FDA Enforcement Class I Ongoing

Palindrome H Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.

Recall: Z-1317-2022 · Reported July 13, 2022

Enforcement

Recall Number
Z-1317-2022
Event ID
90311
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Covidien Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 13, 2022
Initiation Date
June 8, 2022
Classification Date
July 7, 2022
Address
15 Hampshire St, N/A, Mansfield, MA, 02048-1113, United States

Description

Palindrome H Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.

Reason

There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.

Code Info

Code Information organized as "Model No., Serial No., GTIN": 8888119365, 1925200097, 20884521012810; 8888119371, 2031800117, 20884521056562; 8888123405, 2028200168, 20884521012940; 8888123405, 2127200428, 20884521012940; 8888145043, 1932200082, 10884521013223; 8888145043, 2003700042, 10884521013223; 8888145043, 2014900147, 10884521013223; 8888145043, 2023300169, 10884521013223; 8888145043, 2031800113, 10884521013223; 8888145043, 2034400185, 10884521013223; 8888145043, 2104600104, 10884521013223; 8888145043, 2108100054, 10884521013223; 8888145043, 2127200539, 10884521013223; 8888145043, 1919700110, 20884521013220; 8888145043, 1929500076, 20884521013220; 8888145043, 1932200082, 20884521013220; 8888145043, 2000700103, 20884521013220; 8888145043, 2003700042, 20884521013220; 8888145043, 2014900147, 20884521013220; 8888145043, 2019200001, 20884521013220; 8888145043, 2023300169, 20884521013220; 8888145043, 2028200174, 20884521013220; 8888145043, 2031800112, 20884521013220; 8888145043, 2031800113, 20884521013220; 8888145043, 2034400185, 20884521013220; 8888145043, 2034400187, 20884521013220; 8888145043, 2104600100, 20884521013220; 8888145043, 2104600104, 20884521013220; 8888145043, 2108100054, 20884521013220; 8888145043, 2108100055, 20884521013220; 8888145043, 2108100056, 20884521013220; 8888145043, 2109700070, 20884521013220; 8888145043, 2116500163, 20884521013220; 8888145043, 2124500119, 20884521013220; 8888145043, 2127200539, 20884521013220; 8888145043, 2127200540, 20884521013220; 8888145043, 2127200541, 20884521013220; 8888145044, 1922500109, 10884521013247; 8888145044, 1922500112, 10884521013247; 8888145044, 1925200063, 10884521013247; 8888145044, 1933700119, 10884521013247; 8888145044, 2000700104, 10884521013247; 8888145044, 2104600101, 10884521013247; 8888145044, 2106900137, 10884521013247; 8888145044, 2110900104, 10884521013247; 8888145044, 2113400084, 10884521013247; 8888145044, 2116200117, 10884521013247; 8888145044, 2200300118, 10884521013247; 8888145044, 1922500109, 20884521013244; 8888145044, 1922500110, 20884521013244; 8888145044, 1922500112, 20884521013244; 8888145044, 1925200063, 20884521013244; 8888145044, 1929500075, 20884521013244; 8888145044, 1929500092, 20884521013244; 8888145044, 1933700119, 20884521013244; 8888145044, 2000700104, 20884521013244; 8888145044, 2104600101, 20884521013244; 8888145044, 2104600102, 20884521013244; 8888145044, 2104900109, 20884521013244; 8888145044, 2106900137, 20884521013244; 8888145044, 2110900104, 20884521013244; 8888145044, 2113300436, 20884521013244; 8888145044, 2113400084, 20884521013244; 8888145044, 2116200115, 20884521013244; 8888145044, 2116200116, 20884521013244; 8888145044, 2116200117, 20884521013244; 8888145044, 2124500106, 20884521013244; 8888145044, 2124500121, 20884521013244; 8888145044, 2200300118, 20884521013244; 8888145045, 1929500093, 10884521017900; 8888145045, 1933700120, 10884521017900; 8888145045, 2004400084, 10884521017900; 8888145045, 2110900143, 10884521017900; 8888145045, 2113400059, 10884521017900; 8888145045, 1822000073, 20884521017907; 8888145045, 1929500093, 20884521017907; 8888145045, 1933700120, 20884521017907; 8888145045, 2004400084, 20884521017907; 8888145045, 2110900143, 20884521017907; 8888145045, 2113400059, 20884521017907; 8888145045, 2124500087, 20884521017907; 8888145046, 1927500059, 20884521013268; 8888145046, 1927500060, 20884521013268; 8888145069, 2017600095, 20884521056708; 8888145070, 2026000154, 20884521056715; 8888145043C, 1924600101, 10884521056305; 8888145043C, 1926100282, 10884521056305; 8888145043C, 2002900150, 10884521056305; 8888145043C, 2014900156, 10884521056305; 8888145043C, 2017500125, 10884521056305; 8888145043C, 2017500128, 10884521056305; 8888145043C, 2017600093, 10884521056305; 8888145043C, 2026000077, 10884521056305; 8888145043C, 2026000079, 10884521056305; 8888145043C, 2026000080, 10884521056305; 8888145043C, 2026000081, 10884521056305; 8888145043C, 2116200114, 10884521056305; 8888145043C, 2119400127, 10884521056305; 8888145043C, 2125800084, 10884521056305; 8888145043C, 2125800085, 10884521056305; 8888145043C, 2125800086, 10884521056305; 8888145043C, 2125800087, 10884521056305; 8888145043C, 1924600101, 20884521056302; 8888145043C, 1926100282, 20884521056302; 8888145043C, 2002900150, 20884521056302; 8888145043C, 2014900156, 20884521056302; 8888145043C, 2017500125, 20884521056302; 8888145043C, 2017500128, 20884521056302; 8888145043C, 2017600093, 20884521056302; 8888145043C, 2026000077, 20884521056302; 8888145043C, 2026000079, 20884521056302; 8888145043C, 2026000080, 20884521056302; 8888145043C, 2026000081, 20884521056302; 8888145043C, 2116200114, 20884521056302; 8888145043C, 2119400127, 20884521056302; 8888145043C, 2125800086, 20884521056302; 8888145043C, 2125800087, 20884521056302; 8888145043C, 2125800088, 20884521056302; 8888145044C, 1924600102, 10884521056312; 8888145044C, 1926100278, 10884521056312; 8888145044C, 1926100283, 10884521056312; 8888145044C, 1926100284, 10884521056312; 8888145044C, 1928100093, 10884521056312; 8888145044C, 2014900157, 10884521056312; 8888145044C, 2017600094, 10884521056312; 8888145044C, 2023300171, 10884521056312; 8888145044C, 2026000082, 10884521056312; 8888145044C, 2026000083, 10884521056312; 8888145044C, 2026000084, 10884521056312; 8888145044C, 2026000085, 10884521056312; 8888145044C, 2026000086, 10884521056312; 8888145044C, 2026000153, 10884521056312; 8888145044C, 2106900135, 10884521056312; 8888145044C, 2106900213, 10884521056312; 8888145044C, 2108100058, 10884521056312; 8888145044C, 1924600102, 20884521056319; 8888145044C, 1926100278, 20884521056319; 8888145044C, 1926100283, 20884521056319; 8888145044C, 1926100284, 20884521056319; 8888145044C, 1928100093, 20884521056319; 8888145044C, 2014900157, 20884521056319; 8888145044C, 2017600094, 20884521056319; 8888145044C, 2023300171, 20884521056319; 8888145044C, 2026000082, 20884521056319; 8888145044C, 2026000084, 20884521056319; 8888145044C, 2026000085, 20884521056319; 8888145044C, 2026000086, 20884521056319; 8888145044C, 2026000153, 20884521056319; 8888145044C, 2106900135, 20884521056319; 8888145044C, 2106900140, 20884521056319; 8888145044C, 2106900213, 20884521056319; 8888145044C, 2108100058, 20884521056319; 8888145044C, 2120000070, 20884521056319; 8888145045C, 1924600106, 10884521056329; 8888145045C, 1926100279, 10884521056329; 8888145045C, 1926800382, 10884521056329; 8888145045C, 1926800383, 10884521056329; 8888145045C, 2017500126, 10884521056329; 8888145045C, 2026000089, 10884521056329; 8888145045C, 1924600106, 20884521056326; 8888145045C, 1926100279, 20884521056326; 8888145045C, 1926800382, 20884521056326; 8888145045C, 1926800383, 20884521056326; 8888145045C, 2017500126, 20884521056326; 8888145045C, 2026000089, 20884521056326; 8888145045C, 2100500127, 20884521056326; 8888145046C, 2026000091, 10884521056336; 8888145046C, 2026000091, 20884521056333

Distribution

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, DC, GU, & PR. The countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Republic of Korea, Kosovo, Kuwait, Lebanon, Luxembourg, Macao, Malaysia, Maldives, Malta, Mayotte, Mexico, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Viet Nam, Virgin Islands, and Zimbabwe.

Quantity

26,474 units