FDA Enforcement Class II Ongoing

Product Name: GCX Channel Mounting Accessory Model/Catalog Number: H8-300-40-A

Recall: Z-1309-2022 · Reported July 6, 2022

Enforcement

Recall Number
Z-1309-2022
Event ID
90400
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Skytron, Div. The KMW Group, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 6, 2022
Initiation Date
June 2, 2022
Classification Date
June 30, 2022
Address
5085 Corporate Exchange Blvd SE, Grand Rapids, MI, 49512-5515, United States

Description

Product Name: GCX Channel Mounting Accessory Model/Catalog Number: H8-300-40-A

Reason

Under certain circumstances, the mounting screws may loosen or pull out from the attachment point on the Ergon 3 boom when the GCX Channel Mounting Accessory was used with the VST clamp as a monitor mount on the boom.

Code Info

Lot Code: All units of model H8-300-40-A are affected. Code Info: GCX mounting bracket does not have UDI. Ergon 3 booms after September 2016 have UDI. DI of units manufactured after September 2016 is 10841736100889. Units manufactured prior to September 2016 do not have UDI.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada and Japan.

Quantity

731