3M" Steri-Vac" Gas Sterilizer/Aerators, Model #'s 4XL,5XL, 8XL The 3M" Steri-Vac" Gas Sterilizer/Aerator is a compact unit designed to sterilize heat- and/or moisture-sensitive devices. This gas sterilizer/aerator is intended for INDOOR USE ONLY.
Enforcement
- Recall Number
- Z-1318-2014
- Event ID
- 67716
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- 3M Company - Health Care Business
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 9, 2014
- Initiation Date
- March 17, 2014
- Classification Date
- April 2, 2014
- Termination Date
- January 22, 2015
- Address
- 3M Center 2510 Conway Ave , B# 275-5-W-6, N/A, Saint Paul, MN, 55144, United States
Description
3M" Steri-Vac" Gas Sterilizer/Aerators, Model #'s 4XL,5XL, 8XL The 3M" Steri-Vac" Gas Sterilizer/Aerator is a compact unit designed to sterilize heat- and/or moisture-sensitive devices. This gas sterilizer/aerator is intended for INDOOR USE ONLY.
Recently, 3M received a single report of a potential exposure to Ethylene Oxide (EO) at a hospital installation involving 3M" Steri-Vac" Gas Sterilizers. The cause was traced to a rare situation involving a failure of the EO cartridge puncture assembly. This assembly was only in production during the period November 10, 2006 through December 14, 2007 and in service during the period February 7, 20
Model SKU Starting S/N Ending S/N 5XL 70-2007-2303-2 721115 721309 5XL 70-2007-2304-0 820174 820201 5XL 70-2007-2575-5 780066 780102 8XL 70-2006-8493-7 350136 350268 8XL 70-2006-8498-5 450016 450027 4XL 523122, 522205, 522502, 522287, 522655.
Worldwide Distribution - USA (nationwide) Distribution including the states of AL, AZ, CA,CO, FL, GA, HI, IL, IN, IA, KS, KY, MD, MA, MI, MN, MS, NE, NJ, NY, NC, ND, OH, OK,OA, SC, TX, WA, WV and WI. and the countries of COLOMBIA, VENEZUELA, UAE, THAILAND, GERMANY, CHINA, KOREA, CANADA, JAPAN, RUSSIA, INDIA, POLAND, PERU, AUSTRIA, ISRAEL, INDONESIA, SPAIN, CHILE, MALAYSIA, SWITZERLAND, EL SALVADOR, ECUADOR, LEBANON, MEXICO, ARGENTINA, BULGARIA, TAIWAN, TURKEY,ROMANIA, GUATEMALA and NEW ZEALAND.
73 US, 346 OUS