FDA Enforcement Class II Terminated

D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R

Recall: Z-1317-2018 · Reported April 18, 2018

Enforcement

Recall Number
Z-1317-2018
Event ID
79515
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Smith & Nephew, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
April 18, 2018
Initiation Date
March 7, 2018
Classification Date
April 6, 2018
Termination Date
October 31, 2018
Address
1450 E Brooks Rd, Memphis, TN, 38116-1804, United States

Description

D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R

Reason

One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.

Code Info

Lot Number 17GM04522

Distribution

US, Puerto Rico, Malaysia

Quantity

59 units