FDA Enforcement Class II Terminated

In-View Clear Leggings Bulk Non-Sterile

Recall: Z-1302-2021 · Reported March 31, 2021

Enforcement

Recall Number
Z-1302-2021
Event ID
87431
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cardinal Health 200, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 31, 2021
Initiation Date
February 25, 2021
Classification Date
March 24, 2021
Termination Date
January 9, 2024
Address
3651 Birchwood Dr, N/A, Waukegan, IL, 60085-8337, United States

Description

In-View Clear Leggings Bulk Non-Sterile

Reason

Cardinal Health has determined that a manufacturing variance in a plastic film used to manufacture these products is causing the drapes and leggings to stick , which may prevent the products from being unfolded and applied properly.

Code Info

Catalog Number 8430NA; Lot number's: 2460DR2, 2550DR1, 2660DR1, 2470DR2, 2580DR1, 2690DR1, 2480DR2, 2580DR2, 2720DR1, 2520DR1, 2600DR1, 2730DR1, 2530DR1, 2600DR1, 3290DR1, 2540DR1, 2610DR1 3310DR1, 2540DR2, 2650DR1, 3440DR1

Distribution

Distributed nationwide to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Distributed also to Australia and Canada.

Quantity

21343 units