FDA Enforcement Class II Terminated

da Vinci Xi Surgical System, IS4000; Assy, Distal SUJ, Inner, IS4000 Material number: 380662-20 and 380662-21. The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endscopic retractors, electrocautery and accessories for edoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic labaroscopic surgical procedures, general thorascopic surgical procedures and thorascopically-assisted cardiotomy procedures.

Recall: Z-1302-2015 · Reported April 1, 2015

Enforcement

Recall Number
Z-1302-2015
Event ID
70705
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Intuitive Surgical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 1, 2015
Initiation Date
March 3, 2015
Classification Date
March 20, 2015
Termination Date
September 8, 2015
Address
1266 Kifer Rd Bldg 100, Sunnyvale, CA, 94086-5304, United States

Description

da Vinci Xi Surgical System, IS4000; Assy, Distal SUJ, Inner, IS4000 Material number: 380662-20 and 380662-21. The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endscopic retractors, electrocautery and accessories for edoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic labaroscopic surgical procedures, general thorascopic surgical procedures and thorascopically-assisted cardiotomy procedures.

Reason

Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.

Code Info

Material number: 3680662-20: Serial numbers: 1311140273, 1311140370, 1311140385, 1311140148, 1311140099, 1311140152, 1311140081, 1311140473, 1311140561, 1311140409. Material number: 380662-21: Serial numbers: 1311140225, 1311140437, 1311140476, 1311140336, 1311140472, 1311140438, 1311140353, 1311140195, 1311140545, 1311140098 1311140189, 1311140465, 1311140394, 1311140366, 1311140008, 1311140115 1311140014, 1311140371, 1311140238, 1311140306, 1311140332, 1311140404 1311140054, 1311140188, 1311140324, 1311140453, 1311140408, 1311140237 1311140220, 1311140372, 1311140325, 1311140088, 1311140535, 1311140355.

Distribution

Worldwide Distribution: US (nationwide) and countries of: Australia, Belgium, China, France, Germany, India, Israel, Italy, Norway, South Korea, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Quantity

167 total da Vinci Xi affected systems