FDA Enforcement Class II Ongoing

Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967

Recall: Z-1321-2024 · Reported March 27, 2024

Enforcement

Recall Number
Z-1321-2024
Event ID
94032
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Preat Corp
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 27, 2024
Initiation Date
February 9, 2024
Classification Date
March 20, 2024
Address
2625 Skyway Dr Ste B, Santa Maria, CA, 93455-1405, United States

Description

Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967

Reason

Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.

Code Info

Model Number: 9000967 UDI-DI code: 00842092165867 Lot Numbers: 235650, 234294, 233667, 233885, 233439, and 231489

Distribution

U.S. Nationwide distribution in the states of AZ, OH and PA.

Quantity

561 units