FDA Enforcement Class II Ongoing

Steris Corporation Endo Cleaning Indicator VERIFY" RESITEST"

Recall: Z-1313-2024 · Reported March 27, 2024

Enforcement

Recall Number
Z-1313-2024
Event ID
94024
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Mckesson Medical-Surgical Inc. Corporate Office
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
March 27, 2024
Initiation Date
January 9, 2024
Classification Date
March 19, 2024
Address
9954 Maryland Drive, Deep Run Iii Ste. 4000, Richmond, VA, 23233, United States

Description

Steris Corporation Endo Cleaning Indicator VERIFY" RESITEST"

Reason

Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.

Code Info

Catalogue Number: LCC101; UDI/DI: 10724995220072

Distribution

US Nationwide distribution in the states of AZ, PA and TX.

Quantity

25 units