FDA Enforcement Class II Ongoing

Spectral CT 7500 on Rails; Product Code (REF): 728334;

Recall: Z-1318-2026 · Reported February 18, 2026

Enforcement

Recall Number
Z-1318-2026
Event ID
98248
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips North America Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 18, 2026
Initiation Date
January 7, 2026
Classification Date
February 6, 2026
Address
222 Jacobs St, N/A, Cambridge, MA, 02141-2289, United States

Description

Spectral CT 7500 on Rails; Product Code (REF): 728334;

Reason

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

Code Info

Product Code (REF): 728334; UDI-DI: 00884838103627; Serial Numbers: 1005, 1007, 1008;

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Macao, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkmenistan, United Kingdom, Uzbekistan, Vietnam.

Quantity

3 units