FDA Enforcement Class II Ongoing

TENSIVE Conductive Adhesive Gel, REF 016-401600

Recall: Z-1305-2024 · Reported March 27, 2024

Enforcement

Recall Number
Z-1305-2024
Event ID
93969
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Natus Neurology Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 27, 2024
Initiation Date
February 7, 2024
Classification Date
March 19, 2024
Address
3150 Pleasant View Rd, Middleton, WI, 53562-4800, United States

Description

TENSIVE Conductive Adhesive Gel, REF 016-401600

Reason

Natus has become aware that its third-party supplier Parker Laboratories Inc. is conducting a voluntary recall related to their Tensive Conductive Adhesive Gel, Ref. 22-60 Lot Number A0623009. They have received five complaints. These customers reported that the gel is too thick, is discolored (darker than usual), and in one instance, patients reported irritation after procedures using the gel.

Code Info

UDI/DI 00855683006333, Lot Numbers: A0623009

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Quantity

80 units