VITROS Immunodiagnostic Products TSH Reagent Pack
Enforcement
- Recall Number
- Z-1322-2018
- Event ID
- 79508
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Clinical Diagnostic Systems
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- April 18, 2018
- Initiation Date
- January 11, 2018
- Classification Date
- April 11, 2018
- Termination Date
- September 8, 2020
- Address
- 601 Lee Rd Ny, Rochester, NY, 14652-0001, United States
Description
VITROS Immunodiagnostic Products TSH Reagent Pack
VITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency of customer generated calibration curves to fall outside of the calibration quality parameter measuring the gradient of the Customer Calibration Curve between Calibrator Level 2 and Calibrator Level 3.
Lot 5470, expiry date: 24-May-2018 Lot 5430, expiry date: 01-Mar-2018 (not distributed in the US.) Lot 5178 was also affected by this issue, expired on 31-Mar-2017,
Distributors were notified via letter. Lot 5430 was not distributed in the US. Lot 5178 was also affected by this issue, but this lot expired on 31-Mar-2017 so Ortho took no action on this lot.
Lot #5470 (US=8570/OUS=537); Lot #5430 (US=0/OUS=9135); Lot #5178 (US=5178/OUS=4396)