FDA Enforcement
Class II
Terminated
AOA Distalizer Lock Nut Screw, found in products listed as: Distal Jet, Spring Jet, Mesial Jet, and Uprighter Jet. Removable functional orthodontic appliances that are to modify the growth of the jaws in a prescribed growth pattern.
Recall: Z-1314-2014
·
Reported April 9, 2014
Enforcement
- Recall Number
- Z-1314-2014
- Event ID
- 67647
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Allesee Orthodontic Appliances
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 9, 2014
- Initiation Date
- February 24, 2014
- Classification Date
- April 1, 2014
- Termination Date
- August 23, 2014
- Address
- 13931 Spring St, Po Box 725, Sturtevant, WI, 53177-1250, United States
Description
AOA Distalizer Lock Nut Screw, found in products listed as: Distal Jet, Spring Jet, Mesial Jet, and Uprighter Jet. Removable functional orthodontic appliances that are to modify the growth of the jaws in a prescribed growth pattern.
Reason
The Lock Nut component of the Jet appliances may be missing the screw collar, which serves as a mechanical stop to prevent over-tightening in appliance assemby. Without this collar, an over-tightened lock nut may stop the piston assembly from moving freely, subsequently preventing intended movement of the teeth.
Code Info
All Devices/Codes Manufactured from August 28, 2013 to November 6, 2013
Distribution
Worldwide Distribution: US (nationwide) and the country of : Canada.
Quantity
411