FDA Enforcement
Class II
Ongoing
Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967
Recall: Z-1321-2024
·
Reported March 27, 2024
Enforcement
- Recall Number
- Z-1321-2024
- Event ID
- 94032
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Preat Corp
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 27, 2024
- Initiation Date
- February 9, 2024
- Classification Date
- March 20, 2024
- Address
- 2625 Skyway Dr Ste B, Santa Maria, CA, 93455-1405, United States
Description
Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967
Reason
Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.
Code Info
Model Number: 9000967 UDI-DI code: 00842092165867 Lot Numbers: 235650, 234294, 233667, 233885, 233439, and 231489
Distribution
U.S. Nationwide distribution in the states of AZ, OH and PA.
Quantity
561 units