da Vinci Xi Surgical System, IS4000; ASSY,USM,IS4000, Material number: 380647-20 and 380647-22 The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endscopic retractors, electrocautery and accessories for edoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic labaroscopic surgical procedures, general thorascopic surgical procedures and thorascopically-assisted cardiotomy procedures.
Enforcement
- Recall Number
- Z-1304-2015
- Event ID
- 70705
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Intuitive Surgical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 1, 2015
- Initiation Date
- March 3, 2015
- Classification Date
- March 20, 2015
- Termination Date
- September 8, 2015
- Address
- 1266 Kifer Rd Bldg 100, Sunnyvale, CA, 94086-5304, United States
Description
da Vinci Xi Surgical System, IS4000; ASSY,USM,IS4000, Material number: 380647-20 and 380647-22 The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endscopic retractors, electrocautery and accessories for edoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic labaroscopic surgical procedures, general thorascopic surgical procedures and thorascopically-assisted cardiotomy procedures.
Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.
Material number: 380647-20; Serial numbers: 1311140430, 1311140586, 1311140058, 1311140392. 1311140277, 1311140307, 1311140094, 1311140100, 1311140569, 1311140102, 1311140421, 1311140092, 1311140009, 1311140070, 1311140267, 1311140333, 1311140492, 1311140391, 1311140057, 1311140294, 1311140508, 1311140565, 1311140105, 1311140122, 1311140469, 1311140521, 1311140378, 1311140456, 1311140230 1311140573, 1311140065, 1311140164, 1311140194, 1311140511, 1311140291, 1311140304, 1311140212, 1311140222, 1311140001, 1311140137, 1311140344, 1311140381, 1311140341, 1311140538, 1311140168, 1311140523, 1311140274, 1311140033, 1311140091 1311140248, 1311140141, 1311140436, 1311140418, 1311140548 1311140320, 1311140170, 1311140346. Material number 380746-22: Serial numbers: 1311140486 1311140539, 1311140174, 1311140183, 1311140327, 1311140347 1311140482, 1311140145, 1311140066, 1311140302, 1311140477 1311140084, 1311140211, 1311140103, 1311140083, 1311140050, 1311140488, 1311140365, 1311140207, 1311140459, 1311140557, 1311140423, 1311140045, 1311140440, 1311140231.
Worldwide Distribution: US (nationwide) and countries of: Australia, Belgium, China, France, Germany, India, Israel, Italy, Norway, South Korea, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
167 total affected da Vinci Xi systems