FDA Enforcement Class II Ongoing

MicroVue C1-Inhibitor Plus EIA for measuring the amount of functional C1-lnhibitor protein in human plasma or serum, Catalog Number A037, Containing Coated Strips, REF: 4634, with an ID number of 2200 or above

Recall: Z-1308-2023 · Reported April 5, 2023

Enforcement

Recall Number
Z-1308-2023
Event ID
91772
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Quidel Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 5, 2023
Initiation Date
February 1, 2023
Classification Date
March 30, 2023
Address
2005 E State St Ste 100, N/A, Athens, OH, 45701-2125, United States

Description

MicroVue C1-Inhibitor Plus EIA for measuring the amount of functional C1-lnhibitor protein in human plasma or serum, Catalog Number A037, Containing Coated Strips, REF: 4634, with an ID number of 2200 or above

Reason

Kits with affected microtiter plates have the potential to produce low OD values, causing the run to fail, per the Validation requirements of the Package Insert.

Code Info

UDI: 30014613335580; Lot Number: 213783

Distribution

not yet available

Quantity

771 units