FDA Enforcement Class II Terminated

Dynavox T15 Speech Generating Device ( including Batteries)

Recall: Z-1305-2018 · Reported April 11, 2018

Enforcement

Recall Number
Z-1305-2018
Event ID
79433
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Dynavox Systems Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 11, 2018
Initiation Date
December 14, 2017
Classification Date
April 3, 2018
Termination Date
December 11, 2019
Address
2100 Wharton St Ste Ste 400, N/A, Pittsburgh, PA, 15203-1691, United States

Description

Dynavox T15 Speech Generating Device ( including Batteries)

Reason

The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.

Code Info

TL1405 TL1407 Part Numbers: T15: 750423 Date Codes are part of the device s unique Serial Number and represent the year and month the device was manufactured.

Distribution

Nationally

Quantity

307