FDA Enforcement Class II Terminated

RayStation Radiation Therapy Treatment Planning System; -- RayStation 3.5, RayStation 4.0, RayStation 4.5 and RayStation 4.7. For RayStation 4.7, the issue applies also to machines with fixed jaws, regardless of MLC/jaw position. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Recall: Z-1310-2015 · Reported April 1, 2015

Enforcement

Recall Number
Z-1310-2015
Event ID
70686
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
RAYSEARCH LABORATORIES AB
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 1, 2015
Initiation Date
March 4, 2015
Classification Date
March 20, 2015
Termination Date
September 20, 2017
Address
Sveavaegen 9, Stockholm, N/A, Sweden

Description

RayStation Radiation Therapy Treatment Planning System; -- RayStation 3.5, RayStation 4.0, RayStation 4.5 and RayStation 4.7. For RayStation 4.7, the issue applies also to machines with fixed jaws, regardless of MLC/jaw position. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Reason

An issue with photon dose calculation for DMLC (Dynamic MLC) plans for machines where the MLC is positioned above the jaws, e.g. some Elekta linacs. The magnitude of the error depends on the beam model output factor corrections and on the individual DMLC plan characteristics.

Code Info

Build numbers 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, 4.0.3.4, 4.5.0.19, 4.5.1.14 and 4.7.0.15

Distribution

Nationwide Distribution.

Quantity

Domestic: 245 units