NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: (1) 48145; (2) 48145E

Recall: Z-1309-2024 · Reported March 27, 2024

Enforcement

Recall Number
Z-1309-2024
Event ID
94016
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Mc3 Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
March 27, 2024
Initiation Date
October 11, 2023
Classification Date
March 19, 2024
Address
2555 Bishop Cir W, Dexter, MI, 48130-1563, United States

Description

NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: (1) 48145; (2) 48145E

Reason

May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer

Code Info

GTIN: (1) 10854916006888; (2) 10854916006901,20854916006908 (New GTIN (USA only)After reboxing). All Serial Numbers between 481022923 and 481024039

Distribution

US Nationwide distribution.

Quantity

299 units