FDA Enforcement Class II Ongoing

I.B.S¿ 6.5-C Compression screw - diam 6.5mm lg 75mm,Part Number S65 ST175

Recall: Z-1319-2024 · Reported March 27, 2024

Enforcement

Recall Number
Z-1319-2024
Event ID
94063
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
In2Bones, SAS
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
March 27, 2024
Initiation Date
January 26, 2024
Classification Date
March 20, 2024
Address
28 Chemin du petit Bois, Ecully, N/A, France

Description

I.B.S¿ 6.5-C Compression screw - diam 6.5mm lg 75mm,Part Number S65 ST175

Reason

This Field Action is being conducted following the identification of a batch mix-up between I.B.S Compression screws diameter 6.5mm length 75mm and I.B.S Compression screws diameter 6.5mm length 80mm. The screws of the batches concerned by the recall may have different lengths from those issued on their labels.

Code Info

UDI/DI 3760225710685, Batch Number 2306007, exp. 31/MAY/2028

Distribution

US Nationwide distribution in the state of Tennessee.

Quantity

59 units