FDA Enforcement
Class II
Terminated
i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and respiratory based acid-based disturbances.
Recall: Z-1306-2014
·
Reported April 2, 2014
Enforcement
- Recall Number
- Z-1306-2014
- Event ID
- 67513
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Abbott Point Of Care Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 2, 2014
- Initiation Date
- January 1, 2014
- Classification Date
- March 27, 2014
- Termination Date
- October 20, 2014
- Address
- 400 College Rd E, N/A, Princeton, NJ, 08540-6607, United States
Description
i-STAT G3+ Cartridge pH, pCO2 and pO2 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and respiratory based acid-based disturbances.
Reason
Abbott Point of Care has determined that some individual patient results generated with the i-STAT G3+ cartridge lot N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results.
Code Info
List number 03P78-25 :Lot number N13183
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada.
Quantity
48, 450 US, 1400 OUS (49,850 catridges)