FDA Enforcement Class II Ongoing

Access Total T4, REF 33800, IVD, thyroxine reagent assay.

Recall: Z-1303-2024 · Reported March 27, 2024

Enforcement

Recall Number
Z-1303-2024
Event ID
94012
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 27, 2024
Initiation Date
February 8, 2024
Classification Date
March 15, 2024
Address
1000 Lake Hazeltine Dr, Chaska, MN, 55318-1037, United States

Description

Access Total T4, REF 33800, IVD, thyroxine reagent assay.

Reason

Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.

Code Info

Lot numbers: 338555, exp. 31 JUL 2024; 338556, exp. 31 AUG 2024; 338734, exp. 30 SEP 2024; and 338958, exp. 31 OCT 2024; UDI-DI 15099590225742.

Distribution

Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Bangladesh, Bolivia, Brazil, Canada, China, Croatia, Czech Republic, Denmark, Egypt, El Salvador, Georgia, Greece, Hong Kong, India, Kazakhstan, Mexico, Namibia, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Taiwan, Thailand, Turkey, U.A.E, Vietnam, and Zimbabwe.

Quantity

64,009 kits