FDA Enforcement Class II Terminated

Progressa Bed - AC-powered adjustable hospital bed

Recall: Z-1319-2018 · Reported April 18, 2018

Enforcement

Recall Number
Z-1319-2018
Event ID
79516
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hill-Rom, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 18, 2018
Initiation Date
January 12, 2018
Classification Date
April 9, 2018
Termination Date
June 26, 2020
Address
1069 State Road 46 E, Batesville, IN, 47006-7520, United States

Description

Progressa Bed - AC-powered adjustable hospital bed

Reason

A component used within the braking system may break resulting in one or more brake pedals on the bed staying in brake or not being engaged which could result in unwanted bed movement and possible patient or user injury.

Code Info

S159AW9192 through S305AW2321 Beds manufactured from June 8, 2017 to November 1, 2017

Distribution

Worldwide distribution. US nationwide, Canada, Chile, Colombia, Costa Rica, France, Hong Kong, Israel, Jamaica, Japan, Jordan, Malaysia, Mexico, Oman, Panama, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, UAE, and UK.

Quantity

2117