FDA Enforcement Class II Ongoing

SureSigns VS3 NBP, SpO2, Product Number 863071. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).

Recall: Z-1319-2019 · Reported May 22, 2019

Enforcement

Recall Number
Z-1319-2019
Event ID
82684
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips North America, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 22, 2019
Initiation Date
April 19, 2019
Classification Date
May 13, 2019
Address
3000 Minuteman Rd, Andover, MA, 01810-1032, United States

Description

SureSigns VS3 NBP, SpO2, Product Number 863071. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).

Reason

System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.

Code Info

Serial Numbers: All Including UDIs: (01)00884838028890(21) and (01)00884838087075(21) followed by serial number.

Distribution

Worldwide distribution. US Nationwide including Puerto Rico, Andorra, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Dominican Republic, Egypt, France, French Guiana, French Polynesia, Germany, Gibraltar, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea Republic, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Martinique, Mexico, Morocco, Mozambique, Myanmar, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Philippines, Portugal, Qatar, Reunion, Romania, Rwanda, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Kingdom, United Arab Emirates,Venezuela, Vietnam, and Zimbabwe.

Quantity

12543