, 10.0mm x 300mm, 5/7cm ST; 887517980564 BT10-70A300-7 Retrograde Femur 10¿ Bend, 10.0mm x 300mm, 7/7cm ST; 887517980557 BT10-70B300-5 Antegrade Femur Piriformis Straight, 10.0mm x 300mm, 5/7cm ST; 887517980540 BT10-70B300-7 Antegrade Femur Piriformis Straight, 10.0mm x 300mm, 7/7cm ST; 887517980533 BT10-70BT300-5 Antegrade Femur Piriformis Straight, 10.0mm x 300mm, 5/7cm ST; 887517980526 BT10-70BT300-7 Antegrade Femur Piriformis Straight, 10.0mm x 300mm, 7/7cm ST; 887517980519 BT10-70C300-5 Antegrade Tibia 10¿ Bend, 10.0mm x 300mm, 5/7cm ST; 887517980502 BT10-70C300-7 Antegrade Tibia 10¿ Bend, 10.0mm x 300mm, 7/7cm ST; 887517980496 BT10-70D300-5 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 300mm, 5/7cm ST; 887517980489 BT10-70D300-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 300mm, 7/7cm ST; 887517980472 BT10-70DT300-5 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 300mm, 5/7cm ST; 887517980465 BT10-70DT300-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 300mm, 7/7cm ST; 887517980458 BT10-70E300-5 Retrograde Femur Straight, 10.0mm x 300mm, 5/7cm ST; 887517980441 BT10-70E300-7 Retrograde Femur Straight, 10.0mm x 300mm, 7/7cm ST; 887517980434 BT10-70SA300-5 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 300mm, 5/7cm ST; 887517980427 BT10-70SA300-7 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 300mm, 7/7cm ST; 887517980410 BT10-70SB300-5 Recon Antegrade Femur Piriformis Straight, 10.0mm x 300mm, 5/7cm ST; 887517980403 BT10-70SB300-7 Recon Antegrade Femur Piriformis Straight, 10.0mm x 300mm, 7/7cm ST; 887517980397 BT10-70SD300-5 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 300mm, 5/7cm ST; 887517980380 BT10-70SD300-7 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 300mm, 7/7cm ST; 887517980373 BT10-70SE300-5 Retrograde Femur Straight  Three Holes, 10.0mm x 300mm, 5/7cm ST; 887517980366 BT10-70SE300-7 Retrograde Femur Straight  Three Holes, 10.0mm x 300mm, 7/7cm ST; 887517980359 BT10-70SJ300-5 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 300mm, 5/7cm ST; 887517980342 BT10-70SJ300-7 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 300mm, 7/7cm ST; 887517980335 BT10-70V300-5 Retrograde Femur 10¿ Bend, 10.0mm x 300mm, 5/7cm ST; 887517980328 BT10-70V300-7 Retrograde Femur 10¿ Bend, 10.0mm x 300mm, 7/7cm ST; 887517980311 BT10-70X300-5 Retrograde Femur Straight, 10.0mm x 300mm, 5/7cm ST; 887517980304 BT10-70X300-7 Retrograde Femur Straight, 10.0mm x 300mm, 7/7cm ST; 887517980298 BT10-80A320-5 Retrograde Femur 10¿ Bend, 10.0mm x 320mm, 5/8cm ST; 887517980281 BT10-80A320-7 Retrograde Femur 10¿ Bend, 10.0mm x 320mm, 7/8cm ST; 887517980274 BT10-80B320-5 Antegrade Femur Piriformis Straight, 10.0mm x 320mm, 5/8cm ST; 887517980267 BT10-80B320-7 Antegrade Femur Piriformis Straight, 10.0mm x 320mm, 7/8cm ST; 887517980250 BT10-80BT320-5 Antegrade Femur Piriformis Straight, 10.0mm x 320mm, 5/8cm ST; 887517980243 BT10-80BT320-7 Antegrade Femur Piriformis Straight, 10.0mm x 320mm, 7/8cm ST; 887517980236 BT10-80C320-5 Antegrade Tibia 10¿ Bend, 10.0mm x 320mm, 5/8cm ST; 887517980229 BT10-80C320-7 Antegrade Tibia 10¿ Bend, 10.0mm x 320mm, 7/8cm ST; 887517980212 BT10-80D320-5 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 320mm, 5/8cm ST; 887517980205 BT10-80D320-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 320mm, 7/8cm ST; 887517980199 BT10-80DT320-5 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 320mm, 5/8cm ST; 887517980182 BT10-80DT320-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 320mm, 7/8cm ST; 887517980175 BT10-80E320-5 Retrograde Femur Straight, 10.0mm x 320mm, 5/8cm ST; 887517980168 BT10-80E320-7 Retrograde Femur Straight, 10.0mm x 320mm, 7/8cm ST; 887517980151 BT10-80SA320-5 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 320mm, 5/8cm ST; 887517980144 BT10-80SA320-7 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 320mm, 7/8cm ST; 887517980137 BT10-80SB320-5 Recon Antegrade Femur Piriformis Straight, 10.0mm x 320mm, 5/8cm ST; 887517980120 BT10-80SB320-7 Recon Antegrade Femur Piriformis Straight, 10.0mm x 320mm, 7/8cm ST; 887517980113 BT10-80SD320-5 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 320mm, 5/8cm ST; 887517980106 BT10-80SD320-7 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 320mm, 7/8cm ST; 887517980090 BT10-80SE320-5 Retrograde Femur Straight  Three Holes, 10.0mm x 320mm, 5/8cm ST; 887517980083 BT10-80SE320-7 Retrograde Femur Straight  Three Holes, 10.0mm x 320mm, 7/8cm ST; 887517980076 BT10-80SJ320-5 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 320mm, 5/8cm ST; 887517980069 BT10-80SJ320-7 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 320mm, 7/8cm ST; 887517980052 BT10-80V320-5 Retrograde Femur 10¿ Bend, 10.0mm x 320mm, 5/8cm ST; 887517980045 BT10-80V320-7 Retrograde Femur 10¿ Bend, 10.0mm x 320mm, 7/8cm ST; 887517980038 BT10-80X320-5 Retrograde Femur Straight, 10.0mm x 320mm, 5/8cm ST; 887517980021 BT10-80X320-7 Retrograde Femur Straight, 10.0mm x 320mm, 7/8cm ST; 887517980014 BT10-90A340-5 Retrograde Femur 10¿ Bend, 10.0mm x 340mm, 5/9cm ST; 887517980007 BT10-90A340-7 Retrograde Femur 10¿ Bend, 10.0mm x 340mm, 7/9cm ST; 887517979995 BT10-90B340-5 Antegrade Femur Piriformis Straight, 10.0mm x 340mm, 5/9cm ST; 887517979988 BT10-90B340-7 Antegrade Femur Piriformis Straight, 10.0mm x 340mm, 7/9cm ST; 887517979971 BT10-90BT340-5 Antegrade Femur Piriformis Straight, 10.0mm x 340mm, 5/9cm ST; 887517979964 BT10-90BT340-7 Antegrade Femur Piriformis Straight, 10.0mm x 340mm, 7/9cm ST; 887517979957 BT10-90C340-5 Antegrade Tibia 10¿ Bend, 10.0mm x 340mm, 5/9cm ST; 887517979940 BT10-90C340-7 Antegrade Tibia 10¿ Bend, 10.0mm x 340mm, 7/9cm ST; 887517979933 BT10-90D340-50 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 340mm, 5/9cm ST; 887517979926 BT10-90D340-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 340mm, 7/9cm ST; 887517979919 BT10-90DT340-5 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 340mm, 5/9cm ST; 887517979902 BT10-90DT340-7 Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 340mm, 7/9cm ST; 887517979896 BT10-90E340-5 Retrograde Femur Straight, 10.0mm x 340mm, 5/9cm ST; 887517979889 BT10-90E340-7 Retrograde Femur Straight, 10.0mm x 340mm, 7/9cm ST; 887517979872 BT10-90SA340-5 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 340mm, 5/9cm ST; 887517979865 BT10-90SA340-7 Retrograde Femur 5¿ Bend  Three Holes, 10.0mm x 340mm, 7/9cm ST; 887517979858 BT10-90SB340-5 Recon Antegrade Femur Piriformis Straight, 10.0mm x 340mm, 5/9cm ST; 887517979841 BT10-90SB340-7 Recon Antegrade Femur Piriformis Straight, 10.0mm x 340mm, 7/9cm ST; 887517979834 BT10-90SD340-5 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 340mm, 5/9cm ST; 887517979827 BT10-90SD340-7 Recon Antegrade Femur Trochanter 10¿ Bend, 10.0mm x 340mm, 7/9cm ST; 887517979810 BT10-90SE340-5 Retrograde Femur Straight  Three Holes, 10.0mm x 340mm, 5/9cm ST; 887517979803 BT10-90SE340-7 Retrograde Femur Straight  Three Holes, 10.0mm x 340mm, 7/9cm ST; 887517979797 BT10-90SJ340-5 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 340mm, 5/9cm ST; 887517979780 BT10-90SJ340-7 Antegrade Tibia 10¿ Bend  Three Holes, 10.0mm x 340mm, 7/9cm ST; 887517979773 BT10-90V340-5 Retrograde Femur 10¿ Bend, 10.0mm x 340mm, 5/9cm ST; 887517979766 BT10-90V340-7 Retrograde Femur 10¿ Bend, 10.0mm x 340mm, 7/9cm ST; 887517979759 BT10-90X340-5 Retrograde Femur Straight, 10.0mm x 340mm, 5/9cm ST; 887517979742 BT10-90X340-7 Retrograde Femur Straight, 10.0mm x 340mm, 7/9cm ST; 887517979735 BT115-10A360-5 Retrograde Femur 10¿ Bend, 11.5mm x 360mm, 5/10cm ST; 887517979728 BT115-10A360-7 Retrograde Femur 10¿ Bend, 11.5mm x 360mm, 7/10cm ST; 887517979711 BT115-10A380-5 Retrograde Femur 10¿ Bend, 11.5mm x 380mm, 5/10cm ST; 887517979704 BT115-10A380-7 Retrograde Femur 10¿ Bend, 11.5mm x 380mm, 7/10cm ST; 887517979698 BT115-10A400-5 Retrograde Femur 10¿ Bend, 11.5mm x 400mm, 5/10cm ST; 887517979681 BT115-10A400-7 Retrograde Femur 10¿ Bend, 11.5mm x 400mm, 7/10cm ST; 887517979674 BT115-10B360-5 Antegrade Femur Piriformis Straight, 11.5mm x 360mm, 5/10cm ST; 887517979667 BT115-10B360-7 Antegrade Femur Piriformis Straight, 11.5mm x 360mm, 7/10cm ST; 887517979650 BT115-10B380-5 Antegrade Femur Piriformis Straight, 11.5mm x 380mm, 5/10cm ST; 887517979643 BT115-10B380-7 Antegrade Femur Piriformis Straight, 11.5mm x 380mm, 7/10cm ST; 887517979636 BT115-10B400-5 Antegrade Femur Piriformis Straight, 11.5mm x 400mm, 5/10cm ST; 887517979629 BT115-10B400-7 Antegrade Femur Piriformis Straight, 11.5mm x 400mm, 7/10cm ST; 887517979612 BT115-10BT360-5 Antegrade Femur Piriformis Straight, 11.5mm x 360mm, 5/10cm ST; 887517979605 BT115-10BT360-7 Antegrade Femur Piriformis Straight, 11.5mm x 360mm, 7/10cm ST; 887517979599 BT115-10BT380-5 Antegrade Femur Piriformis Straight, 11.5mm x 380mm, 5/10cm ST; 887517979582 BT115-10BT380-7 Antegrade Femur Piriformis Straight, 11.5mm x 380mm, 7/10cm ST; 887517979575 BT115-10BT400-5 Antegrade Femur Piriformis Straight, 11.5mm x 400mm, 5/10cm ST; 887517979568 BT115-10BT400-7 Antegrade Femur Piriformis Straight, 11.5mm x 400mm, 7/10cm ST; 887517979551 BT115-10C360-5 Antegrade Tibia 10¿ Bend, 11.5mm x 360mm, 5/10cm ST; 887517979544 BT115-10C360-7 Antegrade Tibia 10¿ Bend, 11.5mm x 360mm, 7/10cm ST; 887517979537 BT115-10C380-5 Antegrade Tibia 10¿ Bend, 11.5mm x 380mm, 5/10cm ST; 887517979520 BT115-10C380-7 Antegrade Tibia 10¿ Bend, 11.5mm x 380mm, 7/10cm ST; 887517979513 BT115-10C400-5 Antegrade Tibia 10¿ Bend, 11.5mm x 400mm, 5/10cm ST; 887517979506 BT115-10C400-7 Antegrade Tibia 10¿ Bend, 11.5mm x 400mm, 7/10cm ST; 887517979490 BT115-10D360-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 360mm, 5/10cm ST; 887517979483 BT115-10D360-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 360mm, 7/10cm ST; 887517979476 BT115-10D380-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 380mm, 5/10cm ST; 887517979469 BT115-10D380-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 380mm, 7/10cm ST; 887517979452 BT115-10D400-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 400mm, 5/10cm ST; 887517979445 BT115-10D400-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 400mm, 7/10cm ST; 887517979438 BT115-10DT360-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 360mm, 5/10cm ST; 887517979421 BT115-10DT360-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 360mm, 7/10cm ST; 887517979414 BT115-10DT380-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 380mm, 5/10cm ST; 887517979407 BT115-10DT380-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 380mm, 7/10cm ST; 887517979391 BT115-10DT400-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 400mm, 5/10cm ST; 887517979384 BT115-10DT400-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 400mm, 7/10cm ST; 887517979377 BT115-10E360-5 Retrograde Femur Straight, 11.5mm x 360mm, 5/10cm ST; 887517979360 BT115-10E360-7 Retrograde Femur Straight, 11.5mm x 360mm, 7/10cm ST; 887517979353 BT115-10E380-5 Retrograde Femur Straight, 11.5mm x 380mm, 5/10cm ST; 887517979346 BT115-10E380-7 Retrograde Femur Straight, 11.5mm x 380mm, 7/10cm ST; 887517979339 BT115-10E400-5 Retrograde Femur Straight, 11.5mm x 400mm, 5/10cm ST; 887517979322 BT115-10E400-7 Retrograde Femur Straight, 11.5mm x 400mm, 7/10cm ST; 887517979315 BT115-10SA360-5 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 360mm, 5/10cm ST; 887517979308 BT115-10SA360-7 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 360mm, 7/10cm ST; 887517979292 BT115-10SA380-5 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 380mm, 5/10cm ST; 887517979285 BT115-10SA380-7 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 380mm, 7/10cm ST; 887517979278 BT115-10SA400-5 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 400mm, 5/10cm ST; 887517979261 BT115-10SA400-7 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 400mm, 7/10cm ST; 887517979254 BT115-10SB360-5 Recon Antegrade Femur Piriformis Straight, 11.5mm x 360mm, 5/10cm ST; 887517979247 BT115-10SB360-7 Recon Antegrade Femur Piriformis Straight, 11.5mm x 360mm, 7/10cm ST; 887517979230 BT115-10SB380-5 Recon Antegrade Femur Piriformis Straight, 11.5mm x 380mm, 5/10cm ST; 887517979223 BT115-10SB380-7 Recon Antegrade Femur Piriformis Straight, 11.5mm x 380mm, 7/10cm ST; 887517979216 BT115-10SB400-5 Recon Antegrade Femur Piriformis Straight, 11.5mm x 400mm, 5/10cm ST; 887517979209 BT115-10SB400-7 Recon Antegrade Femur Piriformis Straight, 11.5mm x 400mm, 7/10cm ST; 887517979193 BT115-10SD360-5 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 360mm, 5/10cm ST; 887517979186 BT115-10SD360-7 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 360mm, 7/10cm ST; 887517979179 BT115-10SD380-5 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 380mm, 5/10cm ST; 887517979162 BT115-10SD380-7 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 380mm, 7/10cm ST; 887517979155 BT115-10SD400-5 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 400mm, 5/10cm ST; 887517979148 BT115-10SD400-7 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 400mm, 7/10cm ST; 887517979131 BT115-10SE360-5 Retrograde Femur Straight  Three Holes, 11.5mm x 360mm, 5/10cm ST; 887517979124 BT115-10SE360-7 Retrograde Femur Straight  Three Holes, 11.5mm x 360mm, 7/10cm ST; 887517979117 BT115-10SE380-5 Retrograde Femur Straight  Three Holes, 11.5mm x 380mm, 5/10cm ST; 887517979100 BT115-10SE380-7 Retrograde Femur Straight  Three Holes, 11.5mm x 380mm, 7/10cm ST; 887517979094 BT115-10SE400-5 Retrograde Femur Straight  Three Holes, 11.5mm x 400mm, 5/10cm ST; 887517979087 BT115-10SE400-7 Retrograde Femur Straight  Three Holes, 11.5mm x 400mm, 7/10cm ST; 887517979070 BT115-10SJ360-5 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 360mm, 5/10cm ST; 887517979063 BT115-10SJ360-7 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 360mm, 7/10cm ST; 887517979056 BT115-10SJ380-5 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 380mm, 5/10cm ST; 887517979049 BT115-10SJ380-7 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 380mm, 7/10cm ST; 887517979032 BT115-10SJ400-5 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 400mm, 5/10cm ST; 887517979025 BT115-10SJ400-7 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 400mm, 7/10cm ST; 887517979018 BT115-10V360-5 Retrograde Femur 10¿ Bend, 11.5mm x 360mm, 5/10cm ST; 887517979001 BT115-10V360-7 Retrograde Femur 10¿ Bend, 11.5mm x 360mm, 7/10cm ST; 887517978998 BT115-10V380-5 Retrograde Femur 10¿ Bend, 11.5mm x 380mm, 5/10cm ST; 887517978981 BT115-10V380-7 Retrograde Femur 10¿ Bend, 11.5mm x 380mm, 7/10cm ST; 887517978974 BT115-10V400-5 Retrograde Femur 10¿ Bend, 11.5mm x 400mm, 5/10cm ST; 887517978967 BT115-10V400-7 Retrograde Femur 10¿ Bend, 11.5mm x 400mm, 7/10cm ST; 887517978950 BT115-10X360-5 Retrograde Femur Straight, 11.5mm x 360mm, 5/10cm ST; 887517978943 BT115-10X360-7 Retrograde Femur Straight, 11.5mm x 360mm, 7/10cm ST; 887517978936 BT115-10X380-5 Retrograde Femur Straight, 11.5mm x 380mm, 5/10cm ST; 887517978929 BT115-10X380-7 Retrograde Femur Straight, 11.5mm x 380mm, 7/10cm ST; 887517978912 BT115-10X400-5 Retrograde Femur Straight, 11.5mm x 400mm, 5/10cm ST; 887517978905 BT115-10X400-7 Retrograde Femur Straight, 11.5mm x 400mm, 7/10cm ST; 887517978899 BT115-60A280-6 Retrograde Femur 10¿ Bend, 11.5mm x 280mm, 6/6cm ST; 887517978882 BT115-60B280-6 Antegrade Femur Piriformis Straight, 11.5mm x 280mm, 6/6cm ST; 887517978875 BT115-60BT280-6 Antegrade Femur Piriformis Straight, 11.5mm x 280mm, 6/6cm ST; 887517978868 BT115-60C280-6 Antegrade Tibia 10¿ Bend, 11.5mm x 280mm, 6/6cm ST; 887517978851 BT115-60D280-6 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 280mm, 6/6cm ST; 887517978844 BT115-60DT280-6 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 280mm, 6/6cm ST; 887517978837 BT115-60E280-6 Retrograde Femur Straight, 11.5mm x 280mm, 6/6cm ST; 887517978820 BT115-60SA280-6 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 280mm, 6/6cm ST; 887517978813 BT115-60SB280-6 Recon Antegrade Femur Piriformis Straight, 11.5mm x 280mm, 6/6cm ST; 887517978806 BT115-60SD280-6 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 280mm, 6/6cm ST; 887517978790 BT115-60SE280-6 Retrograde Femur Straight  Three Holes, 11.5mm x 280mm, 6/6cm ST; 887517978783 BT115-60SJ280-6 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 280mm, 6/6cm ST; 887517978776 BT115-60V280-6 Retrograde Femur 10¿ Bend, 11.5mm x 280mm, 6/6cm ST; 887517978769 BT115-60X280-6 Retrograde Femur Straight, 11.5mm x 280mm, 6/6cm ST; 887517978752 BT115-70A300-5 Retrograde Femur 10¿ Bend, 11.5mm x 300mm, 5/7cm ST; 887517978745 BT115-70A300-7 Retrograde Femur 10¿ Bend, 11.5mm x 300mm, 7/7cm ST; 887517978738 BT115-70B300-5 Antegrade Femur Piriformis Straight, 11.5mm x 300mm, 5/7cm ST; 887517978721 BT115-70B300-7 Antegrade Femur Piriformis Straight, 11.5mm x 300mm, 7/7cm ST; 887517978714 BT115-70BT300-5 Antegrade Femur Piriformis Straight, 11.5mm x 300mm, 5/7cm ST; 887517978707 BT115-70BT300-7 Antegrade Femur Piriformis Straight, 11.5mm x 300mm, 7/7cm ST; 887517978691 BT115-70C300-5 Antegrade Tibia 10¿ Bend, 11.5mm x 300mm, 5/7cm ST; 887517978684 BT115-70C300-7 Antegrade Tibia 10¿ Bend, 11.5mm x 300mm, 7/7cm ST; 887517978677 BT115-70D300-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 300mm, 5/7cm ST; 887517978660 BT115-70D300-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 300mm, 7/7cm ST; 887517978653 BT115-70DT300-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 300mm, 5/7cm ST; 887517978646 BT115-70DT300-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 300mm, 7/7cm ST; 887517978639 BT115-70E300-5 Retrograde Femur Straight, 11.5mm x 300mm, 5/7cm ST; 887517978622 BT115-70E300-7 Retrograde Femur Straight, 11.5mm x 300mm, 7/7cm ST; 887517978615 BT115-70SA300-5 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 300mm, 5/7cm ST; 887517978608 BT115-70SA300-7 Retrograde Femur 5¿ Bend  Three Holes, 11.5 mm x 300mm, 7/7cm ST; 887517978592 BT115-70SB300-5 Recon Antegrade Femur Piriformis Straight, 11.5mm x 300mm, 5/7cm ST; 887517978585 BT115-70SB300-7 Recon Antegrade Femur Piriformis Straight, 11.5mm x 300mm, 7/7cm ST; 887517978578 BT115-70SD300-5 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 300mm, 5/7cm ST; 887517978561 BT115-70SD300-7 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 300mm, 7/7cm ST; 887517978554 BT115-70SE300-5 Retrograde Femur Straight  Three Holes, 11.5mm x 300mm, 5/7cm ST; 887517978547 BT115-70SE300-7 Retrograde Femur Straight  Three Holes, 11.5mm x 300mm, 7/7cm ST; 887517978530 BT115-70SJ300-5 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 300mm, 5/7cm ST; 887517978523 BT115-70SJ300-7 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 300mm, 7/7cm ST; 887517978516 BT115-70V300-5 Retrograde Femur 10¿ Bend, 11.5mm x 300mm, 5/7cm ST; 887517978509 BT115-70V300-7 Retrograde Femur 10¿ Bend, 11.5mm x 300mm, 7/7cm ST; 887517978493 BT115-70X300-5 Retrograde Femur Straight, 11.5mm x 300mm, 5/7cm ST; 887517978486 BT115-70X300-7 Retrograde Femur Straight, 11.5mm x 300mm, 7/7cm ST; 887517978479 BT115-80A320-5 Retrograde Femur 10¿ Bend, 11.5mm x 320mm, 5/8cm ST; 887517978462 BT115-80A320-7 Retrograde Femur 10¿ Bend, 11.5mm x 320mm, 7/8cm ST; 887517978455 BT115-80B320-5 Antegrade Femur Piriformis Straight, 11.5mm x 320mm, 5/8cm ST; 887517978448 BT115-80B320-7 Antegrade Femur Piriformis Straight, 11.5mm x 320mm, 7/8cm ST; 887517978431 BT115-80BT320-5 Antegrade Femur Piriformis Straight, 11.5mm x 320mm, 5/8cm ST; 887517978424 BT115-80BT320-7 Antegrade Femur Piriformis Straight, 11.5mm x 320mm, 7/8cm ST; 887517978417 BT115-80C320-5 Antegrade Tibia 10¿ Bend, 11.5mm x 320mm, 5/8cm ST; 887517978400 BT115-80C320-7 Antegrade Tibia 10¿ Bend, 11.5mm x 320mm, 7/8cm ST; 887517978394 BT115-80D320-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 320mm, 5/8cm ST; 887517978387 BT115-80D320-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 320mm, 7/8cm ST; 887517978370 BT115-80DT320-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 320mm, 5/8cm ST; 887517978363 BT115-80DT320-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 320mm, 7/8cm ST; 887517978356 BT115-80E320-5 Retrograde Femur Straight, 11.5mm x 320mm, 5/8cm ST; 887517978349 BT115-80E320-7 Retrograde Femur Straight, 11.5mm x 320mm, 7/8cm ST; 887517978332 BT115-80SA320-5 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 320mm, 5/8cm ST; 887517978325 BT115-80SA320-7 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 320mm, 7/8cm ST; 887517978318 BT115-80SB320-5 Recon Antegrade Femur Piriformis Straight, 11.5mm x 320mm, 5/8cm ST; 887517978301 BT115-80SB320-7 Recon Antegrade Femur Piriformis Straight, 11.5mm x 320mm, 7/8cm ST; 887517978295 BT115-80SD320-5 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 320mm, 5/8cm ST; 887517978288 BT115-80SD320-7 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 320mm, 7/8cm ST; 887517978271 BT115-80SE320-5 Retrograde Femur Straight  Three Holes, 11.5mm x 320mm, 5/8cm ST; 887517978264 BT115-80SE320-7 Retrograde Femur Straight  Three Holes, 11.5mm x 320mm, 7/8cm ST; 887517978257 BT115-80SJ320-5 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 320mm, 5/8cm ST; 887517978240 BT115-80SJ320-7 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 320mm, 7/8cm ST; 887517978233 BT115-80V320-5 Retrograde Femur 10¿ Bend, 11.5mm x 320mm, 5/8cm ST; 887517978226 BT115-80V320-7 Retrograde Femur 10¿ Bend, 11.5mm x 320mm, 7/8cm ST; 887517978219 BT115-80X320-5 Retrograde Femur Straight, 11.5mm x 320mm, 5/8cm ST; 887517978202 BT115-80X320-7 Retrograde Femur Straight, 11.5mm x 320mm, 7/8cm ST; 887517978196 BT115-90A340-5 Retrograde Femur 10¿ Bend, 11.5mm x 340mm, 5/9cm ST; 887517978189 BT115-90A340-7 Retrograde Femur 10¿ Bend, 11.5mm x 340mm, 7/9cm ST; 887517978172 BT115-90B340-5 Antegrade Femur Piriformis Straight, 11.5mm x 340mm, 5/9cm ST; 887517978165 BT115-90B340-7 Antegrade Femur Piriformis Straight, 11.5mm x 340mm, 7/9cm ST; 887517978158 BT115-90BT340-5 Antegrade Femur Piriformis Straight, 11.5mm x 340mm, 5/9cm ST; 887517978141 BT115-90BT340-7 Antegrade Femur Piriformis Straight, 11.5mm x 340mm, 7/9cm ST; 887517978134 BT115-90C340-5 Antegrade Tibia 10¿ Bend, 11.5mm x 340mm, 5/9cm ST; 887517978127 BT115-90C340-7 Antegrade Tibia 10¿ Bend, 11.5mm x 340mm, 7/9cm ST; 887517978110 BT115-90D340-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 340mm, 5/9cm ST; 887517978103 BT115-90D340-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 340mm, 7/9cm ST; 887517978097 BT115-90DT340-5 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 340mm, 5/9cm ST; 887517978080 BT115-90DT340-7 Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 340mm, 7/9cm ST; 887517978073 BT115-90E340-5 Retrograde Femur Straight, 11.5mm x 340mm, 5/9cm ST; 887517978066 BT115-90E340-7 Retrograde Femur Straight, 11.5mm x 340mm, 7/9cm ST; 887517978059 BT115-90SA340-5 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 340mm, 5/9cm ST; 887517978042 BT115-90SA340-7 Retrograde Femur 5¿ Bend  Three Holes, 11.5mm x 340mm, 7/9cm ST; 887517978035 BT115-90SB340-5 Recon Antegrade Femur Piriformis Straight, 11.5mm x 340mm, 5/9cm ST; 887517978028 BT115-90SB340-7 Recon Antegrade Femur Piriformis Straight, 11.5mm x 340mm, 7/9cm ST; 887517978011 BT115-90SD340-5 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 340mm, 5/9cm ST; 887517978004 BT115-90SD340-7 Recon Antegrade Femur Trochanter 10¿ Bend, 11.5mm x 340mm, 7/9cm ST; 887517977991 BT115-90SE340-5 Retrograde Femur Straight  Three Holes, 11.5mm x 340mm, 5/9cm ST; 887517977984 BT115-90SE340-7 Retrograde Femur Straight  Three Holes, 11.5mm x 340mm, 7/9cm ST; 887517977977 BT115-90SJ340-5 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 340mm, 5/9cm ST; 887517977960 BT115-

More Code Info

90SJ340-7 Antegrade Tibia 10¿ Bend  Three Holes, 11.5mm x 340mm, 7/9cm ST; 887517977953 BT115-90V340-5 Retrograde Femur 10¿ Bend, 11.5mm x 340mm, 5/9cm ST; 887517977946 BT115-90V340-7 Retrograde Femur 10¿ Bend, 11.5mm x 340mm, 7/9cm ST; 887517977939 BT115-90X340-5 Retrograde Femur Straight, 11.5mm x 340mm, 5/9cm ST; 887517977922 BT115-90X340-7 Retrograde Femur Straight, 11.5mm x 340mm, 7/9cm ST; 887517977915 BT13-10A360-5 Retrograde Femur 10¿ Bend, 13.0mm x 360mm, 5/10cm ST; 887517977908 BT13-10A360-7 Retrograde Femur 10¿ Bend, 13.0mm x 360mm, 7/10cm ST; 887517977892 BT13-10A380-5 Retrograde Femur 10¿ Bend, 13.0mm x 380mm, 5/10cm ST; 887517977885 BT13-10A380-7 Retrograde Femur 10¿ Bend, 13.0mm x 380mm, 7/10cm ST; 887517977878 BT13-10A400-5 Retrograde Femur 10¿ Bend, 13.0mm x 400mm, 5/10cm ST; 887517977861 BT13-10A400-7 Retrograde Femur 10¿ Bend, 13.0mm x 400mm, 7/10cm ST; 887517977854 BT13-10B360-5 Antegrade Femur Piriformis Straight, 13.0mm x 360mm, 5/10cm ST; 887517977847 BT13-10B360-7 Antegrade Femur Piriformis Straight, 13.0mm x 360mm, 7/10cm ST; 887517977830 BT13-10B380-5 Antegrade Femur Piriformis Straight, 13.0mm x 380mm, 5/10cm ST; 887517977823 BT13-10B380-7 Antegrade Femur Piriformis Straight, 13.0mm x 380mm, 7/10cm ST; 887517977816 BT13-10B400-5 Antegrade Femur Piriformis Straight, 13.0mm x 400mm, 5/10cm ST; 887517977809 BT13-10B400-7 Antegrade Femur Piriformis Straight, 13.0mm x 400mm, 7/10cm ST; 887517977793 BT13-10BT360-5 Antegrade Femur Piriformis Straight, 13.0mm x 360mm, 5/10cm ST; 887517977786 BT13-10BT360-7 Antegrade Femur Piriformis Straight, 13.0mm x 360mm, 7/10cm ST; 887517977779 BT13-10BT380-5 Antegrade Femur Piriformis Straight, 13.0mm x 380mm, 5/10cm ST; 887517977762 BT13-10BT380-7 Antegrade Femur Piriformis Straight, 13.0mm x 380mm, 7/10cm ST; 887517977755 BT13-10BT400-5 Antegrade Femur Piriformis Straight, 13.0mm x 400mm, 5/10cm ST; 887517977748 BT13-10BT400-7 Antegrade Femur Piriformis Straight, 13.0mm x 400mm, 7/10cm ST; 887517977731 BT13-10C360-5 Antegrade Tibia 10¿ Bend, 13.0mm x 360mm, 5/10cm ST; 887517977724 BT13-10C360-7 Antegrade Tibia 10¿ Bend, 13.0mm x 360mm, 7/10cm ST; 887517977717 BT13-10C380-5 Antegrade Tibia 10¿ Bend, 13.0mm x 380mm, 5/10cm ST; 887517977700 BT13-10C380-7 Antegrade Tibia 10¿ Bend, 13.0mm x 380mm, 7/10cm ST; 887517977694 BT13-10C400-5 Antegrade Tibia 10¿ Bend, 13.0mm x 400mm, 5/10cm ST; 887517977687 BT13-10C400-7 Antegrade Tibia 10¿ Bend, 13.0mm x 400mm, 7/10cm ST; 887517977670 BT13-10D360-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 360mm, 5/10cm ST; 887517977663 BT13-10D360-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 360mm, 7/10cm ST; 887517977656 BT13-10D380-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 380mm, 5/10cm ST; 887517977649 BT13-10D380-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 380mm, 7/10cm ST; 887517977632 BT13-10D400-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 400mm, 5/10cm ST; 887517977625 BT13-10D400-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 400mm, 7/10cm ST; 887517977618 BT13-10DT360-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 360mm, 5/10cm ST; 887517977601 BT13-10DT360-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 360mm, 7/10cm ST; 887517977595 BT13-10DT380-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 380mm, 5/10cm ST; 887517977588 BT13-10DT380-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 380mm, 7/10cm ST; 887517977571 BT13-10DT400-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 400mm, 5/10cm ST; 887517977564 BT13-10DT400-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 400mm, 7/10cm ST; 887517977557 BT13-10E360-5 Retrograde Femur Straight, 13.0mm x 360mm, 5/10cm ST; 887517977540 BT13-10E360-7 Retrograde Femur Straight, 13.0mm x 360mm, 7/10cm ST; 887517977533 BT13-10E380-5 Retrograde Femur Straight, 13.0mm x 380mm, 5/10cm ST; 887517977526 BT13-10E380-7 Retrograde Femur Straight, 13.0mm x 380mm, 7/10cm ST; 887517977519 BT13-10E400-5 Retrograde Femur Straight, 13.0mm x 400mm, 5/10cm ST; 887517977502 BT13-10E400-7 Retrograde Femur Straight, 13.0mm x 400mm, 7/10cm ST; 887517977496 BT13-10SA360-5 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 360mm, 5/10cm ST; 887517977489 BT13-10SA360-7 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 360mm, 7/10cm ST; 887517977472 BT13-10SA380-5 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 380mm, 5/10cm ST; 887517977465 BT13-10SA380-7 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 380mm, 7/10cm ST; 887517977458 BT13-10SA400-5 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 400mm, 5/10cm ST; 887517977441 BT13-10SA400-7 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 400mm, 7/10cm ST; 887517977434 BT13-10SB360-5 Recon Antegrade Femur Piriformis Straight, 13.0mm x 360mm, 5/10cm ST; 887517977427 BT13-10SB360-7 Recon Antegrade Femur Piriformis Straight, 13.0mm x 360mm, 7/10cm ST; 887517977410 BT13-10SB380-5 Recon Antegrade Femur Piriformis Straight, 13.0mm x 380mm, 5/10cm ST; 887517977403 BT13-10SB380-7 Recon Antegrade Femur Piriformis Straight, 13.0mm x 380mm, 7/10cm ST; 887517977397 BT13-10SB400-5 Recon Antegrade Femur Piriformis Straight, 13.0mm x 400mm, 5/10cm ST; 887517977380 BT13-10SB400-7 Recon Antegrade Femur Piriformis Straight, 13.0mm x 400mm, 7/10cm ST; 887517977373 BT13-10SD360-5 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 360mm, 5/10cm ST; 887517977366 BT13-10SD360-7 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 360mm, 7/10cm ST; 887517977359 BT13-10SD380-5 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 380mm, 5/10cm ST; 887517977342 BT13-10SD380-7 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 380mm, 7/10cm ST; 887517977335 BT13-10SD400-5 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 400mm, 5/10cm ST; 887517977328 BT13-10SD400-7 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 400mm, 7/10cm ST; 887517977311 BT13-10SE360-5 Retrograde Femur Straight  Three Holes, 13.0mm x 360mm, 5/10cm ST; 887517977304 BT13-10SE360-7 Retrograde Femur Straight  Three Holes, 13.0mm x 360mm, 7/10cm ST; 887517977298 BT13-10SE380-5 Retrograde Femur Straight  Three Holes, 13.0mm x 380mm, 5/10cm ST; 887517977281 BT13-10SE380-7 Retrograde Femur Straight  Three Holes, 13.0mm x 380mm, 7/10cm ST; 887517977274 BT13-10SE400-5 Retrograde Femur Straight  Three Holes, 13.0mm x 400mm, 5/10cm ST; 887517977267 BT13-10SE400-7 Retrograde Femur Straight  Three Holes, 13.0mm x 400mm, 7/10cm ST; 887517977250 BT13-10SJ360-5 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 360mm, 5/10cm ST; 887517977243 BT13-10SJ360-7 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 360mm, 7/10cm ST; 887517977236 BT13-10SJ380-5 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 380mm, 5/10cm ST; 887517977229 BT13-10SJ380-7 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 380mm, 7/10cm ST; 887517977212 BT13-10SJ400-5 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 400mm, 5/10cm ST; 887517977205 BT13-10SJ400-7 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 400mm, 7/10cm ST; 887517977199 BT13-10V360-5 Retrograde Femur 10¿ Bend, 13.0mm x 360mm, 5/10cm ST; 887517977182 BT13-10V360-7 Retrograde Femur 10¿ Bend, 13.0mm x 360mm, 7/10cm ST; 887517977175 BT13-10V380-5 Retrograde Femur 10¿ Bend, 13.0mm x 380mm, 5/10cm ST; 887517977168 BT13-10V380-7 Retrograde Femur 10¿ Bend, 13.0mm x 380mm, 7/10cm ST; 887517977151 BT13-10V400-5 Retrograde Femur 10¿ Bend, 13.0mm x 400mm, 5/10cm ST; 887517977144 BT13-10V400-7 Retrograde Femur 10¿ Bend, 13.0mm x 400mm, 7/10cm ST; 887517977137 BT13-10X360-5 Retrograde Femur Straight, 13.0mm x 360mm, 5/10cm ST; 887517977120 BT13-10X360-7 Retrograde Femur Straight, 13.0mm x 360mm, 7/10cm ST; 887517977113 BT13-10X380-5 Retrograde Femur Straight, 13.0mm x 380mm, 5/10cm ST; 887517977106 BT13-10X380-7 Retrograde Femur Straight, 13.0mm x 380mm, 7/10cm ST; 887517977090 BT13-10X400-5 Retrograde Femur Straight, 13.0mm x 400mm, 5/10cm ST; 887517977083 BT13-10X400-7 Retrograde Femur Straight, 13.0mm x 400mm, 7/10cm ST; 887517977076 BT13-60A280-6 Retrograde Femur 10¿ Bend, 13.0mm x 280mm, 6/6cm ST; 887517977069 BT13-60B280-6 Antegrade Femur Piriformis Straight, 13.0mm x 280mm, 6/6cm ST; 887517977052 BT13-60BT280-6 Antegrade Femur Piriformis Straight, 13.0mm x 280mm, 6/6cm ST; 887517977045 BT13-60C280-6 Antegrade Tibia 10¿ Bend, 13.0mm x 280mm, 6/6cm ST; 887517977038 BT13-60D280-6 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 280mm, 6/6cm ST; 887517977021 BT13-60DT280-6 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 280mm, 6/6cm ST; 887517977014 BT13-60E280-6 Retrograde Femur Straight, 13.0mm x 280mm, 6/6cm ST; 887517977007 BT13-60SA280-6 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 280mm, 6/6cm ST; 887517976994 BT13-60SB280-6 Recon Antegrade Femur Piriformis Straight, 13.0mm x 280mm, 6/6cm ST; 887517976987 BT13-60SD280-6 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 280mm, 6/6cm ST; 887517976970 BT13-60SE280-6 Retrograde Femur Straight  Three Holes, 13.0mm x 280mm, 6/6cm ST; 887517976963 BT13-60SJ280-6 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 280mm, 6/6cm ST; 887517976956 BT13-60V280-6 Retrograde Femur 10¿ Bend, 13.0mm x 280mm, 6/6cm ST; 887517976949 BT13-60X280-6 Retrograde Femur Straight, 13.0mm x 280mm, 6/6cm ST; 887517976932 BT13-70A300-5 Retrograde Femur 10¿ Bend, 13.0mm x 300mm, 5/7cm ST; 887517976925 BT13-70A300-7 Retrograde Femur 10¿ Bend, 13.0mm x 300mm, 7/7cm ST; 887517976918 BT13-70B300-5 Antegrade Femur Piriformis Straight, 13.0mm x 300mm, 5/7cm ST; 887517976901 BT13-70B300-7 Antegrade Femur Piriformis Straight, 13.0mm x 300mm, 7/7cm ST; 887517976895 BT13-70BT300-5 Antegrade Femur Piriformis Straight, 13.0mm x 300mm, 5/7cm ST; 887517976888 BT13-70BT300-7 Antegrade Femur Piriformis Straight, 13.0mm x 300mm, 7/7cm ST; 887517976871 BT13-70C300-5 Antegrade Tibia 10¿ Bend, 13.0mm x 300mm, 5/7cm ST; 887517976864 BT13-70C300-7 Antegrade Tibia 10¿ Bend, 13.0mm x 300mm, 7/7cm ST; 887517976857 BT13-70D300-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 300mm, 5/7cm ST; 887517976840 BT13-70D300-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 300mm, 7/7cm ST; 887517976833 BT13-70DT300-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 300mm, 5/7cm ST; 887517976826 BT13-70DT300-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 300mm, 7/7cm ST; 887517976819 BT13-70E300-5 Retrograde Femur Straight, 13.0mm x 300mm, 5/7cm ST; 887517976802 BT13-70E300-7 Retrograde Femur Straight, 13.0mm x 300mm, 7/7cm ST; 887517976796 BT13-70SA300-5 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 300mm, 5/7cm ST; 887517976789 BT13-70SA300-7 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 300mm, 7/7cm ST; 887517976772 BT13-70SB300-5 Recon Antegrade Femur Piriformis Straight, 13.0mm x 300mm, 5/7cm ST; 887517976765 BT13-70SB300-7 Recon Antegrade Femur Piriformis Straight, 13.0mm x 300mm, 7/7cm ST; 887517976758 BT13-70SD300-5 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 300mm, 5/7cm ST; 887517976741 BT13-70SD300-7 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 300mm, 7/7cm ST; 887517976734 BT13-70SE300-5 Retrograde Femur Straight  Three Holes, 13.0mm x 300mm, 5/7cm ST; 887517976727 BT13-70SE300-7 Retrograde Femur Straight  Three Holes, 13.0mm x 300mm, 7/7cm ST; 887517976710 BT13-70SJ300-5 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 300mm, 5/7cm ST; 887517976703 BT13-70SJ300-7 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 300mm, 7/7cm ST; 887517976697 BT13-70V300-5 Retrograde Femur 10¿ Bend, 13.0mm x 300mm, 5/7cm ST; 887517976680 BT13-70V300-7 Retrograde Femur 10¿ Bend, 13.0mm x 300mm, 7/7cm ST; 887517976673 BT13-70X300-5 Retrograde Femur Straight, 13.0mm x 300mm, 5/7cm ST; 887517976666 BT13-70X300-7 Retrograde Femur Straight, 13.0mm x 300mm, 7/7cm ST; 887517976659 BT13-80A320-5 Retrograde Femur 10¿ Bend, 13.0mm x 320mm, 5/8cm ST; 887517976642 BT13-80A320-7 Retrograde Femur 10¿ Bend, 13.0mm x 320mm, 7/8cm ST; 887517976635 BT13-80B320-5 Antegrade Femur Piriformis Straight, 13.0mm x 320mm, 5/8cm ST; 887517976628 BT13-80B320-7 Antegrade Femur Piriformis Straight, 13.0mm x 320mm, 7/8cm ST; 887517976611 BT13-80BT320-5 Antegrade Femur Piriformis Straight, 13.0mm x 320mm, 5/8cm ST; 887517976604 BT13-80BT320-7 Antegrade Femur Piriformis Straight, 13.0mm x 320mm, 7/8cm ST; 887517976598 BT13-80C320-5 Antegrade Tibia 10¿ Bend, 13.0mm x 320mm, 5/8cm ST; 887517976581 BT13-80C320-7 Antegrade Tibia 10¿ Bend, 13.0mm x 320mm, 7/8cm ST; 887517976574 BT13-80D320-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 320mm, 5/8cm ST; 887517976567 BT13-80D320-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 320mm, 7/8cm ST; 887517976550 BT13-80DT320-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 320mm, 5/8cm ST; 887517976543 BT13-80DT320-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 320mm, 7/8cm ST; 887517976536 BT13-80E320-5 Retrograde Femur Straight, 13.0mm x 320mm, 5/8cm ST; 887517976529 BT13-80E320-7 Retrograde Femur Straight, 13.0mm x 320mm, 7/8cm ST; 887517976512 BT13-80SA320-5 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 320mm, 5/8cm ST; 887517976505 BT13-80SA320-7 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 320mm, 7/8cm ST; 887517976499 BT13-80SB320-5 Recon Antegrade Femur Piriformis Straight, 13.0mm x 320mm, 5/8cm ST; 887517976482 BT13-80SB320-7 Recon Antegrade Femur Piriformis Straight, 13.0mm x 320mm, 7/8cm ST; 887517976475 BT13-80SD320-5 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 320mm, 5/8cm ST; 887517976468 BT13-80SD320-7 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 320mm, 7/8cm ST; 887517976451 BT13-80SE320-5 Retrograde Femur Straight  Three Holes, 13.0mm x 320mm, 5/8cm ST; 887517976444 BT13-80SE320-7 Retrograde Femur Straight  Three Holes, 13.0mm x 320mm, 7/8cm ST; 887517976437 BT13-80SJ320-5 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 320mm, 5/8cm ST; 887517976420 BT13-80SJ320-7 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 320mm, 7/8cm ST; 887517976413 BT13-80V320-5 Retrograde Femur 10¿ Bend, 13.0mm x 320mm, 5/8cm ST; 887517976406 BT13-80V320-7 Retrograde Femur 10¿ Bend, 13.0mm x 320mm, 7/8cm ST; 887517976390 BT13-80X320-5 Retrograde Femur Straight, 13.0mm x 320mm, 5/8cm ST; 887517976383 BT13-80X320-7 Retrograde Femur Straight, 13.0mm x 320mm, 7/8cm ST; 887517976376 BT13-90A340-5 Retrograde Femur 10¿ Bend, 13.0mm x 340mm, 5/9cm ST; 887517976369 BT13-90A340-7 Retrograde Femur 10¿ Bend, 13.0mm x 340mm, 7/9cm ST; 887517976352 BT13-90B340-5 Antegrade Femur Piriformis Straight, 13.0mm x 340mm, 5/9cm ST; 887517976345 BT13-90B340-7 Antegrade Femur Piriformis Straight, 13.0mm x 340mm, 7/9cm ST; 887517976338 BT13-90BT340-5 Antegrade Femur Piriformis Straight, 13.0mm x 340mm, 5/9cm ST; 887517976321 BT13-90BT340-7 Antegrade Femur Piriformis Straight, 13.0mm x 340mm, 7/9cm ST; 887517976314 BT13-90C340-5 Antegrade Tibia 10¿ Bend, 13.0mm x 340mm, 5/9cm ST; 887517976307 BT13-90C340-7 Antegrade Tibia 10¿ Bend, 13.0mm x 340mm, 7/9cm ST; 887517976291 BT13-90D340-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 340mm, 5/9cm ST; 887517976284 BT13-90D340-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 340mm, 7/9cm ST; 887517976277 BT13-90DT340-5 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 340mm, 5/9cm ST; 887517976260 BT13-90DT340-7 Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 340mm, 7/9cm ST; 887517976253 BT13-90E340-5 Retrograde Femur Straight, 13.0mm x 340mm, 5/9cm ST; 887517976246 BT13-90E340-7 Retrograde Femur Straight, 13.0mm x 340mm, 7/9cm ST; 887517976239 BT13-90SA340-5 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 340mm, 5/9cm ST; 887517976222 BT13-90SA340-7 Retrograde Femur 5¿ Bend  Three Holes, 13.0mm x 340mm, 7/9cm ST; 887517976215 BT13-90SB340-5 Recon Antegrade Femur Piriformis Straight, 13.0mm x 340mm, 5/9cm ST; 887517976208 BT13-90SB340-7 Recon Antegrade Femur Piriformis Straight, 13.0mm x 340mm, 7/9cm ST; 887517976192 BT13-90SD340-5 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 340mm, 5/9cm ST; 887517976185 BT13-90SD340-7 Recon Antegrade Femur Trochanter 10¿ Bend, 13.0mm x 340mm, 7/9cm ST; 887517976178 BT13-90SE340-5 Retrograde Femur Straight  Three Holes, 13.0mm x 340mm, 5/9cm ST; 887517976161 BT13-90SE340-7 Retrograde Femur Straight  Three Holes, 13.0mm x 340mm, 7/9cm ST; 887517976154 BT13-90SJ340-5 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 340mm, 5/9cm ST; 887517976147 BT13-90SJ340-7 Antegrade Tibia 10¿ Bend  Three Holes, 13.0mm x 340mm, 7/9cm ST; 887517976130 BT13-90V340-5 Retrograde Femur 10¿ Bend, 13.0mm x 340mm, 5/9cm ST; 887517976123 BT13-90V340-7 Retrograde Femur 10¿ Bend, 13.0mm x 340mm, 7/9cm ST; 887517976116 BT13-90X340-5 Retrograde Femur Straight, 13.0mm x 340mm, 5/9cm ST; 887517976109 BT13-90X340-7 Retrograde Femur Straight, 13.0mm x 340mm, 7/9cm ST; 00887517021915 LP4.5-DLF30 PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, 0-Hole; 00887517021908 LP4.5-DLF31 PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, 1-Hole; 00887517021892 LP4.5-DLF32 PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, 2-Hole; 00887517021885 LP4.5-DLF3B PRECICE Plate 4.5mm Distal Femur Plate, Left, 35mm, Bump; 00887517021878 LP4.5-DLF40 PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, 0-Hole; 00887517021861 LP4.5-DLF41 PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, 1-Hole; 00887517021854 LP4.5-DLF42 PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, 2-Hole; 00887517021847 LP4.5-DLF4B PRECICE Plate 4.5mm Distal Femur Plate, Left, 45mm, Bump; 00887517021830 LP4.5-DRF30 PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, 0-Hole; 00887517021823 LP4.5-DRF31 PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, 1-Hole; 00887517021816 LP4.5-DRF32 PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, 2-Hole; 00887517021809 LP4.5-DRF3B PRECICE Plate 4.5mm Distal Femur Plate, Right, 35mm, Bump; 00887517021793 LP4.5-DRF40 PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, 0-Hole; 00887517021786 LP4.5-DRF41 PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, 1-Hole; 00887517021779 LP4.5-DRF42 PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, 2-Hole; 00887517021762 LP4.5-DRF4B PRECICE Plate 4.5mm Distal Femur Plate, Right, 45mm, Bump; 00887517021755 LP4.5-MLT30 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, 0-Hole; 00887517021748 LP4.5-MLT31 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, 1-Hole; 00887517021731 LP4.5-MLT32 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, 2-Hole; 00887517021724 LP4.5-MLT3B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 35mm, Bump; 00887517021717 LP4.5-MLT40 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, 0-Hole; 00887517021700 LP4.5-MLT41 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, 1-Hole; 00887517021694 LP4.5-MLT42 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, 2-Hole; 00887517021687 LP4.5-MLT4B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, Bump; 00887517021670 LP4.5-MRT30 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, 0-Hole; 00887517021663 LP4.5-MRT31 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, 1-Hole; 00887517021656 LP4.5-MRT32 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, 2-Hole; 00887517021649 LP4.5-MRT3B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 35mm, Bump; 00887517021632 LP4.5-MRT40 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, 0-Hole; 00887517021625 LP4.5-MRT41 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, 1-Hole; 00887517021618 LP4.5-MRT42 PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, 2-Hole; 00887517021601 LP4.5-MRT4B PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Right, 45mm, Bump; 00887517021595 LP4.5-PLF30 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, 0-Hole; 00887517021588 LP4.5-PLF31 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, 1-Hole; 00887517021571 LP4.5-PLF32 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, 2-Hole; 00887517021564 LP4.5-PLF3B PRECICE Plate 4.5mm Proximal Femur Plate, Left, 35mm, Bump; 00887517021557 LP4.5-PLF40 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, 0-Hole; 00887517021540 LP4.5-PLF41 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, 1-Hole; 00887517021533 LP4.5-PLF42 PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, 2-Hole; 00887517021526 LP4.5-PLF4B PRECICE Plate 4.5mm Proximal Femur Plate, Left, 45mm, Bump; 00887517021519 LP4.5-PLT30 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, 0-Hole; 00887517021502 LP4.5-PLT31 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, 1-Hole; 00887517021496 LP4.5-PLT32 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, 2-Hole; 00887517021489 LP4.5-PLT3B PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 35mm, Bump; 00887517021472 LP4.5-PLT40 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, 0-Hole; 00887517021465 LP4.5-PLT41 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, 1-Hole; 00887517021458 LP4.5-PLT42 PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, 2-Hole; 00887517021441 LP4.5-PLT4B PRECICE Plate 4.5mm Proximal Tibia Plate, Left, 45mm, Bump; 00887517021434 LP4.5-PRF30 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, 0-Hole; 00887517021427 LP4.5-PRF31 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, 1-Hole; 00887517021410 LP4.5-PRF32 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, 2-Hole; 00887517021403 LP4.5-PRF3B PRECICE Plate 4.5mm Proximal Femur Plate, Right, 35mm, Bump; 00887517021397 LP4.5-PRF40 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, 0-Hole; 00887517021380 LP4.5-PRF41 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, 1-Hole; 00887517021373 LP4.5-PRF42 PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, 2-Hole; 00887517021366 LP4.5-PRF4B PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, Bump; 00887517021359 LP4.5-PRT30 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, 0-Hole; 00887517021342 LP4.5-PRT31 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, 1-Hole; 00887517021335 LP4.5-PRT32 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, 2-Hole; 00887517021328 LP4.5-PRT3B PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 35mm, Bump; 00887517021311 LP4.5-PRT40 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, 0-Hole; 00887517021304 LP4.5-PRT41 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, 1-Hole; 00887517021298 LP4.5-PRT42 PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, 2-Hole; 00887517021281 LP4.5-PRT4B PRECICE Plate 4.5mm Proximal Tibia Plate, Right, 45mm, Bump; 887517950604 PS10.0-50A235 Retrograde Femur 10¿ Bend; 887517950741 PS10.0-50B235 Antegrade Femur Piriformis Straight; 887517950956 PS10.0-50C235 Antegrade Tibia 10¿ Bend; 887517951168 PS10.0-50D235 Antegrade Femur Trochanter 10¿ Bend; 887517951373 PS10.0-50E235 Retrograde Femur Straight; 887517951595 PS10.0-50SJ235 Antegrade Tibia 10¿ Bend; 887517951793 PS10.0-50V235 Retrograde Femur 10¿ Bend; 887517952004 PS10.0-50X235 Retrograde Femur Straight; 887517950543 PS10.0-65A250 Retrograde Femur 10¿ Bend; 887517950758 PS10.0-65B250 Antegrade Femur Piriformis Straight; 887517950963 PS10.0-65C250 Antegrade Tibia 10¿ Bend; 887517951175 PS10.0-65D250 Antegrade Femur Trochanter 10¿ Bend; 887517951380 PS10.0-65E250 Retrograde Femur Straight; 887517951588 PS10.0-65SJ250 Antegrade Tibia 10¿ Bend; 887517951809 PS10.0-65V250 Retrograde Femur 10¿ Bend; 887517952011 PS10.0-65X250 Retrograde Femur Straight; 887517950550 PS10.0-80A265 Retrograde Femur 10¿ Bend; 887517950567 PS10.0-80A280 Retrograde Femur 10¿ Bend; 887517950574 PS10.0-80A305 Retrograde Femur 10¿ Bend; 887517950581 PS10.0-80A335 Retrograde Femur 10¿ Bend; 887517950598 PS10.0-80A365 Retrograde Femur 10¿ Bend; 887517950765 PS10.0-80B265 Antegrade Femur Piriformis Straight; 887517950772 PS10.0-80B280 Antegrade Femur Piriformis Straight; 887517950789 PS10.0-80B305 Antegrade Femur Piriformis Straight; 887517950796 PS10.0-80B335 Antegrade Femur Piriformis Straight; 887517950802 PS10.0-80B365 Antegrade Femur Piriformis Straight; 887517950970 PS10.0-80C265 Antegrade Tibia 10¿ Bend; 887517950987 PS10.0-80C280 Antegrade Tibia 10¿ Bend; 887517950994 PS10.0-80C305 Antegrade Tibia 10¿ Bend; 887517951007 PS10.0-80C335 Antegrade Tibia 10¿ Bend; 887517951014 PS10.0-80C365 Antegrade Tibia 10¿ Bend; 887517951182 PS10.0-80D265 Antegrade Femur Trochanter 10¿ Bend; 887517951199 PS10.0-80D280 Antegrade Femur Trochanter 10¿ Bend; 887517951205 PS10.0-80D305 Antegrade Femur Trochanter 10¿ Bend; 887517951212 PS10.0-80D335 Antegrade Femur Trochanter 10¿ Bend; 887517951229 PS10.0-80D365 Antegrade Femur Trochanter 10¿ Bend; 887517951397 PS10.0-80E265 Retrograde Femur Straight; 887517951403 PS10.0-80E280 Retrograde Femur Straight; 887517951410 PS10.0-80E305 Retrograde Femur Straight; 887517951427 PS10.0-80E335 Retrograde Femur Straight; 887517951434 PS10.0-80E365 Retrograde Femur Straight; 887517951601 PS10.0-80SJ265 Antegrade Tibia 10¿ Bend; 887517951618 PS10.0-80SJ280 Antegrade Tibia 10¿ Bend; 887517951625 PS10.0-80SJ305 Antegrade Tibia 10¿ Bend; 887517951632 PS10.0-80SJ335 Antegrade Tibia 10¿ Bend; 887517951649 PS10.0-80SJ365 Antegrade Tibia 10¿ Bend; 887517951816 PS10.0-80V265 Retrograde Femur 10¿ Bend; 887517951823 PS10.0-80V280 Retrograde Femur 10¿ Bend; 887517951830 PS10.0-80V305 Retrograde Femur 10¿ Bend; 887517951847 PS10.0-80V335 Retrograde Femur 10¿ Bend; 887517951854 PS10.0-80V365 Retrograde Femur 10¿ Bend; 887517952028 PS10.0-80X265 Retrograde Femur Straight; 887517952035 PS10.0-80X280 Retrograde Femur Straight; 887517952042 PS10.0-80X305 Retrograde Femur Straight; 887517952059 PS10.0-80X335 Retrograde Femur Straight; 887517952066 PS10.0-80X365 Retrograde Femur Straight; 887517950819 PS11.5-50B235 Antegrade Femur Piriformis Straight; 887517951021 PS11.5-50C235 Antegrade Tibia 10¿ Bend; 887517951236 PS11.5-50D235 Antegrade Femur Trochanter 10¿ Bend; 887517951441 PS11.5-50E235 Retrograde Femur Straight; 887517951656 PS11.5-50SJ235 Antegrade Tibia 10¿ Bend; 887517951861 PS11.5-50V235 Retrograde Femur 10¿ Bend; 887517952073 PS11.5-50X235 Retrograde Femur Straight; 887517950611 PS11.5-65A250 Retrograde Femur 10¿ Bend; 887517950826 PS11.5-65B250 Antegrade Femur Piriformis Straight; 887517951038 PS11.5-65C250 Antegrade Tibia 10¿ Bend; 887517951243 PS11.5-65D250 Antegrade Femur Trochanter 10¿ Bend; 887517951458 PS11.5-65E250 Retrograde Femur Straight; 887517951663 PS11.5-65SJ250 Antegrade Tibia 10¿ Bend; 887517951878 PS11.5-65V250 Retrograde Femur 10¿ Bend; 887517952080 PS11.5-65X250 Retrograde Femur Straight; 887517950628 PS11.5-80A265 Retrograde Femur 10¿ Bend; 887517950635 PS11.5-80A280 Retrograde Femur 10¿ Bend; 887517950642 PS11.5-80A305 Retrograde Femur 10¿ Bend; 887517950659 PS11.5-80A335 Retrograde Femur 10¿ Bend; 887517950666 PS11.5-80A365 Retrograde Femur 10¿ Bend; 887517950833 PS11.5-80B265 Antegrade Femur Piriformis Straight; 887517950840 PS11.5-80B280 Antegrade Femur Piriformis Straight; 887517950857 PS11.5-80B305 Antegrade Femur Piriformis Straight; 887517950864 PS11.5-80B335 Antegrade Femur Piriformis Straight; 887517950871 PS11.5-80B365 Antegrade Femur Piriformis Straight; 887517951045 PS11.5-80C265 Antegrade Tibia 10¿ Bend; 887517951052 PS11.5-80C280 Antegrade Tibia 10¿ Bend; 887517951069 PS11.5-80C305 Antegrade Tibia 10¿ Bend; 887517951076 PS11.5-80C335 Antegrade Tibia 10¿ Bend; 887517951083 PS11.5-80C365 Antegrade Tibia 10¿ Bend; 887517951250 PS11.5-80D265 Antegrade Femur Trochanter 10¿ Bend; 887517951267 PS11.5-80D280 Antegrade Femur Trochanter 10¿ Bend; 887517951274 PS11.5-80D305 Antegrade Femur Trochanter 10¿ Bend; 887517951281 PS11.5-80D335 Antegrade Femur Trochanter 10¿ Bend; 887517951298 PS11.5-80D365 Antegrade Femur Trochanter 10¿ Bend; 887517951465 PS11.5-80E265 Retrograde Femur Straight; 887517951472 PS11.5-80E280 Retrograde Femur Straight; 887517951489 PS11.5-80E305 Retrograde Femur Straight; 887517951496 PS11.5-80E335 Retrograde Femur Straight; 887517951502 PS11.5-80E365 Retrograde Femur Straight; 887517951670 PS11.5-80SJ265 Antegrade Tibia 10¿ Bend; 887517951687 PS11.5-80SJ280 Antegrade Tibia 10¿ Bend; 887517951694 PS11.5-80SJ305 Antegrade Tibia 10¿ Bend; 887517951700 PS11.5-80SJ335 Antegrade Tibia 10¿ Bend; 887517951717 PS11.5-80SJ365 Antegrade Tibia 10¿ Bend; 887517951885 PS11.5-80V265 Retrograde Femur 10¿ Bend; 887517951892 PS11.5-80V280 Retrograde Femur 10¿ Bend; 887517951908 PS11.5-80V305 Retrograde Femur 10¿ Bend; 887517951915 PS11.5-80V335 Retrograde Femur 10¿ Bend; 887517951922 PS11.5-80V365 Retrograde Femur 10¿ Bend; 887517952097 PS11.5-80X265 Retrograde Femur Straight; 887517952103 PS11.5-80X280 Retrograde Femur Straight; 887517952110 PS11.5-80X305 Retrograde Femur Straight; 887517952127 PS11.5-80X335 Retrograde Femur Straight; 887517952134 PS11.5-80X365 Retrograde Femur Straight; 887517950673 PS13.0-50A235 Retrograde Femur 10¿ Bend; 887517950888 PS13.0-50B235 Antegrade Femur Piriformis Straight; 887517951090 PS13.0-50C235 Antegrade Tibia 10¿ Bend; 887517951304 PS13.0-50D235 Antegrade Femur Trochanter 10¿ Bend; 887517951519 PS13.0-50E235 Retrograde Femur Straight; 887517951724 PS13.0-50SJ235 Antegrade Tibia 10¿ Bend; 887517951939 PS13.0-50V235 Retrograde Femur 10¿ Bend; 887517952141 PS13.0-50X235 Retrograde Femur Straight; 887517950680 PS13.0-65A250 Retrograde Femur 10¿ Bend; 887517950895 PS13.0-65B250 Antegrade Femur Piriformis Straight; 887517951106 PS13.0-65C250 Antegrade Tibia 10¿ Bend; 887517951311 PS13.0-65D250 Antegrade Femur Trochanter 10¿ Bend; 887517951526 PS13.0-65E250 Retrograde Femur Straight; 887517951731 PS13.0-65SJ250 Antegrade Tibia 10¿ Bend; 887517951946 PS13.0-65V250 Retrograde Femur 10¿ Bend; 887517952158 PS13.0-65X250 Retrograde Femur Straight; 887517950697 PS13.0-80A265 Retrograde Femur 10¿ Bend; 887517950703 PS13.0-80A280 Retrograde Femur 10¿ Bend; 887517950710 PS13.0-80A305 Retrograde Femur 10¿ Bend; 887517950727 PS13.0-80A335 Retrograde Femur 10¿ Bend; 887517950734 PS13.0-80A365 Retrograde Femur 10¿ Bend; 887517950901 PS13.0-80B265 Antegrade Femur Piriformis Straight; 887517950918 PS13.0-80B280 Antegrade Femur Piriformis Straight; 887517950925 PS13.0-80B305 Antegrade Femur Piriformis Straight; 887517950932 PS13.0-80B335 Antegrade Femur Piriformis Straight; 887517950949 PS13.0-80B365 Antegrade Femur Piriformis Straight; 887517951113 PS13.0-80C265 Antegrade Tibia 10¿ Bend; 887517951120 PS13.0-80C280 Antegrade Tibia 10¿ Bend; 887517951137 PS13.0-80C305 Antegrade Tibia 10¿ Bend; 887517951144 PS13.0-80C335 Antegrade Tibia 10¿ Bend; 887517951151 PS13.0-80C365 Antegrade Tibia 10¿ Bend; 887517951328 PS13.0-80D265 Antegrade Femur Trochanter 10¿ Bend; 887517951335 PS13.0-80D280 Antegrade Femur Trochanter 10¿ Bend; 887517951342 PS13.0-80D305 Antegrade Femur Trochanter 10¿ Bend; 887517951359 PS13.0-80D335 Antegrade Femur Trochanter 10¿ Bend; 887517951366 PS13.0-80D365 Antegrade Femur Trochanter 10¿ Bend; 8.87518E+11 PS13.0-80E265 Retrograde Femur Straight; 887517951540 PS13.0-80E280 Retrograde Femur Straight; 887517951557 PS13.0-80E305 Retrograde Femur Straight; 887517951564 PS13.0-80E335 Retrograde Femur Straight; 887517951571 PS13.0-80E365 Retrograde Femur Straight; 887517951748 PS13.0-80SJ265 Antegrade Tibia 10¿ Bend; 887517951755 PS13.0-80SJ280 Antegrade Tibia 10¿ Bend; 887517951762 PS13.0-80SJ305 Antegrade Tibia 10¿ Bend; 887517951779 PS13.0-80SJ335 Antegrade Tibia 10¿ Bend; 887517951786 PS13.0-80SJ365 Antegrade Tibia 10¿ Bend; 887517951953 PS13.0-80V265 Retrograde Femur 10¿ Bend; 887517951960 PS13.0-80V280 Retrograde Femur 10¿ Bend; 887517951977 PS13.0-80V305 Retrograde Femur 10¿ Bend; 887517951984 PS13.0-80V335 Retrograde Femur 10¿ Bend; 887517951991 PS13.0-80V365 Retrograde Femur 10¿ Bend; 887517952165 PS13.0-80X265 Retrograde Femur Straight; 887517952172 PS13.0-80X280 Retrograde Femur Straight; 887517952189 PS13.0-80X305 Retrograde Femur Straight; 887517952196 PS13.0-80X335 Retrograde Femur Straight; 887517952202 PS13.0-80X365 Retrograde Femur Straight; 887517952219 SST-010-000 END CAP 10.0/11.5mm, +0mm; 887517952226 SST-010-005 END CAP 10.0/11.5mm, +5mm; 887517952233 SST-010-010 END CAP 10.0/11.5mm, +10mm; 887517952240 SST-010-015 END CAP 10.0/11.5mm, +15mm; 887517952257 SST-010-020 END CAP 10.0/11.5mm, +20mm; 887517952264 SST-013-000 END CAP 13.0mm, +0mm; 887517952271 SST-013-005 END CAP 13.0mm, +5mm; 887517952288 SST-013-010 END CAP 13.0mm, +10mm; 887517952295 SST-013-015 END CAP 13.0mm, +15mm; 887517952301 SST-013-020 END CAP 13.0mm, +20mm; 887517989895 STP3.5-200 PBT SS Locking Screw, 3.5x20mm P.T.; 887517989901 STP3.5-225 PBT SS Locking Screw, 3.5x22.5mm P.T.; 887517989918 STP3.5-250 PBT SS Locking Screw, 3.5x25mm P.T.; 887517989925 STP3.5-275 PBT SS Locking Screw, 3.5x27.5mm P.T.; 887517989932 STP3.5-300 PBT SS Locking Screw, 3.5x30mm P.T.; 887517989949 STP3.5-325 PBT SS Locking Screw, 3.5x32.5mm P.T.; 887517989956 STP3.5-350 PBT SS Locking Screw, 3.5x35mm P.T.; 887517989963 STP3.5-375 PBT SS Locking Screw, 3.5x37.5mm P.T.; 887517989970 STP3.5-400 PBT SS Locking Screw, 3.5x40mm P.T.; 887517989987 STP3.5-425 PBT SS Locking Screw, 3.5x42.5mm P.T.; 887517989994 STP3.5-450 PBT SS Locking Screw, 3.5x45mm P.T.; 887517990006 STP3.5-475 PBT SS Locking Screw, 3.5x47.5mm P.T.; 887517990013 STP3.5-500 PBT SS Locking Screw, 3.5x50mm P.T.; 887517990020 STP3.5-550 PBT SS Locking Screw, 3.5x55mm P.T.; 887517990037 STP3.5-600 PBT SS Locking Screw, 3.5x60mm P.T.; 887517952493 STP4.5-200 4.5mm x 20mm Locking Screws, Partially T; 887517952509 STP4.5-225 4.5mm x 22.5mm Locking Screws, Partially; 887517952516 STP4.5-250 4.5mm x 25mm Locking Screws, Partially T; 887517952523 STP4.5-275 4.5mm x 27.5mm Locking Screws, Partially; 887517952530 STP4.5-300 4.5mm x 30mm Locking Screws, Partially T; 887517952547 STP4.5-325 4.5mm x 32.5mm Locking Screws, Partially; 887517952554 STP4.5-350 4.5mm x 35mm Locking Screws, Partially T; 887517952561 STP4.5-375 4.5mm x 37.5mm Locking Screws, Partially; 887517952578 STP4.5-400 4.5mm x 40mm Locking Screws, Partially T; 887517952585 STP4.5-425 4.5mm x 42.5mm Locking Screws, Partially; 8.87518E+11 STP4.5-450 4.5mm x 45mm Locking Screws, Partially T; 887517952608 STP4.5-475 4.5mm x 47.5mm Locking Screws, Partially; 887517952615 STP4.5-500 4.5mm x 50mm Locking Screws, Partially T; 887517952622 STP4.5-550 4.5mm x 55mm Locking Screws, Partially T; 887517952639 STP4.5-600 4.5mm x 60mm Locking Screws, Partially T; 887517952646 STP4.5-650 4.5mm x 65mm Locking Screws, Partially T; 887517952653 STP4.5-700 4.5mm x 70mm Locking Screws, Partially T; 887517952660 STP4.5-750 4.5mm x 75mm Locking Screws, Partially T; 887517952677 STP4.5-800 4.5mm x 80mm Locking Screws, Partially T; 887517952318 STP4-200 4.0mm x 20mm Locking Screws, Partially T; 887517952325 STP4-225 4.0mm x 22.5mm Locking Screws, Partially; 887517952332 STP4-250 4.0mm x 25mm Locking Screws, Partially T; 887517952349 STP4-275 4.0mm x 27.5mm Locking Screws, Partially; 887517952356 STP4-300 4.0mm x 30mm Locking Screws, Partially T; 887517952363 STP4-325 4.0mm x 32.5mm Locking Screws, Partially; 887517952370 STP4-350 4.0mm x 35mm Locking Screws, Partially T; 887517952387 STP4-375 4.0mm x 37.5mm Locking Screws, Partially; 887517952394 STP4-400 4.0mm x 40mm Locking Screws, Partially T; 887517952400 STP4-425 4.0mm x 42.5mm Locking Screws, Partially; 887517952417 STP4-450 4.0mm x 45mm Locking Screws, Partially T; 887517952424 STP4-475 4.0mm x 47.5mm Locking Screws, Partially; 887517952431 STP4-500 4.0mm x 50mm Locking Screws, Partially T; 887517952448 STP4-550 4.0mm x 55mm Locking Screws, Partially T; 887517952455 STP4-600 4.0mm x 60mm Locking Screws, Partially T; 887517952462 STP4-650 4.0mm x 65mm Locking Screws, Partially T; 887517952479 STP4-700 4.0mm x 70mm Locking Screws, Partially T; 887517952486 STP4-750 4.0mm x 75mm Locking Screws, Partially T; 887517952684 STP5-200 5.0mm x 20mm Locking Screws, Partially T; 887517952691 STP5-225 5.0mm x 22.5mm Locking Screws, Partially; 887517952707 STP5-250 5.0mm x 25mm Locking Screws, Partially T; 887517952714 STP5-275 5.0mm x 27.5mm Locking Screws, Partially; 887517952721 STP5-300 5.0mm x 30mm Locking Screws, Partially T; 887517952738 STP5-325 5.0mm x 32.5mm Locking Screws, Partially; 887517952745 STP5-350 5.0mm x 35mm Locking Screws, Partially T; 887517952752 STP5-375 5.0mm x 37.5mm Locking Screws, Partially; 887517952769 STP5-400 5.0mm x 40mm Locking Screws, Partially T; 887517952776 STP5-425 5.0mm x 42.5mm Locking Screws, Partially; 887517952783 STP5-450 5.0mm x 45mm Locking Screws, Partially T; 887517952790 STP5-475 5.0mm x 47.5mm Locking Screws, Partially; 887517952806 STP5-500 5.0mm x 50mm Locking Screws, Partially T; 887517952813 STP5-550 5.0mm x 55mm Locking Screws, Partially T; 887517952820 STP5-600 5.0mm x 60mm Locking Screws, Partially T; 887517952837 STP5-650 5.0mm x 65mm Locking Screws, Partially T; 887517952844 STP5-700 5.0mm x 70mm Locking Screws, Partially T; 887517952851 STP5-750 5.0mm x 75mm Locking Screws, Partially T; 887517952868 STP5-800 5.0mm x 80mm Locking Screws, Partially T; 00887517021274 LP4.5-SLF30 PRECICE Plate 4.5mm Shaft Plate, Left, ;35mm, Long Magnet;

Distribution

Worldwide distribution - U.S. Nationwide distribution and the countries of Argentina, Austria, Belgium, Canada, Chile, Colombia, Cyprus, Denmark, Finland, France, Germany, Greece, Hong Kong, Israel, Italy, Luxemburg, Netherlands, Portugal, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey and UAE. Consignee List provided only includes email addresses. Complete contact information is pending.

Quantity

2360 units