FDA Enforcement
Class II
Terminated
Drill Tapered NP 3.5x10 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.
Recall: Z-1306-2020
·
Reported February 26, 2020
Enforcement
- Recall Number
- Z-1306-2020
- Event ID
- 84450
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Nobel Biocare Usa Llc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- February 26, 2020
- Initiation Date
- March 14, 2019
- Classification Date
- February 19, 2020
- Termination Date
- February 15, 2024
- Address
- 22715 Savi Ranch Pkwy, Yorba Linda, CA, 92887-4609, United States
Description
Drill Tapered NP 3.5x10 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.
Reason
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
Code Info
(01)07332747065688(10)73415 (01)07332747065688(10)73537 Catalog# 37075
Distribution
Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.