FDA Enforcement Class II Terminated

Penner Patient Stretcher/Transfer Lift System Models 391000-1 and 392000-1 The Pacific Stretcher/Transfer Lift is used with Whirlpool or Aqua-Aire (air bubbling) bathing system intended for use in nursing homes, hospitals, and assisted living facilities to transfer and/or lift patients under the direct supervision of trained staff. Model 392000-1 is equipped with a scale.

Recall: Z-1306-2015 · Reported April 1, 2015

Enforcement

Recall Number
Z-1306-2015
Event ID
70151
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Penner Mfg Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 1, 2015
Initiation Date
January 30, 2015
Classification Date
March 20, 2015
Termination Date
March 10, 2016
Address
102 Grant St Ste A, Aurora, NE, 68818-3200, United States

Description

Penner Patient Stretcher/Transfer Lift System Models 391000-1 and 392000-1 The Pacific Stretcher/Transfer Lift is used with Whirlpool or Aqua-Aire (air bubbling) bathing system intended for use in nursing homes, hospitals, and assisted living facilities to transfer and/or lift patients under the direct supervision of trained staff. Model 392000-1 is equipped with a scale.

Reason

If the patient sits on either wing, not squarely on the center section of the Stretcher Lift, tipping and injury to the operator or patient could result. The device manual was updated with an additional warning statement.

Code Info

Serial numbers: 01050741201, 01071198001, 01071198002, 01071198003, 01071198004, 01091136401, 01091136402, 01091136403, 01101314201, 01101322801, 01111505901, 02071228401, 02071228402, 02071228403, 02071228404, 02080975001, 02080975002, 02091137801, 02091149801, 02121713401, 03050791101, 03050804301, 03050804302, 03050804303, 03050804304, 03050804305, 03050804306, 03071247201, 03071247202, 03071247203, 03071247204, 03080984601, 03080984602, 03080984603, 03091161601, 03101347201, 03121742801, 03122862901, 04040581301, 04040581302, 04040581303, 04040581304, 04040581305, 04040581306, 04040581307, 04040581308, 04040581310, 04040581311, 04040581312, 04040581313, 04040581314, 04040586101, 04050810401, 04050810402, 04050810403, 04050810404, 04050810406, 04050810408, 04050810410, 04061028101, 04061028102, 04061028103, 04061028104, 04061028105, 04081009901, 04091187201, 04111563801, 04111568501, 05071272201, 05071272202, 05071272203, 05071272204, 05071272205, 05081019501, 05091197901, 05091197902, 05091201301, 05111513201, 05111576501, 05121762001, 06050840801, 06050840802, 06050840803, 06050840804, 06050840805, 06050840806, 06081037501, 06081039402, 06081039403, 06101394201, 07071320201, 07071320202, 07071328501, 07071328502, 07071328503, 07091223901, 07091223902, 07101410901, 07101415901, 08040660501, 08040660502, 08061090801, 08061090802, 08061090803, 08061090804, 08101430502, 08101439601, 09081073301, 09081074401, 09081076501, 09101453401, 09101457101, 09111641101, 09111641301, 10061121201, 10061121202, 10061121203, 10061121204, 10071396601, 10071396602, 10071396603, 10071396604, 10101463201, 10101464701, 10111652001, 11040710501, 11040710502, 11040710503, 11040710504, 11040710505, 11040710506, 11040710507, 11040710508, 11040710601, 11040710602, 11040710603, 11040710604, 11040710605, 11040710606, 11040710607, 11040710608, 11081101401, 11081105901, 11081113501, 11091285601, 11091285602, 11101475201, 11101476201, 11101476401, 11101477901, 11101478101, 12101495201, 12111677201, 12111686301, 12111686302, 12111688401, 03121734201S, 04040581301A

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to Canada.

Quantity

158